Cubby Beds CUB2Recalls & reported problems
Sensory Medical, Inc. · Cubby Enclosed Bed Canopy System · Class 1
Quick summary
This Cubby Enclosed Bed canopy system has 133 reports, most describing safety sheet zippers that break or tear and let children get out of the bed. The reports mention 1 death and 7 injuries, including scratches, bruises and a child becoming temporarily trapped. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 133
- Deaths reported
- 1
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 133 adverse event reports that list this exact device (DI 00860007550249), received between Dec 17, 2022 and Jun 30, 2026 — classified as 1 death, 7 injuries, 125 malfunctions.
The most frequently reported specific device problem is “Material Separation” (85 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation85
- Break30
- Material Split, Cut or Torn19
- Fail-Safe Did Not Operate9
- Component Missing3
- Device Dislodged or Dislocated3
- Fail-Safe Problem3
- Insufficient Device Problem Information3
Patient problems
- No Clinical Signs, Symptoms or Conditions94
- Physical Entrapment18
- Abrasion10
- Insufficient Information10
- Bruise/Contusion5
- Laceration(s)4
- Bone Fracture(s)3
- Fall2
Recent adverse event reports
The 25 most recent of 133 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 30, 2026Break
A COMPLAINT WAS REPORTED TO SENSORY MEDICAL BY A DURABLE MEDICAL EQUIPMENT (DME) REPRESENTATIVE. PER REPRESENTATIVE, THE CHILD BROKE THE SAFETY SHEET ZIPPER AND ELOPED THROUGH IT. PER REPRESENTATIVE, THE ZIPPERS ARE DAMAGED ON BOTH SHEETS. PER THE REPRESENTATIVE, THE PARENTS STATED THAT THE CHILD PULLED ON THE ZIPPERS AND BROKE THEM, AND STATED THE PADLOCKS WERE NOT BEING USED TO SECURE THE SAFETY SHEET. FIVE (5) PICTURES WERE PROVIDED BY THE REPRESENTATIVE. PHOTO 1: SAFETY SHEET ZIP-TIED TO THE DOOR ZIPPER. PHOTO 2: SAFETY SHEET ZIP-TIED TO THE DOOR ZIPPER, THE ZIPPER TAB ON THE SAFETY SHEET ZIPPER IS MISSING. PHOTO 3: SAFETY SHEET ZIP-TIED TO THE DOOR ZIPPER, THE ZIPPER TAB ON THE SAFETY SHEET ZIPPER IS MISSING. PHOTO 4: SAFETY SHEET ZIP-TIED TO THE DOOR ZIPPER, THE ZIPPER TAB ON THE SAFETY SHEET ZIPPER IS MISSING. PHOTO 5: SAFETY SHEET ZIP-TIED TO THE DOOR ZIPPER, THE ZIPPER TAB ON THE SAFETY SHEET ZIPPER IS MISSING. ADDITIONALLY, THE REPRESENTATIVE STATED THAT THE PARENTS CONFIRMED THE SAFETY PADLOCKS WERE NOT BEING USED. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25672888 ↗MalfunctionJun 30, 2026Break
THE PARENT REPORTED THAT THEIR CHILD BROKE THE ZIPPER AROUND THE TECHNOLOGY HUB PORTAL AREA AND ELOPED FROM THE BED. PER THE PARENT, THE CHILD INITIALLY BROKE THE ZIPPER WHEN THE TECHNOLOGY HUB WAS INSTALLED AND ELOPED FROM THE BED. THE PARENT REMOVED THE TECHNOLOGY HUB AND INSTALLED THE CLOTH COVER BECAUSE THEY DETERMINED THE CHILD WAS "TOO DESTRUCTIVE" AND WOULD CONTINUE TO PUSH OUT THE TECHNOLOGY HUB. THE PARENT REPORTED THAT THE CHILD CONTINUOUSLY BREAKS THE ZIPPER OFF IT'S TRACK AND REMOVES THE CLOTH COVER AND ELOPE FROM THE BED. ONE (1) PICTURE WAS PROVIDED THAT SHOWS THE TECHNOLOGY HUB PORTAL WITH NO HUB OR COVER INSTALLED, AND THE PADLOCK IS STILL LOCKED TO THE ZIPPER TAB AND LOCKING LOOP. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25672840 ↗MalfunctionJun 25, 2026Material Split, Cut or Torn
THE PARENT REPORTED THAT HER SON RIPPED THE SAFETY SHEET, CRAWLED UNDER THE BED AND BECAME TEMPORARILY ENTRAPPED. PER PARENT, SHE NOTICED A SMALL HOLE IN THE SAFETY SHEET THE DAY PRIOR TO THE EVENT. SHE STATED THAT HER SON WAS ABLE TO RIP THE SHEET AT THE LOCATION OF THE HOLE. THE PARENT STATED THAT HER SON SUSTAINED A SMALL SCRATCH ON HIS LEG AS A RESULT OF THE EVENT.
