Skip to content

Monaco (UDI 00858164002152)Recalls & reported problems

Elekta, Inc. · Class 2

All models of Monaco (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00858164002152), received between Dec 22, 2016 and Mar 27, 2020 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Computer Software Problem8

Patient problems

  • Insufficient Information2
  • No Consequences Or Impact To Patient2
  • No Patient Involvement2
  • Radiation Overdose2

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 27, 2020Computer Software Problem; Computer Software Problem

    DURING AN INTERNAL INVESTIGATION, WHEN EDITING A CONTOUR AND CLICKING ON A CORONAL OR SAGITTAL SLICE, THE 3D VOLUME IS RECREATED UNNECESSARILY WHICH COULD RESULT IN A VOLUME CHANGE.

    Patient problems: No Patient Involvement

    MAUDE report 9888942 ↗
  • MalfunctionJan 12, 2018Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED INCORRECT MONITOR UNITS FOR MONACO 3D PLAN. BASED ON THE AVAILABLE INFORMATION THERE HAVE BEEN NO ACTUAL MISTREATMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7185203 ↗
  • MalfunctionJan 8, 2018Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT THEY HAVE TWO PATIENTS PRESENTING RADIATION DAMAGE FROM ONE AND HALF YEARS AGO. BASED UPON THE AVAILABLE INFORMATION THE PATIENT RECEIVED A DOSE OF RADIATION THAT HAS LED TO SERIOUS INJURY.

    Patient problems: Radiation Overdose

    MAUDE report 7171667 ↗
  • MalfunctionJan 8, 2018Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT THEY HAVE TWO PATIENTS PRESENTING RADIATION DAMAGE FROM ONE AND HALF YEARS AGO. BASED UPON THE AVAILABLE INFORMATION THE PATIENT RECEIVED A DOSE OF RADIATION THAT HAS LED TO SERIOUS INJURY.

    Patient problems: Radiation Overdose

    MAUDE report 7171513 ↗
  • MalfunctionNov 22, 2017Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT MONACO WILL DISPLAY THE INCORRECT MU FOR A 3D PLAN. BASED ON THE AVAILABLE INFORMATION THERE HAS BEEN NO MISTREATMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7052006 ↗
  • MalfunctionOct 11, 2017Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED AN ERROR TO THE DOSE CALCULATION OF A PATIENT TREATED IN (B)(6) 2016. THE ERROR WAS DISCOVERED BY THE SITE DURING ROUTINE QA PROCEDURES.

    Patient problems: Insufficient Information

    MAUDE report 6939523 ↗
  • MalfunctionOct 11, 2017Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT A PATIENT WAS MISTREATED WITH VERSION (B)(4) BEFORE THE UPGRADE TO 5.10.04. IN (B)(6).

    Patient problems: Insufficient Information

    MAUDE report 6939520 ↗
  • MalfunctionDec 22, 2016Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT A PATIENT HAS RECEIVED THE FIRST FRACTION OF THE TREATMENT BEFORE REALISING THAT THE WRONG DOSE WAS CALCULATED IN (B)(6). BASED ON THE AVAILABLE INFORMATION THERE HAS BEEN AN ACTUAL MISTREATMENT, HOWEVER THE SEVERITY DID NOT LEAD TO SERIOUS CLINICAL OUTCOME.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6198322 ↗

FDA clearance and classification

The FDA regulates product code MUJ (system,planning,radiation therapy treatment) under 21 CFR 892.5050 as a Class II device. 21 CFR 892.5050 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Monaco

Has the Monaco been recalled?

No FDA recall record in our data names this exact device identifier (DI 00858164002152), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Monaco?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Dec 22, 2016 and Mar 27, 2020. In those reports, the most common specific device problem is “Computer Software Problem” (8); the most common patient problem is “Radiation Overdose” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Monaco?

None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Monaco, and is it still sold?

The FDA's device database lists Elekta, Inc. as the company responsible for it (brand Monaco, model Monaco). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System,Planning,Radiation Therapy Treatment devices

All System,Planning,Radiation Therapy Treatment devices →