Ti Screw ES 902572Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
All models of Ti Screw ES (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00880304404328), received between Oct 22, 2020 and Apr 14, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture5
- Failure to Advance2
- Material Twisted/Bent2
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Failure of Implant2
- No Consequences Or Impact To Patient2
- Foreign Body In Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 14, 2025Fracture; Fracture
IT WAS REPORTED THAT WHILE THE SURGEON WAS ATTEMPTING TO INSERT THE IMPLANT INTO THE PATIENT'S ROTATOR CUFF, THE IMPLANT FRACTURED. ALL PIECES AND FRAGMENTS OF THE IMPLANT WERE REMOVED FROM THE PATIENT. THERE WAS NO FOREIGN BODY RETAINED OR HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21833901 ↗MalfunctionFeb 28, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING ARTHROSCOPIC SURGERY IN THE SHOULDER JOINT, THE ANCHOR BROKE WHILE SCREWING INTO THE HUMERUS. NO ADVERSE EVENT HAS BEEN REPORTED AS A RESULT OF THE MALFUNCTION. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16454116 ↗InjuryMar 10, 2022Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 13721318 ↗MalfunctionOct 22, 2020Fracture; Failure to Advance; Material Twisted/Bent; Fracture; Failure to Advance; Material Twisted/Bent
IT WAS REPORTED APPROXIMATELY FOUR (4) WEEKS AGO DURING A ROTATOR CUFF REPAIR SURGERY, THE DEVICE FAILED TO ADVANCE THROUGH THE CORTEX. THE SURGEON TRIED TWO (2) DEVICES, ONE (1) BROKE AND ONE (1) BENT. THERE WAS A THIRTY (30) MINUTE DELAY WHILE THE SURGEON LOCATED AND REMOVED THE BROKEN PART FROM THE PATIENT. SURGEON USED A THIRD DEVICE TO COMPLETE THE SURGERY WITH NO ISSUES. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant; No Consequences Or Impact To Patient
MAUDE report 10717396 ↗MalfunctionOct 22, 2020Fracture; Failure to Advance; Material Twisted/Bent; Fracture; Failure to Advance; Material Twisted/Bent
IT WAS REPORTED APPROXIMATELY FOUR (4) WEEKS AGO DURING A ROTATOR CUFF REPAIR SURGERY, THE DEVICE FAILED TO ADVANCE THROUGH THE CORTEX. THE SURGEON TRIED TWO (2) DEVICES, ONE (1) BROKE AND ONE (1) BENT. THERE WAS A THIRTY (30) MINUTE DELAY WHILE THE SURGEON LOCATED AND REMOVED THE BROKEN PART FROM THE PATIENT. SURGEON USED A THIRD DEVICE TO COMPLETE THE SURGERY WITH NO ISSUES. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant; No Consequences Or Impact To Patient
MAUDE report 10717394 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page