Skip to content

OFFSET FLEX CLAMPING 31-555408 (UDI 00880304522749)Recalls & reported problems

Biomet Orthopedics, LLC · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Biomet orthopedic clamping instrument is used manually during hip surgery to prepare the femoral canal for an implant. The 33 reports mostly describe the broach handle breaking or snapping during broaching, with no effect on the patient noted in any of them. There have been no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00880304522749), received between Oct 11, 2019 and May 18, 2026 — classified as 33 malfunctions.

The most frequently reported specific device problem is “Fracture” (33 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture33
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions32
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 18, 2026Fracture

    IT WAS REPORTED THAT THE BACK OF THE FEMORAL BROACH HANDLE BROKE DURING BROACHING. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25201703 ↗
  • MalfunctionFeb 5, 2026Fracture

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24268365 ↗
  • MalfunctionFeb 5, 2026Fracture

    IT WAS REPORTED THAT THE BROACH HANDLE BROKE WHILE THE SURGEON WAS PREPARING THE FEMORAL CANAL. NO PATIENT HARM OR IMPACT REPORTED. IT WAS CONFIRMED THAT NO FURTHER INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24268242 ↗
  • MalfunctionJan 14, 2026Fracture

    IT WAS REPORTED THAT THE INSTRUMENT BROKE WHILE THE SURGEON WAS BROACHING THE FEMORAL CANAL. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24067731 ↗
  • MalfunctionDec 9, 2025Fracture; Fracture

    NO FURTHER INFORMATION AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23756763 ↗
  • MalfunctionOct 24, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23379445 ↗
  • MalfunctionSep 2, 2025Fracture; Fracture

    NO FURTHER INFORMATION AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22937465 ↗
  • MalfunctionMar 28, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE DISTAL PART OF THE BROACH HANDLE FRACTURED AND IS NO LONGER WELDED. THERE IS NO FURTHER INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21720639 ↗
  • MalfunctionJan 13, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING IMPACTION, THE HEAD OF THE BROACH HANDLE BROKE OFF. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21138187 ↗
  • MalfunctionNov 25, 2024Fracture; Fracture

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20773164 ↗
  • MalfunctionNov 6, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20620177 ↗
  • MalfunctionJun 12, 2024Fracture; Fracture

    IT WAS REPORTED THAT WHILE BROACHING FEMUR THE CIRCULAR STRIKE PLATE ON THE END OF THE BROACH HANDLE BROKE OFF. IT APPEARS THE WELD FAILED AFTER BEING USED MANY TIMES. NOTHING FELL INTO THE PATIENT AND THE PROCEDURE WAS COMPLETED USING A SECOND DEVICE. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19522072 ↗
  • MalfunctionMay 14, 2024Fracture; Fracture

    IT WAS REPORTED THAT WHILE EXTRACTING THE TRIAL, THE IMPACTION PLATE BROKE OFF THE BROACH HANDLE. THERE WAS NO HARM OR HEALTH CONSEQUENCES TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19309323 ↗
  • MalfunctionMay 8, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE WELD ON THE STRIKE PLATE BROKE WHILE BROACHING.THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19272796 ↗
  • MalfunctionMar 18, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE STRIKE PLATE BROKE OFF DURING BROACH REMOVAL. THE PROCEDURE WAS COMPLETED WITH ANOTHER INSTRUMENT. THERE WAS NO HARM OR HEALTH IMPACT TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18922305 ↗
  • MalfunctionJan 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT AS THE SURGEON WAS BROACHING, THE IMPACTION PLATE BROKE OFF. THE NURSES HAD TO LOCATE AND OPEN NEW STERILE BROACH HANDLE. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18424318 ↗
  • MalfunctionOct 24, 2023Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17996676 ↗
  • MalfunctionOct 24, 2023Fracture; Fracture

    THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17995302 ↗
  • MalfunctionSep 6, 2023Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17694195 ↗
  • MalfunctionJul 31, 2023Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17423585 ↗
  • MalfunctionSep 15, 2022Fracture; Fracture

    IT WAS REPORTED THAT THE IMPACTION PAD OF THE HANDLE WAS BROKEN DURING A PROCEDURE. THERE WAS NO IMPACT OR CONSEQUENCE TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15432919 ↗
  • MalfunctionJul 21, 2022Fracture; Fracture

    IT WAS REPORTED THE OFFSET BROACH FRACTURED AT THE WELD. NO IMPACT TO THE CASE TIME OR PATIENT. SECOND OFFSET BROACH WAS USED TO FINISH CASE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15074405 ↗
  • MalfunctionJun 2, 2022Fracture; Fracture

    IT WAS REPORTED THE END OF THE BROACH HANDLE BROKE OFF AFTER MULTIPLE STRIKES. SURGEON NOTICED THE STEM INSERTER WAS LOOSE. USED A DIFFERENT HANDLE TO COMPLETE THE CASE. NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14572577 ↗
  • MalfunctionMay 19, 2022Fracture; Fracture

    IT WAS REPORTED THAT THE INSTRUMENT FRACTURED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14437875 ↗
  • MalfunctionMay 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INITIAL HIP ARTHROPLASTY, WHILE THE SURGEON WAS RASPING IN THE FEMORAL CANAL, THE SURGEON STRUCK THE INSTRUMENT WITH A MALLET AND THE INSTRUMENT FRACTURED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14366037 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates