WEB Aneurysm Embolization System W2-9-4Recalls & reported problems
Sequent Medical, Inc. · WEB 27 SL 9x4 · Class 3 · Implant
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00856956005985), received between Aug 30, 2019 and Aug 30, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Difficult or Delayed Positioning” (1 report).
1 other recall cover devices cleared under the same FDA submission (P170032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Difficult or Delayed Positioning1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 30, 2019Difficult or Delayed Positioning; Difficult or Delayed Positioning
IT WAS REPORTED THAT A WEB DEVICE WAS SUCCESSFULLY DETACHED WITHIN AN ANEURYSM, BUT THE PUSHER WIRE WAS ADVANCED FORWARD, INVAGINATING THE WEB DEVICE AND PUSHING IT CLOSER TO THE DOME OF THE ANEUYSM. A SNARE WAS USED TO PULL THE WEB BACK INTO PLACE. THERE WAS NO REPORTED ADVERSE CLINICAL SEQUELA TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8952735 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P170032. One submission often covers many models, so check the product and lot details in each record.
- Z-0170-2023 · Sep 21, 2022 · Open, ClassifiedWEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.
Other Intrasaccular Flow Disruption Device devices
- WEB Detachment Controller WDC-2-MVIMicrovention, Inc. · Intrasaccular Flow Disruption Device1 recall0 reports
- WEB W5-4-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death23 reports
- WEB 17 SL 6x2 W5-6-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death20 reports
- WEB W5-5-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device20 reports
- WEB W5-4.5-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device19 reports
- WEB W5-7-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death16 reports
- WEB W5-6-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death15 reports
- WEB Aneurysm Embolization System W2-8-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device14 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.