Inspire Medical Systems, Inc. · The Model 3024 Inspire II IPG contains electronics and a battery that are sealed inside a titanium case. The IPG is implanted subcutaneously below the clavicle in the upper chest, and connects to the stimulation lead and sensing lead. The IPG has multiple programmable parameters that can be adjusted for a particular patient. Adjustments may be made to the sensing circuits to vary the timing of stimulation pulses and to the stimulation circuits to adjust the energy and duration of stimulation pulses. These adjustments provide the ability to optimize the opening of the upper airway. Based on typical settings from the STAR pivotal trial the longevity of the IPG’s battery will average 10 years. The Model 3024 IPG was originally developed by Medtronic and utilizes the same platform as Medtronic’s Itrel® 3 IPG which is FDA approved for other neurostimulation therapies. Figure 9 provides a photograph of the Model 3024 IPG. · Class 3 · Implant

All models of Inspire Sleep Apnea Implant (24)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Inspire II implantable pulse generator is a battery-powered stimulator placed under the skin of the upper chest that delivers mild pulses to the tongue muscle to keep the airway open during sleep. The 45 reports mostly describe pain and discomfort at the implant site, along with infection, redness, drainage and thinning of the skin over the device; several patients had the unit removed or replaced. There are 41 injuries and 4 deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
45
Deaths reported
0
Injuries reported
41
Recalls naming it
0

Summary

The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 00855728005024), received between Jun 9, 2016 and Nov 11, 2025 — classified as 41 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (4 reports).

3 other recalls cover devices cleared under the same FDA submission (P130008); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem27
  • Migration or Expulsion of Device4
  • Appropriate Device Problem Term/Code Not Available3
  • Insufficient Device Problem Information3
  • Migration3
  • Device Operates Differently Than Expected2
  • Electrical /Electronic Property Problem2
  • Device Operational Issue1

Patient problems

  • Pain10
  • Erosion4
  • Unspecified Infection4
  • Dysphasia3
  • No Code Available3
  • Bacterial Infection2
  • Discomfort2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH INTERMITTENT SHOCKING SENSATIONS AND PAIN AT THE IPG SITE. PHYSICIAN BROUGHT THE PATIENT TO THE OR, REMOVED THE IPG, IMPLANTED A NEWER IPG MODEL, SUTURED, AND CLOSED. THE EXISTING SENSING LEAD WAS ABANDONED AS IT IS NOT USED WITH THE NEW IPG MODEL.

    Patient problems: Pain

    MAUDE report 23524205 ↗
  • InjuryOct 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN REQUESTING AN UPGRADE TO A NEWER MRI-COMPATIBLE IPG AND ALSO REPORTED CHEST PAIN. PHYSICIAN BROUGHT THE PATIENT INTO THE OR, REPLACED THE IPG, AND CLOSED.

    Patient problems: Pain

    MAUDE report 23426610 ↗
  • InjuryMay 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    FOLLOWING A FALL, THE PATIENT PRESENTED TO THE PHYSICIAN WITH REDNESS AND SWELLING AT THE IPG SITE. PHYSICAL EXAMINATION REVEALED SIGNS OF INFECTION. THE PHYSICIAN PRESCRIBED ANTIBIOTICS. THE PATIENT RETURNED WITH PERSISTENT INFECTION AND PURULENT DRAINAGE FROM THE INCISION SITE. THE CHEST INCISION WAS DRAINED. THE NEXT DAY, THE PATIENT WAS TAKEN TO THE OR WHERE THE IPG WAS REMOVED, THE LEADS WERE CAPPED, AND THE INCISION SITE WAS CLOSED. POSTOPERATIVE ANTIBIOTICS WERE PRESCRIBED.

    Patient problems: Purulent Discharge; Unspecified Infection

    MAUDE report 21944977 ↗
  • InjuryMar 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE IPG SITE AFTER UNDERGOING AN IMAGING PROCEDURE. PHYSICIAN BROUGHT THE PATIENT INTO THE OPERATING ROOM AND DECIDED TO REPLACE THE IPG AS IT WAS AN OLDER MODEL. PHYSICIAN OPENED THE INCISION, REPLACED THE IPG, RE-POSITIONED IPG RELATIVE TO ORIGINAL IMPLANT, AND CLOSED.

    Patient problems: Pain

    MAUDE report 18838119 ↗
  • InjuryNov 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH SYSTEMIC MRSA INFECTION. PHYSICIAN BROUGHT THE PATIENT INTO THE OPERATING ROOM AND REMOVED THE ENTIRE INSPIRE SYSTEM.

