POSEY A8070TFRecalls & reported problems
Posey Products, LLC · 8070 SUB-ASSEMBLY · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10190676000996), received between Dec 26, 2024 and Jul 20, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem6
- Insufficient Device Problem Information2
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 20, 2026Mechanical Problem
A PATIENT WAS ABLE TO DISMANTLE PIECES OF THE BED FROM INSIDE THE ZIPPED CLOSED CANOPY. THE PATIENT WAS ABLE TO USE THEIR FINGERS TO TEAR THE ZIPPER TRACT OPEN.\THE PATIENT DID NOT GET OUT, BUT WAS REPEATEDLY MANEUVERING HIS FINGERS TO BREAK THE ZIPPER APART.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25882938 ↗MalfunctionJan 28, 2026Material Puncture/Hole
PATIENT ESCAPED CANOPY THROUGH HOLE IN NETTING. NO INJURIES OF PATIENT OR STAFF WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24205555 ↗MalfunctionJul 1, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
CUSTOMER REPORTED AN INCIDENT WITH THE POSEY BED WHERE A TODDLER WAS FOUND ON THE FLOOR OUTSIDE OF THE BED ENCLOSURE DUE TO AN UNSECURED ZIPPER. THE TODDLER OR ANY STAFF DID NOT SUFFER ANY INJURIES IN THE RELATED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22378766 ↗MalfunctionJun 23, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
CUSTOMER STATES THAT THEIR CUSTOMER HAD A PATIENT WHO FELL OUT OF THE ENCLOSED BED. THE PATIENT WAS ABLE TO PENETRATE THE ZIPPER TRACT NEAR AT BOTTOM PART OF PANEL WITH ALL CLIPS/BUCKLES ENGAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22310018 ↗MalfunctionMar 17, 2025Mechanical Problem; Mechanical Problem
8070 S/N (B)(6) HAS ZIPPER DAMAGE, RIGHT SIDE PANEL HAS ZIPPER TEETH SPLIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21620418 ↗MalfunctionMar 10, 2025Mechanical Problem; Mechanical Problem
S/N (B)(6) HAS ZIPPER DAMAGE, MATTRESS ENVELOPE ZIPPER IS SPLITTING ON THE TRACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21566372 ↗MalfunctionMar 10, 2025Mechanical Problem; Mechanical Problem
SUPPLEMENTAL MEDWATCH SUBMITTED FOR ADDITIONAL INFORMATION RELATED TO A PREVIOUSLY SUBMITTED REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21560474 ↗MalfunctionMar 6, 2025Mechanical Problem; Mechanical Problem
S/N (B)(6) HAS ZIPPER DAMAGE, ZIPPER SPLIT ON RIGHT SIDE PANEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21536619 ↗MalfunctionDec 26, 2024Mechanical Problem; Mechanical Problem
SN (B)(6) ZIPPER IS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21019295 ↗
FDA clearance and classification
The FDA regulates product code OYS (patient bed with canopy/restraints) under 21 CFR 880.6760 as a Class I device. 21 CFR 880.6760 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Enclosed bed canopy system used as passive restraint.
Common questions about the POSEY A8070TF
Has the POSEY A8070TF been recalled?
No FDA recall record in our data names this exact device identifier (DI 10190676000996), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the POSEY A8070TF?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Dec 26, 2024 and Jul 20, 2026. In those reports, the most common specific device problems are “Mechanical Problem” (6) and “Material Puncture/Hole” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the POSEY A8070TF?
None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the POSEY A8070TF FDA approved?
Devices of this type (product code OYS) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Posey Products, LLC.
Who makes the POSEY A8070TF, and is it still sold?
The FDA's device database lists Posey Products, LLC as the company responsible for it (brand POSEY, model A8070TF). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Patient Bed With Canopy/Restraints devices
- Cubby Beds CUB2Sensory Medical, Inc. · Patient Bed With Canopy/Restraints1 death133 reports
- Posey 8060Tidi Products, LLC · Patient Bed With Canopy/Restraints4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page