Skip to content

GlideScopeRecalls & reported problems across all models

Verathon Medical (Canada) ULC · 36 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one GlideScope model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
347
Deaths reported
2
Injuries reported
3
Recalls
2

We found 36 models of GlideScope by Verathon Medical (Canada) ULC in the FDA's device registry (GUDID). Together they are named in 347 adverse event reports in the FDA's MAUDE database, received between Dec 23, 2019 and Sep 29, 2026 — classified as 2 deaths, 3 injuries, 342 malfunctions.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (245 reports).

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 8 reportsDec 2023: 4 reportsJan 2024: 10 reportsFeb 2024: 1 reportsMar 2024: 5 reportsApr 2024: 6 reportsMay 2024: 5 reportsJun 2024: 6 reportsJul 2024: 3 reportsAug 2024: 9 reportsSep 2024: 6 reportsOct 2024: 8 reportsNov 2024: 3 reportsDec 2024: 3 reportsJan 2025: 9 reportsFeb 2025: 5 reportsMar 2025: 3 reportsApr 2025: 8 reportsMay 2025: 9 reportsJun 2025: 7 reportsJul 2025: 7 reportsAug 2025: 12 reportsSep 2025: 4 reportsOct 2025: 6 reportsNov 2025: 6 reportsDec 2025: 8 reportsJan 2026: 5 reportsFeb 2026: 10 reportsMar 2026: 8 reportsApr 2026: 13 reportsMay 2026: 10 reportsJun 2026: 8 reportsJul 2026: 4 reportsAug 2026: 2 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

GlideScope recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Apr 29, 2024Open, Classified

    Due to software issues, their is the potential for loss of image or a degradation of the image when using the monitor with other connectable devices.

    Z-2022-2024 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗

  2. Apr 29, 2024Open, Classified

    Due to software issues, their is the potential for loss of image or a degradation of the image when using the monitor with other connectable devices.

    Z-2021-2024 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗

Most reported GlideScope problems

Device problems

  • Display or Visual Feedback Problem245
  • Erratic or Intermittent Display196
  • Poor Quality Image67
  • No Display/Image58
  • Detachment of Device or Device Component14
  • Intermittent Loss of Power13
  • Break9
  • Power Problem7

Patient problems

  • No Clinical Signs, Symptoms or Conditions332
  • Insufficient Information8
  • Low Oxygen Saturation4
  • No Consequences Or Impact To Patient4
  • Cardiac Arrest2
  • Unintended Extubation2
  • Airway Obstruction1
  • Diminished Pulse Pressure1

GlideScope models

36 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent GlideScope adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Verathon Medical (Canada) ULC products

All Verathon Medical (Canada) ULC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page