The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00879123006240), received between Mar 15, 2023 and Aug 28, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Erratic or Intermittent Display” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Erratic or Intermittent Display2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    A FACILITY REPORTED THAT A GLIDESCOPE SPECTRUM DIRECTVIEW MAC S3 LARYNGOSCOPE WAS CONNECTED TO A GLIDESCOPE SYSTEM IN PREPARATION FOR INTUBATION. IT WAS REPORTED THAT THE LIGHT WAS WORKING DURING PRE-INTUBATION CHECKS; HOWEVER, THE LIGHT FAILED WHEN THE LARYNGOSCOPE WAS INSERTED INTO THE PATIENT'S MOUTH AND PRESSURE WAS APPLIED TO THE LARYNGOSCOPE FOR VISUALIZATION OF THE VOCAL CORDS. THE LARYNGOSCOPE WAS REMOVED FROM THE PATIENT WHICH THE LIGHT WAS THEN WORKING WHEN THERE WAS NO PRESSURE APPLIED ON THE LARYNGOSCOPE. A SECOND ATTEMPT WAS MADE TO INTUBATE AND THE LIGHT FAILED AGAIN. THE PROCEDURE WAS COMPLETED USING A NEW LARYNGOSCOPE WHICH WAS USED WITH NO ISSUES; HOWEVER IT RESULTED IN A DELAY IN INTUBATING THE PATIENT. NO HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20086558 ↗
  • MalfunctionMar 15, 2023Erratic or Intermittent Display; Erratic or Intermittent Display

    A CUSTOMER REPORTED THAT DURING A PATIENT PROCEDURE, USING A GLIDESCOPE SPECTRUM SINGLE-USE DIRECTVIEW MAC S3 VIDEO LARYNGOSCOPE, THE USER EXPERIENCED LIGHT LOSS/IMAGE LOSS DURING AN INTUBATION WHEN THE VIDEO LARYNGOSCOPE WAS MOVED. THE CUSTOMER REPORTED ISOLATING THE ISSUE TO THE MAC S3 VIDEO LARYNGOSCOPE, ITSELF, AND NOT THE CABLE OR MONITOR IT WAS USED WITH DURING THE PROCEDURE. NO DELAY IN THE PROCEDURE, USE OF A BACKUP DEVICE, OR HARM TO THE PATIENT WAS REPORTED.