iMRI 3T S OR-DR, 116118-000Recalls & reported problems
IMRIS Imaging, Inc. · IMRIS SSK OR-DR Partial Body Imaging, 116118-000 · Class 2
All models of iMRI 3T S (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00857534006059), received between Feb 2, 2022 and Nov 11, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device Handling Problem” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K133692); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 30, 2025Class IOpen, Classified
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Z-0162-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Handling Problem3
- Unintended Collision2
- Device Unsafe to Use in Environment1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Bruise/Contusion1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 11, 2025Device Handling Problem; Device Handling Problem
THE PATIENT WAS TRANSPORTED INTO THE MRI DIAGNOSTIC ROOM (ZONE 4) BY A HOSPITAL EMPLOYEE VIA WHEELCHAIR. THE PATIENT WAS IN A SEATED POSITION WHEN THE SEAT AND BACK SUPPORT MATERIAL BEGAN TO FOLD/COLLAPSE DUE TO THE FERROUS PROPERTIES OF THE METAL SUPPORT FRAME AS THE WHEELCHAIR WAS PULLED (ATTRACTED) TO THE MAGNET'S BORE. THE MAGNET WAS INTENTIONALLY QUENCHED BY ONE OF THE MRI TECHNOLOGIST OUT OF CONCERN FOR PATIENT SAFETY. IT WAS REPORTED THE LEFT SIDE OF THE PATIENT'S BODY MADE CONTACT WITH THE BORE. PER INFORMATION VERBALLY PROVIDED TO THE IMRIS CUSTOMER SERVICE ENGINEER, X-RAYS OF THE PATIENT WERE OBTAINED AND REPORTED AS NEGATIVE (NO SPECIFIC ANATOMIC AREA OF INTEREST WAS PROVIDED), MINOR BRUSING OF PATIENT WAS REPORTED; HOWEVER, NO SERIOUS INJURY WAS REPORTED TO THE MRI TECHNOLOGIST OR OTHERS PRESENT WHEN THE EVENT OCCURRED.
Patient problems: Bruise/Contusion
MAUDE report 23528797 ↗MalfunctionMay 21, 2024Mechanical Jam; Mechanical Jam
DURING PMCF SURVEY COLLECTION THE CUSTOMER IDENTIFIED AN EARLIER EVENT WHERE THE MRI SCANNER WAS NOT ABLE TO BE MOVED FROM THE DIAGNOSTIC ROOM TO THE OPERATING ROOM TO OBTAIN INTRA-OPERATIVE DIAGNOSTIC IMAGES. THE MAGNET MOVER PRODUCED AN ERROR MESSAGE AND DISALLOWED LINEAR MOVEMENT INTO THE OPERATING ROOM. THE PATIENT WAS TRANSFERRED TO ANOTHER STATIONARY MR SCANNER AT THE RADIOLOGY DEPARTMENT WHILE STILL UNDER ANESTHESIA. THE CUSTOMER REPORTED THE PROCEDURE COULD BE COMPLETED ALTHOUGH WITH SIGNIFICANT DELAY GREATER THAN 60 MINUTES WHILE THE PATIENT WAS UNDER ANESTHESIA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19367856 ↗MalfunctionFeb 3, 2022Unintended Collision; Device Handling Problem; Unintended Collision; Device Handling Problem
WHILE MOVING THE MAGNET FROM THE OPERATING ROOM TO THE DIAGNOSTIC ROOM ON (B)(6) 2020, A STAFF MEMBER BROUGHT IN AN MR-UNSAFE LINEN CART AND IT WAS DRAWN INTO THE MAGNET. THE MAGNET WAS PHYSICALLY DAMAGED BY THE LINEN CART, BREAKING THE FRONT FUNNEL COVER, THE COLLISION DETECTION RING, AND SOME ELECTRICAL CONNECTIONS INSIDE THE BODY COIL. THERE WAS NO PATIENT INVOLVEMENT AND THE USER DID NOT SUSTAIN ANY INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13446585 ↗MalfunctionFeb 2, 2022Device Unsafe to Use in Environment; Device Unsafe to Use in Environment
ON (B)(6) 2019, AN IMRIS PROGRAM MANAGER WAS INSIDE THE NEWLY CONSTRUCTED AND INSTALLED OR SUITE WITH TWO PHOTOGRAPHERS. A FERROUS OBJECT WAS DRAWN FROM THE HAND OF ONE OF THE PHOTOGRAPHERS INTO THE BORE OF THE MAGNET. THERE WAS NO PATIENT INVOVLEMENT OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13425087 ↗MalfunctionFeb 2, 2022Unintended Collision; Device Handling Problem; Unintended Collision; Device Handling Problem
ON (B)(6) 2019, IT WAS REPORTED THAT A CLEANING PERSONNEL WAS IN THE MAGNETIC RESONANCE SUITE WHILE THE MAGNET WAS DEPLOYED. AN MR-UNSAFE TROLLEY WAS DRAWN INTO THE MAGNET, DAMAGING THE FUNNEL COVER AND COLLISION DETECTION RING. THERE WAS NO PATIENT INVOLVEMENT AND THE USER DID NOT SUSTAIN ANY INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13424965 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133692. One submission often covers many models, so check the product and lot details in each record.
- Z-0164-2026 · Sep 30, 2025 · Open, ClassifiedIMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida
- Z-0163-2026 · Sep 30, 2025 · Open, ClassifiedIMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA
- Z-0161-2026 · Sep 30, 2025 · Open, ClassifiedIMRIS Operating Suite, Neuro III-SV iMRI System, iMRX System, containing Siemens 3 Tesla MRI Magnet, Verio
- Z-1645-2018 · Apr 4, 2018 · TerminatedIMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS intraoperative MRI systems are traditional MRI systems that have been suspended on an overhead rail system. and is designa
- Z-1098-2016 · Feb 10, 2016 · TerminatedIntra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a magnetic resonance diagnostic device (MRDD).
Other System, Nuclear Magnetic Resonance Imaging devices
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging14 recalls12 reports
- Ingenia Elition SPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging13 recalls2 reports
- NA Discovery MR750w 3.0TGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging13 recalls0 reports
- Ingenia Ingenia Ambition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging12 recalls12 reports
- Ingenia Ingenia 3.0T CXPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls59 reports
- Ingenia Smartpath to dStream for 3.0TPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls44 reports
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls5 reports
- MR 7000 Series MR 7700Philips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page