Skip to content

TANDEM 71324046Recalls & reported problems

Smith & Nephew, Inc. · TANDEM INTERNATIONAL BIPOLAR 46MM OUTER DIAMETER 28MM INNER DIAMETER · Class 2 · Implant

All models of TANDEM (58)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
8
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010494375), received between Jan 14, 2020 and May 1, 2026 — classified as 8 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K023743, K211176, K231448); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 15, 2024Open, Classified

    The Bipolar assembly contains an oversized Retainer Ring.

    Z-0326-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated4
  • Adverse Event Without Identified Device or Use Problem3
  • Deformation Due to Compressive Stress1
  • Insufficient Device Problem Information1
  • Manufacturing, Packaging or Shipping Problem1
  • Positioning Failure1

Patient problems

  • Joint Dislocation4
  • No Clinical Signs, Symptoms or Conditions2
  • Failure of Implant1
  • Injury1
  • Loss of Range of Motion1
  • Pain1
  • Post Operative Wound Infection1
  • Rupture1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 1, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT, AFTER AN UKA SURGERY PERFORMED ON (B)(6) 2026, THE PATIENT EXPERIENCED SYMPTOMS OF PROSTHETIC DISLOCATION, AND AN X-RAY REVEALED DETACHMENT OF THE BIPOLAR CUP FROM THE STEM AND THE HEAD. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2026, IN WHICH IT WAS NOTICED THAT THE IMPLANTS WERE NOT BROKEN, THAT THEY HAD BEEN CORRECTLY ASSEMBLED INITIALLY, AND THAT THEY HAD BECOME DETACHED, WITH THE DOME AND RING STILL ATTACHED AND THE HEAD STILL ATTACHED TO THE STEM. THE NEW JOINT REMAINED STABLE AFTER THE NEW COMPONENTS WERE PLACED.

    Patient problems: Joint Dislocation

    MAUDE report 25061998 ↗
  • InjuryJan 23, 2025Device Dislodged or Dislocated; Manufacturing, Packaging or Shipping Problem; Device Dislodged or Dislocated; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT, ON (B)(6) 2024, THE PATIENT CAME IN FOR A HEMIARTHROPLASTY WHERE A TANDEM INTL BIPOLAR 46OD 28ID WAS IMPLANTED WITHOUT COMPLICATIONS. ON (B)(6) 2025 THE PATIENT REPORTED TO ED AND IT WAS FOUND THAT THE IMPLANTED BIPOLAR HAD DISASSOCIATED. THE PATIENT'S IMPLANT DISASSOCIATED IMPACTING THEIR MOBILITY AND INCREASING THEIR PAIN. THE PATIENT WAS BROUGHT IN FOR A REVISION SURGERY ON (B)(6) 2025. NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Failure of Implant; Joint Dislocation; Loss of Range of Motion; Pain

    MAUDE report 21217814 ↗
  • MalfunctionOct 18, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT, DURING A LEFT HIP HEMIARTHROPLASTY PERFORMED DUE TO MEDIAL DISLOCATED FEMORAL NECK FRACTURE, IT BECAME APPARENT THAT THE CONNECTION OF A TANDEM INTL BIPOLAR 46OD 28ID; BETWEEN THE INLAY AND THE RING WAS INADEQUATE. THE INLAY WOULD COME LOOSE AT THE SLIGHTEST TOUCH/TWIST, AND THE METAL RING DID NOT CLOSE AS USUAL (APPARENTLY TOO LARGE). THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A SIZE 44 TANDEM WITH A 28 XS HEAD. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20487471 ↗
  • InjuryApr 13, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT, AFTER A HIP HEMIARTHROPLASTY SURGERY HAD BEEN PERFORMED ON (B)(6) 2023, SEVEN DAYS AFTERWARDS THE PATIENT EXPERIENCED A DISLOCATION DURING TRACTION REHABILITATION TRAINING. IT IS UNKNOWN HOW THIS EVENT WAS SOLVED OR TREATED. PATIENT STATUS IS UNKNOWN.

