Skip to content

Remedy Spectrum GV Spacer GVHDSMRecalls & reported problems

Osteoremedies, LLC · REMEDY SPECTRUM GV Femoral Head - 46mm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00855195006524), received between Jun 11, 2025 and Dec 16, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Loss of or Failure to Bond2

Patient problems

  • Inadequate Osseointegration1
  • Limb Fracture1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 16, 2025Loss of or Failure to Bond; Fracture; Loss of or Failure to Bond; Fracture

    JOURNAL OF ORTHOPAEDIC EXPERIENCE AND INNOVATION VOL. 6, ISSUE2, 2025 INCLUDED RESERACH ARTICLE "STABILITY OF OSTEOREMEDIES REMEDY HIP SPACER FOR TWO-STAGE REVISIONS OF PROSTHETIC JOING INFECTIONS: SINGLE CENTER ANALYSIS" INCLUDED TWO PATIENTS WHO REQUIRED SPACER REPLACMENT. INFORMATION FROM THE ARTICLE: IN THIS SERIES, THE OSTEOREMEDIES® SECOND-GENERATION PRE-MOLDED ANTIBIOTIC HIP SPACER PROVED TO BE A CONVENIENT AND EFFECTIVE OPTION DURING TWO-STAGE REVISIONS FOR MANAGING PJIS, THOUGH IT DEMONSTRATED A SMALL RISK OF IMPLANT FRACTURE AND DISLOCATION. THIS SPACER DEMONSTRATED ONLY 1 FRACTURE AND 1 DISLOCATION OF THE 16 PATIENTS WHO USED BOTH STEM AND HEAD PIECES. OF THE 16 PATIENTS WHO USED BOTH FEMORAL COMPONENTS, A SPACER FRACTURE OCCURRED IN ONE CASE (6.25%). THIS PATIENT UNDERWENT A SPACER EXCHANGE. BASED ON INITIAL POST-OPERATIVE X-RAYS, THE DEGREE OF THREADING OF THE FEMORAL NECK INTO THE HEAD WAS 6 THREADS, BUT AT THE TIME OF FRACTURE, IT HAD REDUCED TO ONLY 2 THREADS. THIS PATIENT DID NOT RECEIVE ADDITIONAL BONE CEMENT, AND THEIR BMI WAS 34.75 KG/M². WHEN COMPARING THIS PATIENT WHO EXPERIENCED A SPACER FRACTURE TO ANOTHER IN OUR COHORT WITH THE SAME THREAD COUNT OF 6 WHO DI

    Patient problems: Inadequate Osseointegration

    MAUDE report 23824711 ↗
  • InjuryJun 11, 2025Loss of or Failure to Bond; Fracture; Loss of or Failure to Bond; Fracture

    THE PATIENT HAD A HIP SPACER IMPLANTED ON (B)(6) 2025 AS PART OF A TWO-STAGE SURGERY TO ADDRESS THE PATIENT'S PRIMARY JOINT INFECTION. POST OPERATIVE IMAGES TAKEN ON (B)(6) 2025 SHOW A BONE FRACTURE IN THE FEMUR THAT IS ASSUMED TO HAVE TAKEN PLACE INTRA-OPERATIVELY DURING EXPLANTATION OF THE PREVIOUS DEVICE AND THE SPACER IMPLANTATION. POST OPERATIVELY THE FEMORAL FRACTURE PROPAGATED ONCE PATIENT WEIGHT WAS APPLIED AND THE SPACER HEAD DISASSOCIATED FROM THE SPACER STEM. ON (B)(6) 2025, THE PATIENT WAS TAKEN BACK TO SURGERY AND REVISED WITH A GIRDLESTONE AS THE PATIENT WAS STILL INFECTED (CULTURE POSITIVE).

    Patient problems: Limb Fracture

    MAUDE report 22188000 ↗

Other Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented devices