Remedy Spectrum GV Spacer GVHDSMRecalls & reported problems
Osteoremedies, LLC · REMEDY SPECTRUM GV Femoral Head - 46mm · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00855195006524), received between Jun 11, 2025 and Dec 16, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Loss of or Failure to Bond2
Patient problems
- Inadequate Osseointegration1
- Limb Fracture1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 16, 2025Loss of or Failure to Bond; Fracture; Loss of or Failure to Bond; Fracture
JOURNAL OF ORTHOPAEDIC EXPERIENCE AND INNOVATION VOL. 6, ISSUE2, 2025 INCLUDED RESERACH ARTICLE "STABILITY OF OSTEOREMEDIES REMEDY HIP SPACER FOR TWO-STAGE REVISIONS OF PROSTHETIC JOING INFECTIONS: SINGLE CENTER ANALYSIS" INCLUDED TWO PATIENTS WHO REQUIRED SPACER REPLACMENT. INFORMATION FROM THE ARTICLE: IN THIS SERIES, THE OSTEOREMEDIES® SECOND-GENERATION PRE-MOLDED ANTIBIOTIC HIP SPACER PROVED TO BE A CONVENIENT AND EFFECTIVE OPTION DURING TWO-STAGE REVISIONS FOR MANAGING PJIS, THOUGH IT DEMONSTRATED A SMALL RISK OF IMPLANT FRACTURE AND DISLOCATION. THIS SPACER DEMONSTRATED ONLY 1 FRACTURE AND 1 DISLOCATION OF THE 16 PATIENTS WHO USED BOTH STEM AND HEAD PIECES. OF THE 16 PATIENTS WHO USED BOTH FEMORAL COMPONENTS, A SPACER FRACTURE OCCURRED IN ONE CASE (6.25%). THIS PATIENT UNDERWENT A SPACER EXCHANGE. BASED ON INITIAL POST-OPERATIVE X-RAYS, THE DEGREE OF THREADING OF THE FEMORAL NECK INTO THE HEAD WAS 6 THREADS, BUT AT THE TIME OF FRACTURE, IT HAD REDUCED TO ONLY 2 THREADS. THIS PATIENT DID NOT RECEIVE ADDITIONAL BONE CEMENT, AND THEIR BMI WAS 34.75 KG/M². WHEN COMPARING THIS PATIENT WHO EXPERIENCED A SPACER FRACTURE TO ANOTHER IN OUR COHORT WITH THE SAME THREAD COUNT OF 6 WHO DI
Patient problems: Inadequate Osseointegration
MAUDE report 23824711 ↗InjuryJun 11, 2025Loss of or Failure to Bond; Fracture; Loss of or Failure to Bond; Fracture
THE PATIENT HAD A HIP SPACER IMPLANTED ON (B)(6) 2025 AS PART OF A TWO-STAGE SURGERY TO ADDRESS THE PATIENT'S PRIMARY JOINT INFECTION. POST OPERATIVE IMAGES TAKEN ON (B)(6) 2025 SHOW A BONE FRACTURE IN THE FEMUR THAT IS ASSUMED TO HAVE TAKEN PLACE INTRA-OPERATIVELY DURING EXPLANTATION OF THE PREVIOUS DEVICE AND THE SPACER IMPLANTATION. POST OPERATIVELY THE FEMORAL FRACTURE PROPAGATED ONCE PATIENT WEIGHT WAS APPLIED AND THE SPACER HEAD DISASSOCIATED FROM THE SPACER STEM. ON (B)(6) 2025, THE PATIENT WAS TAKEN BACK TO SURGERY AND REVISED WITH A GIRDLESTONE AS THE PATIENT WAS STILL INFECTED (CULTURE POSITIVE).
Patient problems: Limb Fracture
MAUDE report 22188000 ↗
Other Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented devices
- STAGEONE 431209Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 recalls0 reports
- SELF CENTERING 1035-45-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented84 reports
- LFIT 6260-9-036Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented80 reports
- ACTIS 101011060Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented66 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165228Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall1 death12 reports
- SELF CENTERING 1035-44-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented60 reports
- TANDEM 71324046Smith & Nephew, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall9 reports
- TAPERLOC COMPLETE XR 123 MICRO FEMORAL 51-145170Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented59 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page