CoNextions TR Tendon Repair System-Combo Kit FA0001Recalls & reported problems

Conextions Medical Inc · The CoNextions TR Tendon Repair System consists of a single-use, sterile implant consisting of two identical stainless steel anchors implanted in either end of the injured tendon and connected by two loops of UHMWPE fiber. The implant is provided pre-loaded into an Implant Mechanism and with a Deployment Mechanism to facilitate placement. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00861445000303), received between Sep 1, 2023 and Sep 27, 2024 — classified as 2 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Rupture2
  • Muscle/Tendon Damage1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT SUFFERED A LACERATION OF THE FLEXOR POLLICIS LONGUS (FPL) TENDON OF THE RIGHT THUMB. THE TENDON WAS SURGICALLY REPAIRED USING A TWO STRAND SUTURE REPAIR AND A CONEXTIONS TR TENDON REPAIR IMPLANT. APPROXIMATELY A WEEK AFTER THE PROCEDURE, THE PATIENT REPORTED AN INABILITY TO MOVE THE THUMB. THE SURGEON SURGICALLY EXPLORED THE AREA AND FOUND THE REPAIR HAD RUPTURED WITH THE CONEXTIONS TR IMPLANT HAVING PULLLED OUT OF THE DISTAL SIDE OF THE REPAIR. THE CONEXTIONS TR IMPLANT WAS FULLY REMOVED AND THE TENDON WAS REAPPROXIMATED USING SUTURE.

    Patient problems: Rupture

    MAUDE report 20321146 ↗
  • InjurySep 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT SUFFERED A LACERATION TO THE EXTENSOR POLLICIS LONGUS (EPL) TENDON OF THE LEFT THUMB. THE TENDON WAS SURGICALLY REPAIRED WITH A CONEXTOINS TR TENDON REPAIR IMPLANT. 14 DAYS POST-OPERATIVE THE PATIENT WAS MOVING HEAVY FURNITURE, FELT A "POP" AND WAS NO LONGER ABLE TO EXTEND THEIR THUMB. THE SURGEON SURGICALLY EXPLORED THE AREA AND FOUND THE REPAIR HAD RUPTURED. THE CONEXTIONS TR TENDON REPAIR IMPLANT WAS FULLY REMOVED AND THE TENDON, AND A EIP TO EPL TENDON TRANSFER WAS PERFORMED TO ADDRESS THE ISSUE WITH SUTURE USED TO APPROXIMATE THE TWO TENDONS TOGETHER.

    Patient problems: Muscle/Tendon Damage; Rupture

    MAUDE report 17671543 ↗