EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190I (REF N3802440)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA lll COLONOVIDEOSCOPE (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EVIS EXERA III colonoscope is a flexible endoscope for examining the colon. Many reports describe contamination found on scopes and cleaning that was not adequate. The records include 1,944 reports, 12 injuries and no deaths. There is 1 recall, open since September 2023, because some repaired scopes had water left in the channel, a possible source of contamination. Reports rose from 252 in the year before to 472 in the most recent year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,967
Deaths reported
0
Injuries reported
12
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1,967 adverse event reports that list this exact device (DI 04953170305177), received between May 11, 2018 and Sep 30, 2026 — classified as 12 injuries, 1,955 malfunctions.

The most frequently reported specific device problem is “Microbial Contamination of Device” (782 reports).

1 FDA recall record name this device's identifier, 1 still open.

8 other recalls cover devices cleared under the same FDA submission (K131780); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 30 reportsDec 2023: 33 reportsJan 2024: 38 reportsFeb 2024: 35 reportsMar 2024: 31 reportsApr 2024: 52 reportsMay 2024: 32 reportsJun 2024: 35 reportsJul 2024: 30 reportsAug 2024: 30 reportsSep 2024: 36 reportsOct 2024: 43 reportsNov 2024: 21 reportsDec 2024: 6 reportsJan 2025: 16 reportsFeb 2025: 19 reportsMar 2025: 29 reportsApr 2025: 17 reportsMay 2025: 25 reportsJun 2025: 37 reportsJul 2025: 19 reportsAug 2025: 31 reportsSep 2025: 16 reportsOct 2025: 16 reportsNov 2025: 35 reportsDec 2025: 36 reportsJan 2026: 66 reportsFeb 2026: 31 reportsMar 2026: 46 reportsApr 2026: 61 reportsMay 2026: 63 reportsJun 2026: 44 reportsJul 2026: 45 reportsAug 2026: 29 reportsSep 2026: 39 reportsOct 2026: 0 reports20242025202666
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0329-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Microbial Contamination of Device782
  • Failure to Clean Adequately709
  • Device Reprocessing Problem177
  • Communication or Transmission Problem98
  • No Display/Image74
  • Contamination62
  • Poor Quality Image35
  • Contamination /Decontamination Problem32

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,764
  • No Known Impact Or Consequence To Patient134
  • No Consequences Or Impact To Patient49
  • Insufficient Information14
  • Perforation5
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bacterial Infection1
  • Caustic/Chemical Burns1

Recent adverse event reports

The 25 most recent of 1,967 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026No Display/Image

    The customer reported that no image was displayed during inspection for use involving an olympus colonovideoscope. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949574 ↗
  • MalfunctionSep 30, 2026Device Reprocessing Problem

    The customer reported the colonovideoscope had a separate issue. During the initial evaluation of the device, foreign material was observed in the Air Water channel. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948661 ↗
  • MalfunctionSep 30, 2026Microbial Contamination of Device

    It was reported that the Colonovideoscope tested positive for an unspecified amount of Pseudomonas. The issue was found during a routine culture of the scope. The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944710 ↗
  • MalfunctionSep 30, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, it was observed that the scope cover unit had foreign objects. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941996 ↗
  • MalfunctionSep 30, 2026Erratic or Intermittent Display

    The customer reported image flickering during inspection for use involving an Olympus colonovideoscope. There was no patient involvement reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941963 ↗
  • MalfunctionSep 30, 2026Communication or Transmission Problem

    The customer reported scope communication error B30 during inspection for use involving an Olympus colonovideoscope. There was no patient involvement reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941852 ↗
  • MalfunctionSep 29, 2026Erratic or Intermittent Display; Poor Quality Image

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, video flickering, vertical lines, and a pixelated image were observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936528 ↗
  • MalfunctionSep 28, 2026Microbial Contamination of Device

    It was reported that, during reprocessing, the COLONOVIDEOSCOPE tested positive for 4Colony Forming Units (CFUs) of pseudomonas aeruginosa. The device was retested on July 27, 2026, and it tested positive for 7 CFUs of unspecified bacteria. The device was tested again on August 04, 2026, and tested positive for more than >66 CFUs of unspecified bacteria. All channels were sampled. The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918426 ↗
  • MalfunctionSep 28, 2026Communication or Transmission Problem; Output Problem

