Skip to content

GlideScope® Video Baton QC LargeRecalls & reported problems

Verathon Medical (Canada) ULC · GlideScope® Video Baton QC Large · Class 1

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00879123007797), received between Aug 13, 2024 and Apr 29, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Display or Visual Feedback Problem5
  • Erratic or Intermittent Display3
  • Poor Quality Image2
  • Crack1
  • Loose or Intermittent Connection1
  • No Display/Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2025Erratic or Intermittent Display; Display or Visual Feedback Problem; Loose or Intermittent Connection; Erratic or Intermittent Display; Display or Visual Feedback Problem; Loose or Intermittent Connection

    A CUSTOMER REPORTED THAT DURING AN INTUBATION USING A GLIDESCOPE VIDEO BATON QUICKCONNECT LARGE, THE IMAGE WAS GOING IN AND OUT. FURTHERMORE, THE LIGHT FAILED TO ILLUMINATE CONSISTENTLY AND THE GLIDESCOPE CORE 2M QUICKCONNECT CABLE WAS NOT STAYING CONNECTED TO THE VIDEO BATON. THE PROCEDURE WAS COMPLETED USING A BACKUP DEVICE WHICH WAS MADE AVAILABLE IN AN UNSPECIFIED AMOUNT OF TIME. NO DELAY IN THE PROCEDURE OR HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21928738 ↗
  • MalfunctionJan 14, 2025Crack; Display or Visual Feedback Problem; Poor Quality Image; Crack; Display or Visual Feedback Problem; Poor Quality Image

    A CUSTOMER REPORTED WHILE USING A GLIDESCOPE VIDEO BATON QUICKCONNECT LARGE DURING A PATIENT PROCEDURE, THERE WAS A CRACK IN THE MIDDLE OF THE CAMERA RESULTING IN A BIG, WHITE LINE AND USERS BEING UNABLE TO SEE. THE PROCEDURE WAS COMPLETED USING A BACKUP DEVICE WHICH WAS MADE AVAILABLE IN AN USPECIFIED AMOUNT OF TIME. NO DELAY IN THE PROCEDURE OR HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21152256 ↗
  • MalfunctionOct 15, 2024Erratic or Intermittent Display; Display or Visual Feedback Problem; Erratic or Intermittent Display; Display or Visual Feedback Problem

    A CUSTOMER REPORTED WHILE USING A GLIDESCOPE VIDEO BATON QUICKCONNECT LARGE DURING A PATIENT PROCEDURE, THE LIGHT WAS FLASHING ON AND OFF AND INTERMITTENTLY DISPLAYING THE MESSAGE 'CAMERA NOT CONNECTED.' THE CUSTOMER TRIED SWAPPING THE VIDEO BATON FROM THE A PORT TO THE B PORT OF THE MONITOR BUT THE ISSUE PERSISTED. NO DELAY IN THE PROCEDURE, USE OF A BACKUP DEVICE, OR HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20457999 ↗
  • MalfunctionSep 10, 2024Erratic or Intermittent Display; Display or Visual Feedback Problem; Erratic or Intermittent Display; Display or Visual Feedback Problem

    THE CUSTOMER REPORTED THAT DURING A PATIENT PROCEDURE, USING A GLIDESCOPE VIDEO BATON QUICKCONNECT LARGE, THE PICTURE ON THE CONNECTED GLIDESCOPE CORE 10-INCH MONITOR FREEZES WHEN THE GLIDESCOPE CORE QUICKCONNECT CABLE AND VIDEO BATON ARE MOVED. NO DELAY IN THE PROCEDURE, USE OF A BACKUP DEVICE, OR HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20182304 ↗
  • MalfunctionAug 13, 2024No Display/Image; Display or Visual Feedback Problem; Poor Quality Image; No Display/Image; Display or Visual Feedback Problem; Poor Quality Image

    THE CUSTOMER REPORTED THAT DURING A PATIENT PROCEDURE, USING A GLIDESCOPE VIDEO BATON QUICKCONNECT LARGE, THE LIGHT WAS DIM WHEN TURNED ON; HOWEVER, THE SCREEN WENT BLACK ONCE THE VIDEO BATON WAS INSERTED INTO THE PATIENT'S AIRWAY. NO DELAY IN THE PROCEDURE, USE OF A BACKUP DEVICE, OR HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19973618 ↗

Other Laryngoscope, Rigid devices

All Laryngoscope, Rigid devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.