EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190L (REF N3802540)Recalls & reported problems
Olympus Medical Systems Corp. · "PCF-H190L EVIS EXERA III HD SLIM COLONS." · Class 2
All models of EVIS EXERA lll COLONOVIDEOSCOPE (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
Reports for this Olympus colonoscope mostly concern cleaning: residue or contamination left after reprocessing (997 reports), along with image and connection problems. Reports rose sharply, from 209 to 1,646 over the past year. Of 2,416 reports, 29 describe injuries and 2 deaths. It is named in 3 open recalls related to repairs and drying of scope channels.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,513
- Deaths reported
- 2
- Injuries reported
- 29
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 2,513 adverse event reports that list this exact device (DI 04953170305191), received between Nov 15, 2018 and Sep 30, 2026 — classified as 2 deaths, 29 injuries, 2,482 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (1,033 reports).
3 FDA recall records name this device's identifier, 3 still open.
6 other recalls cover devices cleared under the same FDA submission (K131780); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jan 11, 2024Class IIOpen, Classified
Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Z-1211-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Oct 16, 2023Class IIOpen, Classified
Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
Z-0419-2024 · Root cause: Process control · Recall details · FDA record ↗
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0341-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Clean Adequately1,033
- No Display/Image380
- Communication or Transmission Problem360
- Poor Quality Image208
- Residue After Decontamination191
- Device Reprocessing Problem186
- Microbial Contamination of Device132
- Erratic or Intermittent Display79
Patient problems
- No Clinical Signs, Symptoms or Conditions2,446
- No Known Impact Or Consequence To Patient16
- Insufficient Information10
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available7
- No Consequences Or Impact To Patient7
- Perforation7
- Bowel Perforation6
- Hemorrhage/Blood Loss/Bleeding4
Recent adverse event reports
The 25 most recent of 2,513 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Poor Quality Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954447 ↗MalfunctionSep 30, 2026Failure to Clean Adequately
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952653 ↗MalfunctionSep 30, 2026Residue After Decontamination
The customer reported the colonovideoscope had a black piece that had removed from inside the tube. During the initial evaluation of the device, foreign material in the biopsy channel was observed. There was no patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950806 ↗MalfunctionSep 30, 2026Poor Quality Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950006 ↗MalfunctionSep 30, 2026Failure to Clean Adequately
The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, foreign material was observed inside the nozzle, and white residue was observed within the air/water supply channel. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945620 ↗MalfunctionSep 30, 2026Failure to Clean Adequately
The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, Air water supply, channel and cylinder had white residue. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945504 ↗MalfunctionSep 30, 2026Failure to Clean Adequately
The customer reported that the colonovideoscope had a separate issue. During the initial evaluation of the returned device, a dirty angle wire was observed. No patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941790 ↗MalfunctionSep 30, 2026Communication or Transmission Problem
It was reported that the error message Scope Communication Error occurred during a diagnostic colonoscopy while using the colonovideoscope. The procedure was completed with a similar device. There were no reports of patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941772 ↗MalfunctionSep 29, 2026Residue After Decontamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50938889 ↗MalfunctionSep 29, 2026No Display/Image; Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50938711 ↗MalfunctionSep 29, 2026Failure to Clean Adequately
The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, white residue was observed in the auxiliary water flow, and the nozzle was clogged with white residue. No patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937208 ↗MalfunctionSep 29, 2026No Display/Image
The customer reported no image during inspection for use involving an Olympus colonovideoscope. There were no reports of patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935818 ↗MalfunctionSep 29, 2026Failure to Clean Adequately
The customer reported that the colonovideoscope instrument was dirty during reprocessing. During the initial evaluation of the device, the connecting tube was observed to be sticky. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935736 ↗MalfunctionSep 29, 2026Residue After Decontamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935701 ↗MalfunctionSep 29, 2026Erratic or Intermittent Display
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934362 ↗MalfunctionSep 29, 2026Failure to Clean Adequately
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931799 ↗MalfunctionSep 29, 2026Failure to Clean Adequately
The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, white residue was observed in the air/water channel and cylinder and in the auxiliary water channel. No patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931770 ↗MalfunctionSep 29, 2026Residue After Decontamination; Communication or Transmission Problem
The customer reported the Colonovideoscope displayed an E216 scope communication error. During the initial evaluation of the device, that white residue in the air/water channel and cylinder was observed. No patient harm was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927292 ↗MalfunctionSep 29, 2026Device Reprocessing Problem
The customer reported that the colonovideoscope had a separate issue. During the initial evaluation of the device, it was observed that the light guide lens had foreign objects. No patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927120 ↗MalfunctionSep 28, 2026Residue After Decontamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926190 ↗MalfunctionSep 28, 2026No Display/Image
The customer reported static screen without image during inspection for use involving an Olympus colonovideoscope. There were no reports of patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923241 ↗MalfunctionSep 28, 2026Erratic or Intermittent Display
The customer reported an image that was all staticky, like snow during inspection for use involving an Olympus Colonovideoscope. There was no patient involvement reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917260 ↗MalfunctionSep 28, 2026Failure to Clean Adequately
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50916018 ↗MalfunctionSep 28, 2026Failure to Clean Adequately
The customer reported the colonovideoscope had a separate issue. During the initial evaluation of the device, it was observed that foreign objects were in the air water tube and cylinder and jet tube. No patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915287 ↗MalfunctionSep 28, 2026Failure to Clean Adequately
The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, white residue was observed within the air/water channel, air/water cylinder, and auxiliary water tube. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50912619 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131780. One submission often covers many models, so check the product and lot details in each record.
- Z-1210-2024 · Jan 11, 2024 · Open, ClassifiedColonoscope, Model Number PCF-H190DL.
- Z-0416-2024 · Oct 16, 2023 · Open, ClassifiedGIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
- Z-0418-2024 · Oct 16, 2023 · Open, ClassifiedPCF-H190DL, EVIS EXERA III COLONOVIDEOSCOPE
- Z-0417-2024 · Oct 16, 2023 · Open, ClassifiedGIF-HQ190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
- Z-0415-2024 · Oct 16, 2023 · Open, ClassifiedCF-HQ190L EVIS EXERA III COLONOVIDEOSCOPE
- Z-0056-2025 · Sep 14, 2023 · Open, ClassifiedEVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light sources, monitors, EndoTherapy accessories, and other ancil
Other Colonoscope and accessories, flexible/rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls1,315 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page