VOCSN-VC VOCSN-VC Pro, EnglishRecalls & reported problems
Ventec Life Systems, Inc. · VOCSN-VC Pro, English: The VC configuration of VOCSN contains Ventilation and Cough therapies. The VOCSN Unified Respiratory System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It may be used in invasive and non-invasive applications. The VOCSN is intended for pediatric through adult patients weighing at least 5 kg. It is intended for use in home, hospital, institutional and transport settings, including portable applications. The integral cough assist option is intended for patients who are additionally unable to cough or clear secretions effectively. www.VentecLife.com · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 21
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 00855573007433), received between Feb 6, 2019 and Mar 10, 2026 — classified as 21 malfunctions.
The most frequently reported specific device problem is “Output Problem” (15 reports).
5 other recalls cover devices cleared under the same FDA submission (K162877); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Output Problem15
- Unexpected Shutdown3
- Display or Visual Feedback Problem2
- Connection Problem1
- Power Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions19
- No Patient Involvement2
Recent adverse event reports
The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 10, 2026Output Problem
IT WAS REPORTED TO VENTEC THAT THE DEVICE'S EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT; HOWEVER, THERE WAS NO PATIENT HARM AS A RESULT OF THE REPORTED ISSUE. THE PATIENT WAS PLACED ON THEIR BACKUP VENTILATOR FOR SUPPORT. NO FURTHER DETAILS ABOUT THE PATIENT WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24561055 ↗MalfunctionJan 6, 2025Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED TO VENTEC THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT IT UNEXPECTEDLY POWERED OFF BY ITSELF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21085113 ↗MalfunctionNov 18, 2024Power Problem; Power Problem
IT WAS REPORTED TO VENTEC THAT THE DEVICE FROZE IN THE MIDDLE OF A SOFTWARE UPDATE AND IS NOW INOPERABLE. VENTEC HAS MADE MULTIPLE ATTEMPTS TO CONTACT THE REPORTER TO GET ADDITIONAL INFORMATION ABOUT THE REPORTED ISSUE, BUT NO RESPONSE HAS BEEN RECEIVED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20717576 ↗MalfunctionJul 18, 2024Output Problem; Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE'S EVALUATION VENTEC OBSERVED THAT IT WAS PROVIDING LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19774021 ↗MalfunctionJun 21, 2024Output Problem; Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT THE VENTILATOR MEASURED LOWER PEEP (POSITIVE END EXPIRATORY PRESSURE) THAN SET AND PROVIDED A LOWER THAN EXPECTED BREATH RATE. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19582339 ↗MalfunctionJun 7, 2024Output Problem; Output Problem
IT WAS REPORTED TO VENTEC THAT THE DEVICE'S "O2 SENSOR NOT WORKING." NO FURTHER DETAILS WERE PROVIDED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19481545 ↗MalfunctionMay 17, 2024Output Problem; Output Problem
IT WAS INITIALLY REPORTED TO VENTEC THAT THE DEVICE HAD PROVIDED A LOW FIO2 (FRACTION OF INSPIRED OXYGEN) ALERT DURING AN EVENT INVOLVING A PATIENT. VENTEC REACHED OUT TO THE REPORTER IN ORDER TO OBTAIN ADDITIONAL INFORMATION ABOUT THE PATIENT AND THE EVENT, AND WAS ADVISED THAT THEY WEREN'T SURE IF THEY COULD RELEASE THAT INFORMATION, AND THAT "I KNOW THE VENT WAS NOT ON THE PATIENT AND NO PATIENT HARM." VENTEC HAS REACHED OUT TO THE REPORTER MULTIPLE TIMES IN ORDER TO GET CLARIFICATION AS TO WHETHER OR NOT THERE WAS PATIENT USE. NO RESPONSE HAS BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19347742 ↗MalfunctionFeb 20, 2024Output Problem; Output Problem
