Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00860000122375), received between Nov 19, 2021 and Nov 19, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Capsular Bag Tear1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 19, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    A (B)(6) PATIENT UNDERWENT CATARACT SURGERY IN THE LEFT EYE ON (B)(6) 2021 WHERE THE MILOOP AND MICOR LENS FRAGMENTATION SYSTEM (EXTRACTOR AND DRIVE) WERE USED TO FRAGMENT AND REMOVE THE CATARACTOUS LENS. THE PATIENT'S POSTERIOR CAPSULE TORE DURING SURGERY. THERE WAS NO LOSS OF VITREOUS FLUID, NO VITRECTOMY, AND NO CHANGE TO THE INTRAOCULAR LENS IMPLANTED. THE SURGEON REPORTED THE MICOR "FELT RUNNING AT TOO HIGH THROTTLE" AND HE "GOT TOO CLOSE TO THE [CAPSULAR] BAG". THE SURGICAL VIDEO REVEALED THE FOLLOWING ADDITIONAL INFORMATION AND INSIGHT. IT SHOULD BE NOTED THAT THE VIDEO RESOLUTION WAS LOW AND IT WAS NOT POSSIBLE TO IDENTIFY EXACTLY WHEN THE CAPSULAR TEAR OCCURRED. THE PATIENT HAD A DARK IRIS AND A SMALL PUPIL WHICH REQUIRED USE OF AN IRIS RETRACTOR AT THE BEGINNING OF SURGERY. THE MILOOP WAS USED TO BISECT THE PATIENT'S GRADE 3 CATARACTOUS LENS. THERE WERE NO PROBLEMS WITH THE TECHNIQUE USED TO EXECUTE THE MILOOP PROCEDURE. THERE WAS NO EVIDENCE OF A DEVICE MALFUNCTION WITH THE MILOOP OR MICOR DEVICES AND THERE WAS NO EVIDENCE OF AN OVERSPEED CONDITION. BASED ON THE VIDEO EVIDENCE, IT IS POSSIBLE THAT THE IRIS RETRACTOR AND SURGICAL TECHNIQUE DURING ITS REMOVAL AND/OR BAG PO

    Patient problems: Capsular Bag Tear

    MAUDE report 12845129 ↗

Other Unit, Phacofragmentation devices

All Unit, Phacofragmentation devices →