Skip to content

GlideScope® Spectrum QC Hyperangle S3Recalls & reported problems

Verathon Medical (Canada) ULC · GlideScope® Spectrum QC Hyperangle S3 · Class 1

All models of GlideScope (36)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00879123007926), received between May 20, 2026 and May 20, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Display or Visual Feedback Problem1
  • Poor Quality Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2026Display or Visual Feedback Problem; Poor Quality Image

    A CUSTOMER REPORTED THAT DURING AN EMERGENCY CESAREAN SECTION PROCEDURE, THE GLIDESCOPE SPECTRUM SINGLE-USE QC HYPERANGLE S3 BLADE PROVIDED A VERY CLOUDY IMAGE, RESULTING IN POOR VISUALIZATION. THE BLADE WAS REPLACED WITH ANOTHER BLADE, WHICH PERFORMED AS EXPECTED. NO DELAY IN PROCEDURE OR HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25231608 ↗

Other Laryngoscope, Rigid devices

All Laryngoscope, Rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page