Skip to content

Worley™ Adv. LVI 00884450483954Recalls & reported problems

Merit Medical Systems, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884450483954), received between Oct 28, 2025 and Oct 28, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Fracture1
  • Material Integrity Problem1
  • Material Separation1

Patient problems

  • Foreign Body In Patient1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2025Break; Fracture; Material Separation; Material Integrity Problem; Break; Fracture; Material Separation; Material Integrity Problem

    THE ACCOUNT ALLEGES THAT DURING PLACEMENT OF A DUAL ACCESS INTRODUCER (DAI-TRC), WHILE ADVANCING THE WORLEY ADVANCED SELECTIVE INTRODUCER TOWARD THE CORONARY SINUS TRIBUTARY, THE DISTAL TIP OF THE DEVICE DETACHED. AN AMPLATZ GOOSE NECK SNARE SYSTEM WAS INTRODUCED TO CAPTURE THE DETACHED SEGMENT. DURING RETRIEVAL, THE TIP FRAGMENTED FURTHER, AND ONE PORTION MIGRATED TOWARD THE PULMONARY ARTERY. DUE TO THE RISK OF ADDITIONAL FRAGMENTATION AND POTENTIAL VESSEL INJURY, RETRIEVAL EFFORTS WERE DISCONTINUED, AND TWO FRAGMENTS OF THE WORLEY INTRODUCER TIP WERE LEFT IN SITU. FOLLOWING THE EVENT, THE PHYSICIAN NOTED THAT A RETAINED FRAGMENT IN THE CORONARY SINUS CAUSED THROMBOSIS OF A CORONARY SINUS TRIBUTARY, AND ANOTHER FRAGMENTED PORTION HAD MIGRATED INTO THE SEGMENTAL PULMONARY ARTERY, POSING A RISK FOR THROMBOSIS. THE PATIENT WAS PLACED ON LIFELONG ANTICOAGULATION THERAPY. THE PROCEDURE WAS PERFORMED VIA A VENOUS APPROACH.

    Patient problems: Foreign Body In Patient; Thrombosis/Thrombus

    MAUDE report 23405387 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page