    Patient problems: Joint Dislocation

    MAUDE report 16731825 ↗
  • InjuryDec 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A PARTIAL ARTHROPLASTY HAD BEEN PERFORMED ON (B)(6) 2021 TO REPAIR A FEMUR FRACTURE, THE PATIENT HAD TWO CONSECUTIVE FALLS AFTERWARDS, WHICH CAUSED TEARS IN ALL SOFT TISSUE SUPPORT STRUCTURES AROUND THE HIP JOINT. REDAPT SLVLS MONO STEM 240MM SZ 17 SO, COCR 12/14 FEM HEAD 28 +0 AND TANDEM INTL BIPOLAR 46OD 28ID WERE EXPLANTED. THE PATIENT WAS LEFT WITH A GURDLE STONE HIP. THE PATIENT CURRENT HEALTH STATUS IS GOOD.

    Patient problems: Rupture

    MAUDE report 13089690 ↗
  • InjuryJun 14, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A HIP HEMIARTHROPLASTY HAD BEEN PERFORMED WITH A TANDEM SYSTEM ON (B)(6) 2021 TO TREAT A FEMUR FRACTURE, THE PATIENT SUFFERED FROM A HIP DISLOCATION SIX DAYS AFTERWARDS. IT IS UNKNOWN HOW THIS ADVERSE EVENT WAS ADDRESSED AND SO IS THE PATIENT CURRENT STATUS OF HEALTH.

    Patient problems: Joint Dislocation

    MAUDE report 11991459 ↗
  • InjuryJan 6, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED ON (B)(6) 2020 FOR A FEMORAL NECK FRACTURE AND UNDERWENT ARTIFICIAL FEMORAL HEAD REPLACEMENT USING THIS HEMI HIP PROSTHESIS COMPONENT (PRIMARY SURGERY). POSTOPERATIVELY, ON (B)(6) 2020, THE PATIENT DEVELOPED SYMPTOMS OF INFECTION SUCH AS WOUND REDNESS AND SWELLING WITH FLUID EXUDATION. TREATMENT: IMMEDIATE DEBRIDEMENT, LOCAL IODOPHOR AND SALINE IRRIGATION WERE GIVEN, AND THE ABOVE SYMPTOMS GRADUALLY IMPROVED AND DISAPPEARED.

    Patient problems: Post Operative Wound Infection

    MAUDE report 11123104 ↗
  • InjuryDec 6, 2020Deformation Due to Compressive Stress; Insufficient Device Problem Information; Deformation Due to Compressive Stress; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING A THR EVERYTHING WAS GOING NORMAL AND DR. DECIDED TO POSITION THE 46MM CUP (REF: (B)(4) LOT: R2008440) AND HEAD 28MM +4 (REF: (B)(4)) BUT POLYETHYLENE RING DID NOT FIT CORRECTLY AND SURGEON DECIDED TO IMPACT IT. DURING THE IMPACTION, A SMALL PIECE IN THE RING DAMAGED. NO BACKUP DEVICE WAS AVAILABLE AND ITEM WAS LEFT DAMAGED INSIDE OF THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10955865 ↗
  • InjuryJan 14, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED SINCE THE HEAD WAS OUT OF THE BIPOLAR BALL HEAD ONE DAY AFTER PRIMARY SURGERY. PATIENT WAS IN GOOD CONDITION.

    Patient problems: Injury

    MAUDE report 9586278 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K023743, K211176, K231448. One submission often covers many models, so check the product and lot details in each record.

  • Z-0350-2017 · Aug 29, 2016 · Terminated
    TANDEM (TM) COCR SHELL UHMWPE LINER BIPOLAR, REF NUMBERS: 71322040, 71322041, 71322042, 71322043, 71322044, 71322045, 71322046, 71322047, 71322048, 71322049, 71322050, 71322051,
  • Z-0351-2017 · Aug 29, 2016 · Terminated
    TANDEM (TM) INTL COCR SHELL UHMWPE LINER BIPOLAR, REF NUMBERS: 71324038, 71324039, 71324040, 71324041, 71324042, 71324043, 71324044, 71324045, 71324046, 71324047, 71324048, 7132

Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices

All Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page