WEB Aneurysm Embolization System W2-8-4Recalls & reported problems
Sequent Medical, Inc. · WEB 27 SL 8x4 · Class 3 · Implant
All models of WEB Aneurysm Embolization System (12)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00856956005817), received between Jul 22, 2020 and Aug 8, 2022 — classified as 1 death, 2 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 other recall cover devices cleared under the same FDA submission (P170032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Difficult or Delayed Separation1
- Difficult to Remove1
- Separation Failure1
- Unintended Movement1
Patient problems
- Eye Pain1
- Hemorrhage/Blood Loss/Bleeding1
- Intracranial Hemorrhage1
- No Consequences Or Impact To Patient1
- Rupture1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 8, 2022Unintended Movement; Difficult or Delayed Separation; Unintended Movement; Difficult or Delayed Separation
IT WAS REPORTED THAT THE WEB DEVICE DID NOT FULLY DETACH ON THE FIRST DETACHMENT. THE VIA CATHETER WAS ADVANCED SLOWLY, AND WEB DETACHED. THE CASE WAS REPORTED TO BE DIFFICULT BECAUSE OF THE ANATOMY AND ANGLE OF THE WAS ANEURYSM IN WITH THE VIA 27 AND CAUSED THE WEB DEVICE TO ROTATE. AS A RESULT, THE ANEURYSM HAD A SMALL RUPTURE/BLEED AFTER, WHICH WAS TREATED WITH AN INFLATED A COMPETITOR BALLOON DEVICE FOR A FEW MINUTES, AND THE BLEEDING WAS THEN STOPPED. THE PATIENT WAS MONITORED OVERNIGHT AND DID NOT RE-RUPTURE. THE PATIENT WOKE UP OKAY AND INTACT, AND REPORTED TO BE SPEAKING AND RESPONDING CONVERSATIONALLY. THE PATIENT WAS HELD AT THE HOSPITAL FOR A FEW EXTRA DAYS FOR OBSERVATION BEFORE DISCHARGE.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Rupture
MAUDE report 15188175 ↗DeathMay 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A WEB DEVICE WAS IMPLANTED IN A RUPTURED ANEURYSM OF THE LEFT INTERNAL CAROTID ARTERY BIFURCATION ON (B)(6) 2020 WITH GOOD OCCLUSION OF THE ANEURYSM AND STAGNATE FLOW IN THE SAC. THE PATIENT WAS EVALUATED ONE MONTH POST-OP AND THE ANEURYSM DID NOT SHOW UP ON MRA. THE ANEURYSM WAS DETERMINED TO BE OCCLUDED. AT THREE MONTHS POST-OP, THE PATIENT DEVELOPED EYE PAIN. IMAGING REVEALED THE ANEURYSM HAD GROWN SIGNIFICANTLY, THE ANEURYSM APPEARED TO HAVE SIGNIFICANT FILLING IN THE NECK AND BASE OF THE ANEURYSM SAC, AND THE WEB WAS COMPACTED INTO THE DOME. THE PATIENT EXPERIENCED ACUTE SYMPTOMS WAS DIAGNOSED WITH A LARGE SUBDURAL HEMATOMA AND SUBARACHNOID HEMORRHAGE. THE PATIENT LATER PASSED AWAY.
Patient problems: Eye Pain; Intracranial Hemorrhage
MAUDE report 11853150 ↗InjuryJul 22, 2020Difficult to Remove; Separation Failure; Detachment of Device or Device Component; Difficult to Remove; Separation Failure; Detachment of Device or Device Component
IT WAS REPORTED THAT TREATMENT WAS PERFORMED FOR A MIDDLE CEREBRAL ARTERY (MCA) BIFURCATION ANEURYSM. AFTER PLACEMENT OF THE WEB IN THE ANEURYSM, THE WEB DID NOT DETACH AFTER TWO (2) ATTEMPTS. THE WEB APPEARED TO HAVE DETACHED WITH THE THIRD ATTEMPT; HOWEVER, WHEN THE PUSHER WIRE AND MICROCATHETER WERE REMOVED, THE WEB WAS PULLED OUT OF THE ANEURYSM AND THEN DETACHED IN THE INTERNAL CAROTID ARTERY (ICA). THE WEB WAS REMOVED FROM THE ICA WITH A STENT RETRIEVER. THERE WAS NO REPORTED PATIENT INJURY. THE PATIENT WAS REPORTED TO HAVE NO DEFICIENCIES THE DAY AFTER THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10312043 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P170032 on Dec 31, 2018. Submitted by MicroVention, Inc. as “Woven EndoBridge (WEB) Aneurysm Embolization System”. The FDA review took 459 days. The FDA has since approved 23 supplements (changes to the device or its labeling), most recently on Nov 13, 2026. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
FDA definition of this device type: An intrasaccular flow disruption device is intended for use to treat wide-neck intracranial aneurysms in the neurovasculature through an endovascular approach.
Recalls under the same FDA submission
These recalls involve devices cleared under P170032. One submission often covers many models, so check the product and lot details in each record.
- Z-0170-2023 · Sep 21, 2022 · Open, ClassifiedWEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.
Common questions about the WEB Aneurysm Embolization System W2-8-4
Has the WEB Aneurysm Embolization System W2-8-4 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00856956005817), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the WEB Aneurysm Embolization System W2-8-4?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jul 22, 2020 and Aug 8, 2022. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (1), “Difficult or Delayed Separation” (1) and “Difficult to Remove” (1); the most common patient problems are “Eye Pain” (1), “Hemorrhage/Blood Loss/Bleeding” (1) and “Intracranial Hemorrhage” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the WEB Aneurysm Embolization System W2-8-4?
1 report is classified as a death, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the WEB Aneurysm Embolization System W2-8-4 FDA approved?
It was approved through premarket approval (PMA) P170032 on Dec 31, 2018.
Is the WEB Aneurysm Embolization System W2-8-4 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the WEB Aneurysm Embolization System W2-8-4, and is it still sold?
The FDA's device database lists Sequent Medical, Inc. as the company responsible for it (brand WEB Aneurysm Embolization System, model W2-8-4). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Intrasaccular Flow Disruption Device devices
- WEB Detachment Controller WDC-2-MVIMicrovention, Inc. · Intrasaccular Flow Disruption Device1 recall0 reports
- WEB W5-4-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death23 reports
- WEB 17 SL 6x2 W5-6-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death21 reports
- WEB W5-5-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device20 reports
- WEB W5-4.5-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device19 reports
- WEB W5-7-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death17 reports
- WEB W5-6-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death15 reports
- WEB Aneurysm Embolization System W2-8-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device14 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page