Patient problems: Abrasion; Physical Entrapment
MAUDE report 25630031 ↗MalfunctionJun 25, 2026Material Separation; Material Split, Cut or Torn
THE PARENT REPORTED THAT HER SON UNZIPPED THE SAFETY SHEET ZIPPER AND BECAME TEMPORARILY ENTRAPPED IN THE BED FRAME. SHE EXPLAINED THAT HER SON FIGURED OUT HOW TO UNDO THE S-CLIP THAT WAS SECURING THE SAFETY SHEET ZIPPER AND HE CRAWLED UNDER THE SHEET AND BECAME STUCK IN THE FRAME. THE PARENT WAS ABLE TO REMOVE THE CHILD FROM THE BED. SHE STATED THAT HER CHILD SUSTAINED A SMALL BRUISE ON HIS CHEST AND DID NOT REQUIRE MEDICAL ATTENTION OR INTERVENTION. THE PARENT STATED SHE DID NOT RECEIVE PADLOCKS WITH THE BED AND ONLY RECEIVED THE S-CLIPS. ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER CONFIRMS THAT S-CLIPS WERE BEING USED TO SECURE THE SAFETY SHEET ZIPPER, AND A TEAR IN THE SEAM OF THE MESH DOOR. THE PARENT PROVIDED TWO RELEVANT PHOTOS. ONE PHOTO SHOWS A TEAR IN THE SEAM OF THE MESH DOOR AND CANOPY WALL. ONE PHOTO SHOWS THE SAFETY SHEET SECURED WITH AN S-CLIP. SEPARATELY, THE PARENT REPORTED THAT THERE IS A SMALL TEAR IN THE CANOPY FABRIC BETWEEN THE MESH DOOR AND THE CANOPY WALL.
Patient problems: Bruise/Contusion; Physical Entrapment
MAUDE report 25629557 ↗MalfunctionJun 17, 2026Material Separation
THE PARENT REPORTED THAT HER CHILD SEPARATED THE SAFETY SHEET ZIPPER AND ELOPED THROUGH THE MATTRESS AREA. IN ADDITION, THE PARENT REPORTED THAT HER CHILD UNZIPPED THE SAFETY SHEET AND THE CLOTH COVER OF THE TECHNOLOGY HUB PORTAL AND ELOPED FROM THE BED. THE PARENT CONFIRMED THAT SHE WAS NOT USING THE NECESSARY LOCKS BECAUSE SHE WAS NOT PROVIDED THEM AT THE INSTALLATION OF THE BED BY THE PROVIDER. THE PARENT ALSO REPORTED THAT THE BED SLATS UNDER THE MATTRESS ARE BENT OR BROKEN DUE TO THE CHILD JUMPING AROUND INSIDE THE BED. SHE EXPLAINED THAT SHE NOTICED THE DAMAGE A COUPLE MONTHS PRIOR BUT DID NOT REPORT THE ISSUE, UNTIL SHE WAS HAVING ISSUES WITH THE SAFETY SHEET AND CLOTH COVER ZIPPERS. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25538177 ↗MalfunctionJun 12, 2026Break; Material Separation