    Patient problems: Unspecified Infection

    MAUDE report 18096922 ↗
  • InjuryMay 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE IPG SITE WHEN LAYING ON THE RIGHT SIDE AND THINNING AREA OF THE TISSUE NEAR THE IPG INCISION. PHYSICIAN BROUGHT THE PATIENT INTO THE OR, EXCISED THE AREA OF SCAR THAT HAD THINNED, REMOVED SOME SCAR TISSUE, AND PLACED THE NEW IPG IN THE SAME POCKET. THIS RESOLVED THE ISSUE.

    Patient problems: Pain

    MAUDE report 16888886 ↗
  • InjuryMay 9, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED IPG EXPOSED AT THE CHEST INCISION SITE 5 MONTHS POST COMPLETION OF CHEMOTHERAPY FOR BREAST CANCER. PHYSICIAN PLACED THE PATIENT ON ANTIBIOTICS AND BROUGHT THE PATIENT INTO THE OR 1 WEEK LATER AND REMOVED THE ENTIRE SYSTEM.

    Patient problems: Erosion

    MAUDE report 14342981 ↗
  • InjuryFeb 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REPORTED TO THE PHYSICIAN THAT THE IPG HAS BEEN MOVING MOBILE AND WAS CAUSING PAIN. PHYSICIAN BROUGHT THE PATIENT INTO THE OPERATING ROOM AND REPLACED THE IPG. THIS RESOLVED THE ISSUE.

    Patient problems: Pain

    MAUDE report 13548633 ↗
  • InjuryFeb 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE SENSE LEAD SITE, REPORTED A LACK OF EFFICACY, AND REQUESTED REMOVAL OF INSPIRE SYSTEM. PHYSICIAN REMOVED THE INSPIRE SYSTEM. THIS RESOLVED THE ISSUE.

    Patient problems: Pain

    MAUDE report 13548567 ↗
  • InjuryOct 14, 2021Unintended Electrical Shock; Unintended Electrical Shock

    PATIENT UNDERWENT CONTRAINDICATED MRI SCAN OF THE LOWER BACK. DURING THE SCAN, THE PATIENT EXPERIENCED INTENSE BURNING STIMULATION IN HIS NECK, CHEST, AND EAR. IMMEDIATELY FOLLOWING THE SCAN, PATIENT REPORTS THAT THEY WERE STILL EXPERIENCING PAIN IN THE THROAT, CHEST, AND EAR AS WELL AS A HEADACHE. THE MRI MACHINE WAS ON FOR LESS THAN ONE MINUTE AND THE DOCTOR CONFIRMED NO TISSUE DAMAGE TOOK PLACE FROM THE EVENT.

    Patient problems: Pain

    MAUDE report 12633214 ↗
  • InjuryFeb 23, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    PATIENT PRESENTED TO THE PHYSICIAN WITH THINNING SKIN OVER THE IPG. PHYSICIAN SURGICALLY REPLACED THE DEVICE AND RESOLVED THE PATIENT¿S ISSUE.

    Patient problems: Erosion

    MAUDE report 11366080 ↗
  • InjurySep 21, 2020Migration or Expulsion of Device; Migration or Expulsion of Device

    PATIENT PRESENTED TO PHYSICIAN WITH PAIN AND THINNING OF THE SKIN AT THE IPG SITE. SURGEON REPLACED THE IPG ON (B)(6) 2020 TO ADDRESS PATIENT''S PAIN AS WELL AS TO ADDRESS FUTURE MRI NEEDS.

    Patient problems: Pain

    MAUDE report 10563387 ↗
  • InjuryJul 10, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT EXPERIENCED PAIN AT THE IPG SITE. THE PHYSICIAN DETERMINED THAT THE IPG WAS SUPERFICIAL AND DECIDED TO REPLACE IT WITH THE SMALLER IPG (MODEL 3028) ON (B)(6) 2020.

    Patient problems: Erosion

    MAUDE report 10258382 ↗
  • InjuryDec 13, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT EXPERIENCED INTERMITTENT SENSE LEAD PAIN SYMPTOMATIC OF INTERCOSTAL NEURALGIA. THE SURGEON REMOVED THE INSPIRE SYSTEM (B)(6) 2019.

    Patient problems: Pain

    MAUDE report 9467624 ↗
  • InjurySep 30, 2019Migration; Migration

    THE PHYSICIAN FOUND VIA X-RAY THAT THE IPG HAS MIGRATED DOWNWARD. THE PHYSICIAN PERFORMED REVISION SURGERY ON (B)(6) 2019 TO RE-SECURE THE IPG WITH SUTURES. ISSUE IS NOW RESOLVED WITHOUT COMPLICATIONS.

    Patient problems: Erosion

    MAUDE report 9135549 ↗
  • InjuryJul 11, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ANTIBIOTICS (DICLOXACILLIN) PRESCRIBED TO TREAT SWELLING AT THE SITE OF THE INCISION.

