NEUWAVE LK20Recalls & reported problems
Neuwave Medical, Inc. · NEUWAVE LK Probe 20CM 17ga · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00853719006371), received between Mar 13, 2020 and Feb 15, 2023 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break2
- Appropriate Device Problem Term/Code Not Available1
- Material Twisted/Bent1
Patient problems
- Foreign Body In Patient2
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 15, 2023Material Twisted/Bent; Appropriate Device Problem Term/Code Not Available; Material Twisted/Bent; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING ABLATION CASE, THE PROBE BROKE (BENT) DURING THE ABLATION CASE INSIDE THE PATIENT. IN ADDITIONAL TO THAT, THE ABLATION ZONE HAS BEEN SHIFTING BACKWARD 2 CM FROM WHAT IT WAS EXPECTED, WHICH MAKE THE ABLATION AREA HIT THE LIVER SURFACE. THE PROCEDURE WAS DELAYED 20 MINUTES. THE PROCEDURE WAS STOPPED AND THE TUMOR WASN'T ABLATED. THE PATIENT EXPERIENCED SEVERE PAIN.
Patient problems: Pain
MAUDE report 16381125 ↗InjuryDec 2, 2020Break; Break
IT WAS REPORTED THAT DURING A LIVER ABLATION THE PROBE BROKE WHILE IN THE PATIENTS LIVER. THERE WERE NO PATIENT CONSEQUENCES AND THERE IS NO ADDITIONAL INFORMATION.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 10931434 ↗InjuryMar 13, 2020Break
IT WAS REPORTED THAT DURING AN ABLATION THE PROBE BROKE INSIDE THE PATIENT. THE METAL TIP WAS LEFT INSIDE THE LIVER. A SIMILAR DEVICE WAS USED TO COMPLETE THE CASE.
Patient problems: Foreign Body In Patient
MAUDE report 9829033 ↗
Other SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES devices
- Emprint CA20L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls1 death43 reports
- Emprint CA15L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls1 death41 reports
- Emprint CA30L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls2 reports
- Emprint CA15L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall49 reports
- Solero 700106001-USAngiodynamics, Inc. · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES92 reports
- NEUWAVE PR15XTNeuwave Medical, Inc. · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES90 reports
- Emprint CA20L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall18 reports
- Emprint CA30L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall1 death9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