    The customer reported not read by the light source during an inspection for use involving an Olympus Colonovideoscope. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916649 ↗
  • MalfunctionSep 25, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50896697 ↗
  • MalfunctionSep 24, 2026Gas/Air Leak

    The customer reported dirty instrument during the leak test during reprocessing involving a colonovideoscope. There was no patient involvement reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50888666 ↗
  • MalfunctionSep 24, 2026Device Reprocessing Problem

    The customer reported the colonovideoscope had a separate issue. During the initial evaluation of the device, it was observed that connecting tube, bending section cover, adhesives around light guide lens, objective lens, c-cover, jet tube and scope connector case unit and protector of universal cord on scope connector side had foreign object. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50886477 ↗
  • MalfunctionSep 24, 2026Device Reprocessing Problem

    The customer reported the colonovideoscope head had a separate issue. During the initial evaluation of the device it was observed, Air/Water Channel has foreign material. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885984 ↗
  • MalfunctionSep 24, 2026Erratic or Intermittent Display; No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50876586 ↗
  • MalfunctionSep 23, 2026Microbial Contamination of Device

    It was reported that, during reprocessing, the Colonovideoscope tested positive for >100 Colony Forming Units (CFUs) of Klebsiella pneumoniae. All channels were sampled. The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872360 ↗
  • MalfunctionSep 22, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50857955 ↗
  • MalfunctionSep 22, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope to the service center for a separate issue. During the device analysis, the technical assistance indicates that a foreign objects auxiliary jet channel was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50850383 ↗
  • MalfunctionSep 21, 2026Corroded

    During onsite inspection of colonovideoscope, it was found . There was no patient involvement.The customer reported the colonovideoscope had a separate issue. During the initial evaluation of the device it was observed that Connector Equalization valve cover was corroded. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50835273 ↗
  • MalfunctionSep 19, 2026Erratic or Intermittent Display; Poor Quality Image

    It was reported that the customer allegation was involving an Olympus colonovideoscope.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50832748 ↗
  • MalfunctionSep 18, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, it was observed that white foreign material was present within the jet channel. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50813483 ↗
  • MalfunctionSep 15, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50781621 ↗
  • MalfunctionSep 14, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope, which is an Olympus asset with no reported complaint. During the evaluation of the device repair technician found white foreign material jet tube, there was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50750596 ↗
  • MalfunctionSep 14, 2026Failure to Clean Adequately

    The customer reported the colonovideoscope had a separate issue. During the initial evaluation of the device, it was observed that foreign objects were in the air water tube and cylinder. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50750473 ↗
  • MalfunctionSep 11, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50740359 ↗
  • MalfunctionSep 11, 2026Microbial Contamination of Device

    Patient problems: Insufficient Information

    MAUDE report 50738683 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131780. One submission often covers many models, so check the product and lot details in each record.

  • Z-1210-2024 · Jan 11, 2024 · Open, Classified
    Colonoscope, Model Number PCF-H190DL.
  • Z-0419-2024 · Oct 16, 2023 · Open, Classified
    PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE
  • Z-0416-2024 · Oct 16, 2023 · Open, Classified
    GIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
  • Z-0417-2024 · Oct 16, 2023 · Open, Classified
    GIF-HQ190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
  • Z-0418-2024 · Oct 16, 2023 · Open, Classified
    PCF-H190DL, EVIS EXERA III COLONOVIDEOSCOPE
  • Z-0415-2024 · Oct 16, 2023 · Open, Classified
    CF-HQ190L EVIS EXERA III COLONOVIDEOSCOPE
  • Z-0341-2024 · Sep 25, 2023 · Open, Classified
    Colonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-
  • Z-0056-2025 · Sep 14, 2023 · Open, Classified
    EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light sources, monitors, EndoTherapy accessories, and other ancil

Other Colonoscope and accessories, flexible/rigid devices

All Colonoscope and accessories, flexible/rigid devices →