Skip to content

Tricera™ Controller R-10001Recalls & reported problems

Relign Corporation · System Controller with Footswitch · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00859273007014), received between Feb 8, 2023 and Apr 29, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Frayed” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Frayed3
  • Connection Problem1
  • Device Difficult to Program or Calibrate1
  • Key or Button Unresponsive/not Working1
  • Loose or Intermittent Connection1
  • Pressure Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Material Frayed; Loose or Intermittent Connection

    IT WAS REPORTED THAT OUTSIDE OF SURGERY, THE UNIT HAD AN ISSUE WITH THE FOOTWSWITCH, THERE WAS A WIRE FRAYED. THE UNIT HAD WIRES EXPOSED/FRAYED AND A DANGLING STRAIN RELIEF THERE WAS NO PATIENT INVOLVEMENT. DUE DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25029013 ↗
  • MalfunctionNov 4, 2024Material Frayed; Key or Button Unresponsive/not Working; Material Frayed; Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT DURING SET UP, THE PEDALS OR BUTTONS ON FOOTSWITCH NO WORKED, CHANGED OVER FOOTSWITCH AND ALL OK. ON FURTHER INSPECTION THERE WERE SOME EXPOSED WIRES ON THE CABLE, METAL AND COLORED WIRES WERE VISIBLE. THERE WAS NO PATIENT HARM/INJURY OR SURGICAL DELAY AS THERE WAS NO PATIENT INVOLVEMENT. DUE DILIGENCE IS COMPLETE, NO FURTHER INFORMATION IS AVAILABLE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20599192 ↗
  • MalfunctionJun 25, 2024Material Frayed; Device Difficult to Program or Calibrate; Pressure Problem; Material Frayed; Device Difficult to Program or Calibrate; Pressure Problem

    NO ADDITIONAL EVENT INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19604932 ↗
  • MalfunctionFeb 8, 2023Connection Problem; Connection Problem

    IT IS REPORTED THAT THE FOOT PEDAL DOES NOT GET RECOGNIZED WHEN IT IS PLUGGED IN. NO HARM OR DELAY WAS REPORTED. AT PRODUCT EVALUATION/INVESTIGATION THERE WAS CORROSION ON THE TIP OF SEVERAL PINS, SALINE CORROSION ON ALL WIRES AND A DISCONNECTED CENTER WIRE. EXPOSED WIRES WERE PRESENT. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION. NO ADDITIONAL EVENT INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16329693 ↗