    Patient problems: Cellulitis

    MAUDE report 8782611 ↗
  • InjuryJul 9, 2019Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration

    DURING A REVISION SURGERY TO RESTORE THERAPY, THE SURGEON FOUND EXTENSIVE DAMAGE TO BOTH LEADS DUE TO TWIDDLER'S SYNDROME. THE STIMULATION LEAD WAS FOUND FRACTURED IN THE RIGHT SUBLINGUAL REGION NEAR THE SUBMANDIBULAR GLAND. THE REMAINDER OF THE STIMULATION LEAD RETRACTED INTO THE CHEST WALL AND COILED IN THE AREA OF THE BATTERY PACK. THE SENSING LEAD WAS ALSO DAMAGED.

    Patient problems: Twiddlers Syndrome

    MAUDE report 8773740 ↗
  • InjuryJul 9, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT DEVELOPED VERY MILD ERYTHEMA OVER THE IPG SITE NOTED ON THE POST-OP VISIT AND WAS STARTED ON AN ANTIBIOTIC. IT COMPLETELY RESOLVED, UNEVENTFULLY ON FOLLOW UP VISIT. DATE OF EVENT UNKNOWN.

    Patient problems: Erythema

    MAUDE report 8772068 ↗
  • InjuryJun 21, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DATE OF EVENT START IS UNKNOWN; ASSOCIATED STUDY VISIT (B)(6) 2017. PATIENT WAS ADMITTED TO THE HOSPITAL WITHIN ONE MONTH OF IMPLANT FOR THE TREATMENT OF NECK SWELLING, EPISODES OF DIZZINESS, SWEATING, AND SEEING SPOTS. ADVERSE EVENT RESOLVED BY (B)(6) 2017.

    Patient problems: Dizziness

    MAUDE report 8723049 ↗
  • InjuryJun 20, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT COMPLAINED OF A LISP WITH NO NERVE WEAKNESS 3 MONTHS AFTER IMPLANT. THE PHYSICIAN REPORTED THAT THIS RESOLVED WITHOUT INTERVENTION.

    Patient problems: Dysphasia

    MAUDE report 8717365 ↗
  • InjuryJun 12, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT COMPLAINED OF LISPING, RIGHT LOWER LIP NUMBNESS, AND A CROOKED SMILE AT HIS POST-OP, 1-MONTH, AND 2-MONTH APPOINTMENTS. THE PHYSICIAN CONFIRMED THAT THE ISSUE IS NOW RESOLVED.

    Patient problems: Dysphasia; Facial Nerve Paralysis

    MAUDE report 8693651 ↗
  • InjuryJun 12, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CHEST SEROMA. THE PATIENT HAD BEEN DOING SOME HEAVY LIFTING AFTER IMPLANT AND THE SEROMA RESULTED. THE SURGEON, DR. (B)(6), EVACUATED THE SEROMA IN OFFICE AND SENT THE PATIENT HOME. THE PHYSICIAN CONFIRMED THAT THE SEROMA WAS RESOLVED.

    Patient problems: Seroma

    MAUDE report 8693322 ↗
  • InjuryJun 12, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RIGHT CHEST SEROMA AT 5 WEEKS POST-OP. ISSUE REPORTED AS RESOLVED BY THE PHYSICIAN.

    Patient problems: Seroma

    MAUDE report 8693296 ↗
  • InjuryJun 11, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SLURRED SPEECH WAS REPORTED AT THE POST-TITRATION VISIT AND FINAL VISIT (1 YEAR APART). THE PHYSICIAN CONFIRMS THAT SLURRED SPEECH ISSUES HAVE RESOLVED.

    Patient problems: Dysphasia

    MAUDE report 8689388 ↗
  • InjuryJun 11, 2019Migration; Migration

    THE ANCHOR OF THE STIMULATION LEAD WAS NOTED TO HAVE MIGRATED. THE SURGEON REPOSITIONED THE ANCHOR IN A REVISION SURGERY WHICH RESOLVED THE ISSUE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8689312 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P130008. One submission often covers many models, so check the product and lot details in each record.

  • Z-2516-2026 · May 11, 2026 · Open, Classified
    Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire Stimulation Lead (Model 4063) is design
  • Z-2515-2026 · May 11, 2026 · Open, Classified
    Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire Respirato
  • Z-2969-2018 · Jun 12, 2018 · Terminated
    Inspire II Model 3024 Implantable Pulse Generator (IPG) for Obstructive Sleep Apnea, Inspire Part #900-003-020, Product Usage: The device is an implanted component of the Inspir

Other Stimulator, Hypoglossal Nerve, Implanted, Apnea devices

All Stimulator, Hypoglossal Nerve, Implanted, Apnea devices →