Visual-ICE MRI Cryoablation Console FPRCH7000Recalls & reported problems

Galil Medical Inc. · MRI Cryoablation System Console · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00859019006318), received between May 21, 2022 and Aug 7, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Insufficient Cooling” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Cooling3
  • Defective Device1
  • Gas/Air Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2024Defective Device; Defective Device

    IT WAS REPORTED THAT A MASSIVE ICEBALL WAS CREATED FASTER THAN EXPECTED AND THE PROCEDURE HAD TO BE CANCELLED. THIS VI MRI CRYOABLATION CONSOLE WAS SELECTED FOR MRI CRYOABLATION OF A DESMOID TUMOR. DURING THE PROCEDURE, ALL SIXTEEN NEEDLES WERE TESTED AND PASSED; HOWEVER, THE STICK FUNCTION ON THE CONSOLE WAS NOT WORKING. THEY PHYSICIAN OPTED TO TURNED THE 10% FREEZE OPTION ON TO GET A SIMILAR STICK EFFECT, BUT UPON SCANNING THE PATIENT THEY OBSERVED A MASSIVE SIZE ICE BALL (8CM IN 10 MINUTES). THIS IS NOT NORMAL AND APPEARS THE MACHINE WAS NOT PRODUCING THE CORRECT ICE AMOUNT. THE THAW FUNCTION WAS QUICKLY TURNED ON TO PREVENT ICE FROM GROWING MORE AND THE PROCEDURE WAS CANCELLED (POST-SEDATION). THE PATIENT WAS MONITORED CLOSE AND THERE HAVE BEEN NO PATIENT COMPLICATIONS OR ADDITIONAL TREATMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19933415 ↗
  • MalfunctionNov 29, 2022Insufficient Cooling; Gas/Air Leak; Insufficient Cooling; Gas/Air Leak

    IT WAS REPORTED THAT THIS UNIT WAS GENERATING SUBOPTIMAL ICE. A VI MRI CRYOABLATION CONSOLE WAS USED IN A CRYOABLATION PROCEDURE. THE CONSOLE DID NOT GENERATE SUFFICIENT ICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15880504 ↗
  • MalfunctionJun 22, 2022Insufficient Cooling; Insufficient Cooling

    IT WAS REPORTED THAT INSUFFICIENT ICE OCCURRED. A CRYO CONSOLE AND AN ICEROD 90 DEGREE MRI NEEDLE WERE SELECTED FOR TREATMENT OF A KIDNEY INJURY OF 25X21. DURING THE FIRST FREEZE AT 10 MINUTES, AN ICEBALL OF 44X17 WAS NOTICED. THE PHYSICIAN EXPECTED TO SEE AN ICEBALL OF 45X26. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. DURING THE MAGNETIC RESONANCE IMAGING (MRI) EXAM, FOUR MONTHS AFTER THE PROCEDURE, THE ICEBALL SIZE WAS DEEMED NOT SUFFICIENT IN COVERING THE TARGETED TISSUE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14780721 ↗
  • MalfunctionMay 21, 2022Insufficient Cooling; Insufficient Cooling

    IT WAS REPORTED THAT THERE WAS INSUFFICIENT ICE. A VISUAL ICE MRI CRYOABLATION CONSOLE WAS SELECTED FOR A CRYOABLATION PROCEDURE. IT WAS REPORTED THAT THERE WERE ISSUES WITH GETTING THE RIGHT SIZE ICEBALLS. THE DEVICE SEEMED TO BE WORKING AT 50% CAPACITY. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14466260 ↗

Other Unit, Cryosurgical, Accessories devices

All Unit, Cryosurgical, Accessories devices →