Visual-ICE MRI Cryoablation Console FPRCH7000Recalls & reported problems
Galil Medical Inc. · MRI Cryoablation System Console · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00859019006318), received between May 21, 2022 and Aug 7, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Insufficient Cooling” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Cooling3
- Defective Device1
- Gas/Air Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 7, 2024Defective Device; Defective Device
IT WAS REPORTED THAT A MASSIVE ICEBALL WAS CREATED FASTER THAN EXPECTED AND THE PROCEDURE HAD TO BE CANCELLED. THIS VI MRI CRYOABLATION CONSOLE WAS SELECTED FOR MRI CRYOABLATION OF A DESMOID TUMOR. DURING THE PROCEDURE, ALL SIXTEEN NEEDLES WERE TESTED AND PASSED; HOWEVER, THE STICK FUNCTION ON THE CONSOLE WAS NOT WORKING. THEY PHYSICIAN OPTED TO TURNED THE 10% FREEZE OPTION ON TO GET A SIMILAR STICK EFFECT, BUT UPON SCANNING THE PATIENT THEY OBSERVED A MASSIVE SIZE ICE BALL (8CM IN 10 MINUTES). THIS IS NOT NORMAL AND APPEARS THE MACHINE WAS NOT PRODUCING THE CORRECT ICE AMOUNT. THE THAW FUNCTION WAS QUICKLY TURNED ON TO PREVENT ICE FROM GROWING MORE AND THE PROCEDURE WAS CANCELLED (POST-SEDATION). THE PATIENT WAS MONITORED CLOSE AND THERE HAVE BEEN NO PATIENT COMPLICATIONS OR ADDITIONAL TREATMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19933415 ↗MalfunctionNov 29, 2022Insufficient Cooling; Gas/Air Leak; Insufficient Cooling; Gas/Air Leak
IT WAS REPORTED THAT THIS UNIT WAS GENERATING SUBOPTIMAL ICE. A VI MRI CRYOABLATION CONSOLE WAS USED IN A CRYOABLATION PROCEDURE. THE CONSOLE DID NOT GENERATE SUFFICIENT ICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15880504 ↗MalfunctionJun 22, 2022Insufficient Cooling; Insufficient Cooling
IT WAS REPORTED THAT INSUFFICIENT ICE OCCURRED. A CRYO CONSOLE AND AN ICEROD 90 DEGREE MRI NEEDLE WERE SELECTED FOR TREATMENT OF A KIDNEY INJURY OF 25X21. DURING THE FIRST FREEZE AT 10 MINUTES, AN ICEBALL OF 44X17 WAS NOTICED. THE PHYSICIAN EXPECTED TO SEE AN ICEBALL OF 45X26. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. DURING THE MAGNETIC RESONANCE IMAGING (MRI) EXAM, FOUR MONTHS AFTER THE PROCEDURE, THE ICEBALL SIZE WAS DEEMED NOT SUFFICIENT IN COVERING THE TARGETED TISSUE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14780721 ↗MalfunctionMay 21, 2022Insufficient Cooling; Insufficient Cooling
IT WAS REPORTED THAT THERE WAS INSUFFICIENT ICE. A VISUAL ICE MRI CRYOABLATION CONSOLE WAS SELECTED FOR A CRYOABLATION PROCEDURE. IT WAS REPORTED THAT THERE WERE ISSUES WITH GETTING THE RIGHT SIZE ICEBALLS. THE DEVICE SEEMED TO BE WORKING AT 50% CAPACITY. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14466260 ↗
Other Unit, Cryosurgical, Accessories devices
- IceFORCE® 2.1 CX 90° Cryoablation Needle FPRPR3602Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls4 deaths274 reports
- IceRod® 1.5 CX 90° Cryoablation Needle FPRPR3533Boston Scientific Corporation · Unit, Cryosurgical, Accessories1 recall168 reports
- IcePearl® 2.1 CX 90° Cryoablation Needle FPRPR3601Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls78 reports
- IceFORCE™ 2.1 CX L 90° Cryoablation Needle FPRPR3618Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls6 reports
- IceFORCE® 2.1 CX Cryoablation Needle FPRPR3604Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls3 reports
- IcePearl™ 2.1 CX L 90° Cryoablation Needle FPRPR3617Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls1 report
- IceRod® He-Thaw Prostate Cryoablation Kit / Presice® System FPRPR4570Galil Medical Ltd · Unit, Cryosurgical, Accessories2 recalls0 reports
- IceRod® i-Thaw® Prostate Cryoablation Kit / Presice® System FPRPR4070Galil Medical Ltd · Unit, Cryosurgical, Accessories2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page