THE PARENT REPORTED THAT TWO OF THEIR SAFETY SHEETS HAVE BROKEN DUE TO NORMAL USE BY THEIR CHILD WHO HAS "STRONG PULLING" BEHAVIORS. FOR ONE, THE ZIPPER PULL WAS TORN FREE OF THE FABRIC IT WAS ANCHORED TO. ON THE OTHER, THE ZIPPER DETACHED FROM THE BOTTOM HALF THAT ALLOWS THE ZIPPER TEETH TO ENGAGE AND CLICK INTO PLACE AND IS THEREBY UNUSABLE. THE PARENT STATED THAT THEIR CHILD WAS ABLE TO ELOPE FROM THE BED DUE TO THE BROKEN SAFETY SHEET ZIPPERS. THE PARENT STATED THAT THE TWO SAFETY SHEET ZIPPERS BROKE WITHIN A MONTH OR TWO OF EACH OTHER. THE PARENT CONFIRMED THE PADLOCKS WERE BEING USED AT THE TIME OF THE EVENT. ONE PHOTO WAS PROVIDED TO SENSORY MEDICAL AND IT SHOWS THE ZIPPER SLIDER CAME OFF THE TRACK / TEETH TAPE SIDE. THERE WAS NO PHOTO PROVIDED OF THE REPORT OF THE ZIPPER STOP COMING OFF THE SAFETY SHEET. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25490576 ↗MalfunctionJun 12, 2026Break
PER PARENT, THE CHILD BROKE OFF THE TECH HUB PORTAL ZIPPER SLIDER OFF OF THE CANOPY AND ELOPED THROUGH THE PORTAL WINDOW. PER PARENT, THERE WAS NO DAMAGE TO THE TECH HUB PORTAL ZIPPER PRIOR TO THE ELOPEMENT EVENT. PER PARENT, THEY HAD CONTINUED USE OF THE BED AFTER THE DAMAGE, AND THE CHILD HAS ELOPED MULTIPLE TIMES THROUGH THE TECH HUB PORTAL. ONE (1) PICTURE WAS PROVIDED THAT SHOWS AN EMPTY TECH HUB PORTAL WINDOW WITH THE ZIPPER SLIDER REMOVED. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25490160 ↗MalfunctionJun 12, 2026Break
A COMPLAINT WAS REPORTED TO SENSORY MEDICAL BY A DURABLE MEDICAL EQUIPMENT (DME) REPRESENTATIVE. PER REPRESENTATIVE, THE CHILD BROKE THE SAFETY SHEET ZIPPER AND ELOPED THROUGH IT. PER REPRESENTATIVE, THE PARENTS STATED THAT THE CHILD PULLED ON THE ZIPPER AND BROKE IT. TWO (2) PICTURES WERE PROVIDED BY THE REPRESENTATIVE. BOTH IMAGES SHOW THE SIDE OF THE CUBBY BED WITHOUT A SAFETY SHEET INSTALLED. NO IMAGES WERE PROVIDED THAT SHOW THE DAMAGE TO THE SAFETY SHEET. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25488202 ↗MalfunctionJun 12, 2026Break; Material Separation
PER CAREGIVER, THE CHILD BROKE THE SAFETY SHEET ZIPPER AND THE PADLOCK DURING THE NIGHT LEADING TO THE CHILD ELOPING OUT FROM UNDER THE BED. PER CAREGIVER, THE CHILD REMOVED THE SAFETY SHEET ZIPPER TAB. PER CAREGIVER, BASED ON THEIR VISUAL ASSESSMENT, THE CHILD CHEWED ON THE TEETH OF THE SAFETY SHEET AS WELL AS THE PADLOCK. PER CAREGIVER, THE CHILD RECEIVED SCRAPES FROM THE ELOPEMENT WHICH DID NOT REQUIRE MEDICAL ATTENTION.
