DiLumen Endolumenal Interventional Platform, 130 cm DiLumen, 130 cmRecalls & reported problems
Lumendi LLC · DiLumen consists of a soft flexible sheath that fits over standard colonoscopes. The device employs two balloons, one behind the bending section of the colonoscope and the second in front of the tip of the colonoscope. When both balloons are deployed, and inflated the area in between is stabilized. This Therapeutic Zone (TZ) facilitates stabilization, insufflation and manipulation of the tissue thus allowing the clinician to effectively treat the target area. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00865309000225), received between Jul 6, 2021 and Aug 1, 2022 — classified as 4 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Mechanical Problem4
Patient problems
- Bowel Perforation4
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryAug 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
FOLLOWING A UNDERWATER ENDOLUMINAL PROCEDURE AND UPON REMOVAL OF THE ENDOSCOPE WITH DILUMEN ATTACHED, A 3CM PERFORATION APPROXIMATELY 30-35 CM FROM THE ANAL VERGE WAS NOTED REQUIRING SURGICAL INTERVENTION TO REPAIR. THE DOCTOR DID NOT ALLEGE A DEVICE MALFUNCTION..
Patient problems: Bowel Perforation
MAUDE report 15139785 ↗MalfunctionMay 8, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT HAD A TORTUOUS COLON THAT WAS DIFFICULT TO NAVIGATE. DURING NAVIGATION THE DOCTOR PUSHED THE DEVICE TOO DESPITE FEELING RESISTANCE. ONE PUSH ROD SEPARATED FROM THE FORE BALLOON SKIRT.. NO ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14332336 ↗InjuryJan 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING AN ENDOLUMINAL PROCEDURE A PERFORATION OF THE SIGMOID COLON OCCURRED REQUIRING SURGICAL INTERVENTION TO REPAIR. THE CUSTOMER DID NOT ALLEGE THAT THE DEVICE MALFUNCTIONED.
Patient problems: Bowel Perforation
MAUDE report 13154428 ↗InjuryDec 20, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NOTE: THIS EVENT OCCURED IN (B)(6) 2021 BUT WAS NOT REPORTED TO LUMENDI UNTIL 11/22/2021. WHEN THE NEW SALES MANAGER VISITED THE ACCOUNT, HE WAS INFORMED A PATIENT WAS READMITTED IN LATE (B)(6) 2021 WITH ABDOMINAL PAIN SHORTLY AFTER ESD WITH THE DEVICE AND WAS SUBSEQUENTLY FOUND TO HAVE A PERFORATION ON THE LEFT SIDE OF THE COLON THAT REQUIRED SURGICAL INTERVENTION. THE USER REPORTED AT THE TIME OF THE ENDOSCOPIC INTERVENTION HE HAD DIFFICULTY NAVIGATING THE ENDOSCOPE THROUGH THE SIGMOID COLON AS THE PATIENT HAD PREVIOUS PELVIC SURGERY.
Patient problems: Bowel Perforation
MAUDE report 13046353 ↗MalfunctionOct 11, 2021Mechanical Problem; Mechanical Problem
DURING POLYPECTOMY, DOCTOR NOTICED MIGRATION OF DEVICE FORE BALLOON AND THEN NOTICED THAT THE FORE BALLOON WAS COMPLETELY SEPARATED FROM DEVICE WITH BOTH PUSH RODS BROKEN. DOCTOR REMOVED DILUMEN DEVICE AND THEN SEPARATELY WENT BACK IN WITH A SNARE TO REMOVE FB AND PUSH ROD SEGMENTS. POST-PROCEDURE, X RAY WAS DONE WHICH REVEALED NO PERFORATION OF THE COLON.
Patient problems: Insufficient Information
MAUDE report 12605859 ↗MalfunctionSep 8, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT HAD A TORTUOUS COLON THAT WAS DIFFICULT TO NAVIGATE. THE ENDOSCOPE WAS POSITIONED SLIGHTLY FORWARD OF THE FORE BALLOON. THE FORE BALLOON WAS IN THE FULLY RETRACTED POSITION. UPON PULLING THE ENDOSCOPE BACKWARD THE DOCTOR NOTED THE DAMAGE TO THE PUSH RODS. ONE PUSH ROD SEPARATED FROM THE SADDLE TIP AND THE OTHER PUSH ROD BROKE AT THE FORE BALLOON SKIRT. NO ADVERSE EVENT.
Patient problems: Insufficient Information
MAUDE report 12442580 ↗MalfunctionAug 24, 2021Mechanical Problem; Mechanical Problem
PATIENT HAD A 3 CM LESION IN THE ASCENDING COLON. PATIENT HAD A VERY TORTUOUS ANATOMY AND THE DOCTOR HAD DIFFICULTY NAVIGATING TO THE LESION SITE. HE WAS ABLE TO SEE THE LESION IN THE ENDOSCOPIC VIEW.. WHEN HE HAD TROUBLE GOING ANY FURTHER, HE STARTED TO MANIPULATE THE DILUMEN FORE BALLOON BY TRYING TO EXTEND IT PAST THE LESION. HE WAS UNABLE TO GET PAST THE LESION AND PULLED THE DEVICE OUT AND THAT IS WHEN WE SAW THE PUSH ROD ON THE LEFT BECAME SEPARATED. THE CASE WAS FINISHED WITH ONLY THE ADULT PENTAX COLONOSCOPE AND THE PATIENT WAS DISCHARGED THAT SAME DAY. NO ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12363020 ↗InjuryJul 6, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PERFORATION OF RECTO-SIGMOID COLON. PATIENT- (B)(6)-YEAR-OLD, FEMALE WITH A TORTUOUS COLON AND A FIXED SIGMOID COLON DUE TO PREVIOUS OOPHORECTOMY AND APPENDECTOMY. THE ENDOSCOPE AND DILUMEN ENTERED THE ANUS AND THE DOCTOR PROCEEDED TO INTRODUCE THE ENDOSCOPE/DILUMEN THROUGH THE RECTUM AND UP INTO THE SIGMOID. APPROX. 8 MINUTES LATER BLEEDING WOULD BE SEEN AT THE ANUS. AFTER REMOVING BOTH DILUMEN AND THE SCOPE, THE SCOPE WAS REINSERTED WITHOUT DILUMEN, AND A PERFORATION (APPROXIMATELY 4 CM) WAS DISCOVERED IN THE AREA OF THE RECTO-SIGMOID COLON. THE PERFORATION WAS CLOSED ENDOSCOPICALLY AND CONFIRMED UNDER FLUOROSCOPY. THERE WAS NO FURTHER SEQUELAE FROM THE INCIDENT. THE ATTENDING DOCTOR INDICATED THAT IT WAS LIKELY THAT AGGRESSIVE NAVIGATION OF THE ENDOSCOPE/DEVICE IN A NARROW AND FIXED SIGMOID PATIENT'S ANATOMY CAUSED THE PERFORATION AND WAS NOT SPECIFICALLY DEVICE RELATED.
Patient problems: Bowel Perforation
MAUDE report 12125374 ↗
Other Colonoscope And Accessories, Flexible/Rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page