Skip to content

Benesta Tissue Removal Device CAL-TR1511Recalls & reported problems

Caldera Medical, Inc. · The Benesta Tissue Removal Device is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retained products of conception. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00857492008096), received between Jun 20, 2023 and Sep 11, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Material Separation2
  • Dull, Blunt1
  • Failure to Cut1
  • Material Deformation1
  • Material Twisted/Bent1
  • Mechanical Problem1
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2026Product Quality Problem; Dull, Blunt; Failure to Cut

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50735532 ↗
  • MalfunctionNov 10, 2023Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component

    THE OUTER SHEATH CAME OFF PRIOR TO INSERTING THE TRD INTO THE HYSTEROSCOPE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18116130 ↗
  • MalfunctionJun 20, 2023Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    CALDERA REP NOTICED THE TRDS CUTTING WINDOW APPEARED "OPEN" EVEN THOUGH THE ACTIVATION BUTTON WAS NOT PRESSED. AS THE SURGEON ATTEMPTED TO GUIDE THE SCOPE AND TRD INTO OPTIMAL POSITION, THE OUTER TRD SHEATH BEGAN SLIDING OFF. SECOND TRD WAS USED AND PROCEDURE WAS SUCCESSFULLY COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17167439 ↗
  • MalfunctionJun 20, 2023Material Separation; Material Deformation; Material Twisted/Bent; Material Separation; Material Deformation; Material Twisted/Bent

    TRD OUTER CUTTER DETACHED FROM DEVICE DURING THE PROCEDURE. SECOND TRD WAS USED AND PROCEDURE WAS COMPLETED. CALDERA REP NOTICED THE OUTER CUTTER WAS BENT DURING THE PROCEDURE, CAUSED BY TORQUING THE DEVICE WHILE INSIDE THE WORKING CHANNEL OF THE HYSTEROSCOPE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17167295 ↗

Other Hysteroscope (And Accessories) devices

All Hysteroscope (And Accessories) devices →