Skip to content

INSPIRATION 7i Ventilator INSPIRATION 7i Ventilator 17" screenRecalls & reported problems

Event Medical, Ltd. · INSPIRATION 7i Ventilator System, 17" screen · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00860008873637), received between Aug 17, 2023 and Aug 17, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “No Display/Image” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • No Display/Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 17, 2023No Display/Image; No Display/Image

    THE FOLLOWING WAS REPORTED TO EVENT MEDICAL: IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE SCREEN OF THE VENTILATOR WENT BLACK. THE EQUIPMENT WAS VENTILATING AND WORKING ALL THE TIME, WITHOUT STOPPING VENTILATION,THE PATIENT WAS PLACED ON AN ALTERNATE VENTILATOR WITH NO INJURY REPORTED. THE THIRD PARTY SERVICE PERSONNEL REPLACED THE DEVICE'S DISPLAY WHICH CORRECTED THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17567904 ↗

Other Ventilator, Continuous, Facility Use devices

All Ventilator, Continuous, Facility Use devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page