pNeuton miniRecalls & reported problems
Airon Corporation · pNeuton Ventilator, mini · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00853678006023), received between Dec 11, 2015 and Dec 11, 2015.
The most frequently reported specific device problem is “Maintenance Does Not Comply To Manufacturers Recommendations” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Maintenance Does Not Comply To Manufacturers Recommendations1
- Use of Device Problem1
Patient problems
- Missing Value Reason1
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
Dec 11, 2015Use of Device Problem; Maintenance Does Not Comply To Manufacturers Recommendations; Adverse Event Without Identified Device or Use Problem
PATIENT WAS BROUGHT TO ANTEROOM OF THE MRI FOR SCAN. CUSTOMARY FOR THE MRI SCAN THE PATIENT WAS INTUBATED IN THE MRI SUITE IMMEDIATELY PRIOR TO THE SCAN. VENTILATION WAS ESTABLISHED USING AN ANESTHESIA MACHINE (DRAEGER) POST INDUCTION/INTUBATION. THE PATIENT SETTINGS WERE PC: PIP 18, PEEP 3, TI 0.7 SEC, WITH RATE OF 30 BPM. MEASURED TIDAL VOLUMES WERE MID 30S WITH APPROPRIATE VITAL SIGNS AND CHEST RISE, MONITORING PETCO2 OF 45 AS REPORTED BY THE RESPIRATORY THERAPIST. THE PNEUTON MINI OPERATIONAL VERIFICATION WAS COMPLETED AT THE BEDSIDE WITHOUT ANY NOTED ERRORS. PNEUTON MINI PATIENT SETTINGS WERE INITIALLY PLACED ON IMV/CPAP: PIP 20, PEEP 5, TI 0.4 SEC, TE 2.0 SEC WITH FLOW RATE 8 LPM. CLINICAL OBSERVATION NOTED A DECREASE IN PATIENT CHEST WALL MOVEMENT AND PETCO2 DROPPED TO 25 WITH ABSENCE OF AN ALVEOLAR PLATEAU. DESPITE MULTIPLY VENTILATOR CHANGES AT THE BEDSIDE; TI SEC, FLOW RATE AND PEEP, THEY WERE UNABLE TO OBTAIN A NORMAL CAPNOGRAPH AND CHEST WALL MOVEMENT REMAINED LESS THAN THE ANESTHESIA MACHINE. A SECOND PNEUTON MINI WAS SET-UP FOR PATIENT APPLICATION, PASSED OPERATIONAL VERIFICATION AND APPLIED THE SAME PATIENT SETTINGS, CHANGES IN DIFFERENT VENTILATOR SETTINGS AS NOTED
Patient problems: Missing Value Reason; No Information; No Known Impact Or Consequence To Patient
MAUDE report 5290618 ↗
Other Ventilator, Continuous, Facility Use devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.