Patient problems: Abrasion
MAUDE report 25488161 ↗MalfunctionJun 8, 2026Material Split, Cut or Torn
PER PARENT, THE CHILD TORE THE SAFETY SHEET FABRIC AND ESCAPED THROUGH THE TEAR. PER PARENT, THE DAMAGE OCCURRED IN THE CENTER OF THE SHEET, AND THE ELOPEMENT ONLY OCCURRED ONCE. PER PARENT, THEY DID NOT NOTICE THE TEAR UNTIL AFTER THE ELOPEMENT. ONE (1) PICTURE WAS PROVIDED OF THE SAFETY SHEET. IN THE PICTURE, THERE ARE TWO SEWN AREAS ALONG THE LENGTH OF THE SAFETY SHEET, SHOWING WHERE THE TEARS WERE SEWN TOGETHER BY THE PARENT(S). PER PARENT, THEY REPAIRED (SEWED) THE TEAR TO PREVENT FURTHER ELOPEMENT. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25430225 ↗MalfunctionJun 8, 2026Break; Device Dislodged or Dislocated
THE PARENT REPORTED THAT HER SON BROKE THE TECHNOLOGY HUB ZIPPER AREA AND WAS ABLE TO ELOPE FROM THE BED THROUGH THE PORTAL AREA. SHE STATED THAT THE CHILD WAS ABLE TO BREAK THE S-CLIPS (WHICH ONE WAS BEING USED TO SECURE THE ZIPPER, AS WELL AS THE PADLOCK) FROM INSIDE OF THE BED. SHE STATED THAT SHE DOES NOT KNOW HOW HE WAS ABLE TO BREAK THE ZIPPER OR S-CLIPS FROM INSIDE THE BED. A PHOTO PROVIDED BY THE PARENT SHOWS THE TECHNOLOGY HUB ZIPPER SECURED WITH THE LOCK AND AN S-CLIP, SECURED THROUGH A PUNCHED HOLE IN THE LOCKING LOOP, WHICH WAS TORN/BROKE BY THE CHILD. IT APPEARS THAT THE S-CLIP WAS BEING USED IN LIEU OF THE BROKEN/TORN LOCKING LOOP, AND IT IS CONFIRMED THROUGH THE PHOTO THAT THE S-CLIP IS BROKEN. THERE WAS NO INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25429624 ↗MalfunctionJun 2, 2026Break
THE COMPLAINT WAS REPORTED TO SENSORY MEDICAL BY A DURABLE MEDICAL EQUIPMENT (DME) REPRESENTATIVE. PER REPRESENTATIVE, THE TECH HUB PORTAL ZIPPER TEETH ON THE CANOPY AND TECH HUB BROKE DURING REGULAR USE. PER REPRESENTATIVE, THE CHILD ELOPED DUE TO THE DAMAGES. PER REPRESENTATIVE, THE TECH HUB WAS INSTALLED AT THE TIME THAT THE ZIPPER BROKE, AND THERE WAS NO PREVIOUSLY ACKNOWLEDGED DAMAGE PRIOR TO THE ELOPEMENT. THREE (3) PHOTOS WERE PROVIDED THAT SHOW THE DAMAGE TO THE TECHNOLOGY HUB PORTAL ZIPPER. TWO (2) PHOTOS SHOW DAMAGE TO THE TECH HUB PORTAL ZIPPER ON THE CANOPY SIDE. ONE (1) PHOTO SHOWS DAMAGE TO THE TECH HUB PORTAL ZIPPER ON THE TECH HUB SIDE. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25363510 ↗MalfunctionMay 29, 2026Material Separation
THE PARENT REPORTED THAT THE ZIPPER THAT ATTACHES THE SHEET TO THE FABRIC ENCLOSURE KEEPS SPLITTING OPEN. THE CHILD IS BECAME ENTRAPPED BETWEEN THE MATTRESS AND THE FRAME. THE PARENT STATED THAT HER DAUGHTER WEDGED HERSELF INSIDE THE CUBBY BED BETWEEN THE MATTRESS AND THE CANOPY WALL, WITH THE SAFETY SHEET IN PLACE, CAUSING THE SAFETY SHEET ZIPPER TO SEPARATE. THE CHILD WAS STUCK IN THAT LOCATION UNTIL THE PARENTS FOUND HER. THE ZIPPER DID NOT APPEAR TO BE DAMAGED, PER THE PARENT, AND WAS ZIPPED AND LOCKED AS REQUIRED PRIOR TO THE REPORTED INCIDENT. THE PARENT STATED THAT THIS ISSUE HAPPENED TWICE, 3-4 DAYS APART. THE CUSTOMER PROVIDED ONE PHOTO FOR REVIEW BY SENSORY MEDICAL. THE PHOTO DISPLAYS THAT THE SAFETY SHEET ZIPPER IS SEPARATED. THERE WAS NO INJURY AS A RESULT OF THIS EVENT.