IT WAS REPORTED TO VENTEC THAT THE DEVICE'S EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18744596 ↗MalfunctionFeb 20, 2024Connection Problem; Output Problem; Connection Problem; Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT THE VENTILATOR WAS UNABLE TO MAINTAIN PEEP (POSITIVE END EXPIRATORY PRESSURE), THAT ITS EXHALED TIDAL VOLUME (VTE) LEVELS WERE LOW AND THAT IT WAS DISPLAYING A PATIENT CIRCUIT DISCONNECT ALARM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18744562 ↗MalfunctionJul 18, 2023Output Problem; Output Problem
IT WAS REPORTED TO VENTEC THAT THE VENTILATOR PROVIDED AN ALARM INDICATING VERY LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17344870 ↗MalfunctionMay 11, 2023Output Problem; Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED MAINTENANCE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT THE VENTILATOR PROVIDED AN ALARM INDICATING VERY LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16916756 ↗MalfunctionApr 28, 2023Output Problem; Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED MAINTENANCE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT THE VENTILATOR PROVIDED AN ALARM INDICATING VERY LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16829730 ↗MalfunctionApr 27, 2023Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED TO VENTEC THAT THE DEVICE NEEDED MAINTENANCE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THE DEVICE CONTINUOUSLY REBOOTED BY ITSELF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16827962 ↗MalfunctionMar 15, 2023Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED DATA IN THE DEVICE'S MEMORY WHICH INDICATED THAT IT HAD PREVIOUSLY REBOOTED MULTIPLE TIMES BY ITSELF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16550436 ↗MalfunctionMar 15, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT ITS LCD TOUCHSCREEN WAS UNRESPONSIVE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16549969 ↗MalfunctionMar 7, 2023Output Problem; Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED SERVICE. DURING THE DEVICE EVALUATION, VENTEC OBSERVED THAT THE VENTILATOR WAS UNABLE TO MAINTAIN PEEP (POSITIVE END EXPIRATORY PRESSURE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16499225 ↗MalfunctionFeb 14, 2023Output Problem; Output Problem
IT WAS REPORTED TO VENTEC THAT THE DEVICE WAS ALARMING DUE TO LOW FIO2 (FRACTION OF INSPIRED OXYGEN) READINGS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16376413 ↗MalfunctionJan 31, 2023Output Problem; Output Problem
IT WAS REPORTED THAT THE DEVICE NEEDED MAINTENANCE. UPON EVALUATION OF THE DEVICE, VENTEC OBSERVED THAT THE VENTILATOR WAS UNABLE TO MAINTAIN PEEP (POSITIVE END EXPIRATORY PRESSURE). THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16273141 ↗MalfunctionJan 16, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED TO VENTEC BY A THIRD PARTY SERVICE PROVIDER THAT THE LCD TOUCHSCREEN ON THE DEVICE WAS UNRESPONSIVE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16174906 ↗MalfunctionMay 11, 2020Output Problem; Output Problem
IT WAS REPORTED THAT THE DEVICE FAILED THE PRE-USE TEST AND WOULD NOT VENTILATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 10045045 ↗MalfunctionFeb 6, 2019Output Problem; Output Problem
IT WAS REPORTED THAT A VENTILATOR WAS BEING SET UP AND DID NOT PASS THE PRE-USE TEST. IT WAS NOTED THAT THE UNIT WAS NOT VENTILATING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8313041 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K162877. One submission often covers many models, so check the product and lot details in each record.
- Z-2525-2026 · May 8, 2026 · Open, ClassifiedVOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
- Z-1942-2026 · Mar 25, 2026 · Open, ClassifiedVOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
- Z-1196-2025 · Feb 3, 2025 · Open, ClassifiedVOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002
- Z-1070-2024 · Jan 9, 2024 · Open, ClassifiedVOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides c
- Z-1902-2021 · May 10, 2021 · Open, ClassifiedVentec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900
Other Ventilator, Continuous, Facility Use devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