Skip to content

SafeSheath® Worley 00884450162286Recalls & reported problems

Merit Medical · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884450162286), received between Jan 6, 2022 and Jan 6, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 6, 2022Difficult to Remove; Difficult to Remove

    THE ACCOUNT ALLEGES THAT DURING A CRTD PROCEDURE, WHILE ATTEMPTING TO REMOVE THE SHEATH, THEY WERE UNABLE TO REMOVE THE ENTIRE SHEATH DUE TO STRONG RESISTANCE. THE CLINICAN CUT THE SHEATH WITH A SCALPEL LEAVEING ONE PART OF THE SHEATH ON THE ELECTRODE AND THE OTHER REMAINED INSIDE THE PATIENT'S VEIN. THE PROCEDURE WAS REPORTED TO BE COMPLETED WITH NO ADDITIONAL PATIENT CONSEQUENCE TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13177577 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page