Patient problems: Physical Entrapment
MAUDE report 25340506 ↗MalfunctionMay 13, 2026Break; Device Dislodged or Dislocated
PER PARENT, THE CHILD WAS PUSHING ON THE TECHNOLOGY HUB RESULTING IN THE ZIPPER TAB FOR THE TECHNOLOGY HUB PORTAL BREAKING. PER PARENT, THE CHILD ELOPED THROUGH THE TECHNOLOGY HUB PORTAL AFTER IT WAS PUSHED THROUGH AND THERE WAS NO DAMAGE PRESENT BEFORE THE EVENT OCCURRED. PER PARENT, THE TECHNOLOGY HUB WAS SECURED WITH A PADLOCK DURING THE EVENT. FOUR (4) PHOTOS WERE PROVIDED THAT SHOW THE DAMAGE TO THE TECHNOLOGY HUB PORTAL ZIPPER. TWO (2) OF THE IMAGES SHOW AN EMPTY TECHNOLOGY HUB PORTAL WINDOW, WITH THE DAMAGED ZIPPER TAB ATTACHED TO THE CANOPY LOCKING LOOP BY A PADLOCK. THE OTHER TWO (2) PHOTOS SHOW THE DAMAGED TECHNOLOGY HUB PORTAL ZIPPER SLIDER MISSING THE TAB AND ATTACHED TO THE TECHNOLOGY HUB. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25164921 ↗MalfunctionMay 5, 2026Material Separation
THE PARENT REPORTED THAT HER SON WAS STIMMING IN THE BED AND PULLED ON THE SHEET, CAUSING THE ZIPPER TO TEAR OR SEPARATE AT THE SEAM. HER SON WAS THEN ABLE TO ELOPE FROM THE BED AT THE AREA WHERE THE SAFETY SHEET ZIPPER SEPARATED ALONG THE SEAM. THE PARENT CONFIRMED THE LOCK WAS IN PLACE AT THE TIME OF THE EVENT. THE PARENT PROVIDED THREE PHOTOS UPON REQUEST FROM SENSORY MEDICAL. TWO PHOTOS SHOW THE SAFETY SHEET ZIPPER TORN ALONG THE SHEET SEAM. ONE PHOTO SHOWS THE TORN ZIPPER IN THE BED. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25084656 ↗MalfunctionMay 5, 2026Material Separation
A COMPLAINT WAS REPORTED TO SENSORY MEDICAL BY A DURABLE MEDICAL EQUIPMENT (DME) REPRESENTATIVE. PER REPRESENTATIVE, THE CHILD BROKE THE SAFETY SHEET ZIPPER AND ELOPED THROUGH IT. PER REPRESENTATIVE, THE MOM SAID THE LOCK WAS INSTALLED AT THE TIME OF THE EVENT AND THE CHILD PULLED HARD ENOUGH TO BREAK THE ZIPPER WITH THE LOCK STILL ATTACHED. A PICTURE PROVIDED SHOWS THE UNINSTALLED SAFETY SHEET WITH A MISSING ZIPPER SLIDER, AND ANOTHER SHEET WITH A DAMAGED ZIPPER WITH FRAYING THREADS. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25084494 ↗MalfunctionMay 1, 2026Material Split, Cut or Torn
PER PARENT, THE CANOPY SEPARATED AT THE SEAM WHERE THE ZIPPER HOLDS THE CANOPY TO THE SAFETY SHEET. PER PARENT, HER DAUGHTER ELOPED THROUGH THE DAMAGED SEAM AND OUT THROUGH THE BOTTOM OF THE BED. PER PARENT, SHE DID NOT NOTICE THE SEAM SEPARATION NEAR THE ZIPPER UNTIL AFTER HER DAUGHTER ESCAPED. ONE (1) VIDEO WAS PROVIDED SHOWING THE DAMAGE TO THE CANOPY. THE VIDEO SHOWED A TEAR IN THE SEAM BETWEEN THE CANOPY AND THE CANOPY SIDE SAFETY SHEET ZIPPER APPROXIMATELY 10 INCHES LONG. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25060805 ↗MalfunctionApr 24, 2026Material Separation; Device Dislodged or Dislocated
THE PARENT REPORTED THAT HER CHILD WAS ABLE TO UNZIP THE TECHNOLOGY HUB AND ELOPE FROM THE BED THROUGH THE TECHNOLOGY HUB PORTAL AREA. SHE CONFIRMED THE TECHNOLOGY HUB WAS ZIPPED AND LOCKED AT THE TIME OF THE EVENT. THE PARENT CONFIRMED THAT THE CHILD WAS ABLE TO UNZIP THE TECHNOLOGY HUB, WHILE IT WAS LOCKED IN PLACE, AND CLIMB OUT OF THE BED. THE PARENT STATED THAT SHE DID NOT NOTICE ANY DAMAGE TO THE TECHNOLOGY HUB AS A RESULT OF THE EVENT. SHE ALSO STATED THAT SHE IS UNABLE TO PROVIDE ANY PHOTOS OF THE REPORTED DAMAGE. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24990160 ↗MalfunctionApr 22, 2026Material Split, Cut or Torn
PER THE PARENT, HER SON WAS KICKING THE CUBBY BED BY THE SAFETY SHEET ZIPPER AND IT RIPPED, THEN THE CHILD FELL THROUGH THE RIPPED SHEET AND BECAME STUCK UNDERNEATH THE BED. THE MOTHER INDICATED THE CHILD WAS STUCK FOR APPROXIMATELY ONE MINUTE BEFORE THE PARENT WAS ABLE TO REMOVE HIM FROM THE BED. THE PARENT CONFIRMED THAT SHE WAS NOT USING THE LOCK ON THE SAFETY SHEET AT THE TIME OF THE EVENT. THE RIPPED SHEET WAS DISCARDED BY THE PARENT. THERE WERE NO PHOTOS OF THE RIPPED SHEET PROVIDED FOR REVIEW. THE PARENT PROVIDED A PHOTO OF A SECOND SHEET, NOT INVOLVED IN THE EVENT REPORTED, SHOWING THE ZIPPER PULL TAB IS MISSING. THE PARENT EXPLAINED THE ZIPPER ON THIS SECOND SHEET IS FUNCTIONAL BUT BROKEN. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; Physical Entrapment
MAUDE report 24960391 ↗InjuryApr 22, 2026Break
PER PARENT, THEIR SON BROKE THE SAFETY SHEET ZIPPER TEETH, PREVENTING THE SHEET FROM BEING INSTALLED. PER PARENT, THEY CONTINUED USE OF THE DAMAGED SAFETY SHEET WITHOUT THE SHEET BEING FULLY ASSEMBLED, THEN THE CHILD BECAME ENTRAPPED BETWEEN THE MATTRESS AND THE FRAME. PER PARENT, THE ENTRAPMENT LED TO THE DISLOCATION OF THE CHILD'S KNEE, WHICH RESULTED IN A VISIT TO THE EMERGENCY ROOM. PER PARENT, AT THE ER, AN HCP FIXED THE CHILD'S KNEE AND SENT THEM HOME. PER PARENT, THE CHILD IS DOING "OKAY" SINCE THE VISIT. PER PARENT, A SIMILAR EVENT HAS OCCURRED 3 TIMES, WITH TWO REQUIRING TRIPS TO THE ER.
Patient problems: Joint Dislocation; Physical Entrapment
MAUDE report 24959906 ↗MalfunctionApr 13, 2026Material Split, Cut or Torn
PER PARENT, THE CHILD TORE THE SHEET FABRIC AND ESCAPED THROUGH THE TEAR. PER PARENT, THE DAMAGE OCCURRED NEAR THE ZIPPER AND THE ELOPEMENT ONLY OCCURRED ONCE. PER PARENT, THEY BELIEVE THE DAMAGE OCCURRED DUE TO THE AGE OF THE SHEET AND THE CONSTANT WASHING, AND SHE DID NOT NOTICE THE TEAR UNTIL AFTER THE ELOPEMENT. ONE (1) PICTURE WAS PROVIDED OF THE SAFETY SHEET TEAR. IN THE PICTURE, THERE IS A TEAR ALONG THE ZIPPER COVERED IN TAPE. PER PARENT, SHE ATTEMPTED TO TAPE SHUT THE TEAR TO PREVENT FURTHER ELOPEMENT. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24866835 ↗MalfunctionMar 30, 2026Break
THE COMPLAINANT IS A DURABLE MEDICAL EQUIPMENT (DME) REPRESENTATIVE. PER THE DME REPRESENTATIVE, THE CANOPY-SIDE SAFETY SHEET ZIPPER HAD BROKEN AND THE CHILD ELOPED FROM THE BED. PER THE REPRESENTATIVE, THE BROKEN ZIPPER WAS NOTICED AFTER THE CHILD ELOPED FROM THE BED. THE REPRESENTATIVE CONFIRMED THE LOCKS WERE BEING USED AT THE TIME OF THE EVENT. THE DME REPRESENTATIVE PROVIDED TWO (2) PHOTOS THAT APPEAR TO CONFIRM THE DAMAGE TO THE CANOPY SIDE SAFETY SHEET ZIPPER. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24721763 ↗MalfunctionMar 30, 2026Break; Material Separation
THE PARENT REPORTED THAT THE SAFETY SHEET ZIPPER SEPARATED AND HER CHILD ELOPED FROM THE BED BETWEEN THE MATTRESS AND THE CANOPY WALL. THE PARENT INSPECTED THE INSIDE OF THE BED AFTER THE CHILD ELOPED AND DISCOVERED THAT THE ZIPPER HAD SEPARATED. AFTER ELOPEMENT, SHE ATTEMPTED TO UNZIP THE ZIPPER, IT GOT STUCK AND THE ZIPPER PULL TAB BROKE OFF. THE PARENT STATED THE ISSUE HAPPENED APPROXIMATELY 1 ½ WEEKS PRIOR TO HER REPORTING THE EVENT. SHE CONFIRMED THE LOCKS WERE BEING USED AT THE TIME OF THE EVENT. THE PARENT PROVIDED ONE PHOTO OF THE SAFETY SHEET. REVIEW OF THE PHOTO CANNOT CONFIRM THE REPORTED DAMAGE TO THE SHEET. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24721246 ↗MalfunctionMar 18, 2026Material Separation; Material Split, Cut or Torn
THE COMPLAINT WAS REPORTED TO SENSORY MEDICAL BY A DURABLE MEDICAL EQUIPMENT (DME) REPRESENTATIVE. PER THE REPRESENTATIVE, THE CHILD RIPPED ONE OF THE SHEETS ON THE LONG SIDE NEAR WHERE THE ZIPPER STARTS AND ENDS THEN BECAME ENTRAPPED BETWEEN THE MATTRESS AND THE FRAME DURING AN ATTEMPT TO ESCAPE. PER THE REPRESENTATIVE, THE CHILD WAS ABLE TO PULL ON A LOOSE THREAD AND CAUSE THE RIP. PER THE REPRESENTATIVE, THEY DO NOT THINK IT WAS BECAUSE OF A MANUFACTURING DEFECT, AND THE PARENT WAS AWARE OF THE THREAD BEFORE ELOPEMENT OCCURRED. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: Physical Entrapment
MAUDE report 24630520 ↗MalfunctionMar 17, 2026Material Separation
PER PARENT, ONE OF THEIR SAFETY SHEETS SEPARATES FROM THE CANOPY SIDE SAFETY SHEET ZIPPER DURING USE, RESULTING IN THEIR CHILD ELOPING. PER PARENT, IT ONLY HAPPENED WITH ONE SAFETY SHEET, AND THE DAMAGE WAS NOTICED AFTER THE CHILD ELOPED. PER PARENT, THEY HAVE DISCONTINUED USE OF THAT SAFETY SHEET, AND A PADLOCK WAS IN USE AT THE TIME OF ELOPEMENT. THERE WAS NO REPORT OF INJURY AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24618337 ↗
Other Patient Bed With Canopy/Restraints devices
- POSEY A8070TFPosey Products, LLC · Patient Bed With Canopy/Restraints10 reports
- Posey 8060Tidi Products, LLC · Patient Bed With Canopy/Restraints4 reports