Arcadia™ GRecalls & reported problems

Merit Medical Systems, Inc. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884450815489), received between Jul 16, 2025 and Jul 16, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Fracture1
  • Material Integrity Problem1
  • Material Separation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 16, 2025Break; Fracture; Material Separation; Material Integrity Problem; Break; Fracture; Material Separation; Material Integrity Problem

    THE ACCOUNT ALLEGES THAT THE KYPHOPLASTY BALLOONS WERE DEPLOYED INTO THE VERTEBRAL BODIES OF THE T12 VERTEBRA UNDER FLUOROSCOPIC GUIDANCE AND SATISFACTORY BALLOON FILL WAS NOTED AT EACH LEVEL BILATERALLY. ON THE RIGHT SIDE, DURING BALLOON INFLATION, THE BALLOON RUPTURED DESPITE NO PRESSURE OVER 400MHG. DURING REMOVAL OF THE BALLOON, THE SMALL PLATINUM-IRIDIUM MARKER WAS RETAINED WITHIN THE VERTEBRAL BODY. AN INTRAOPERATIVE DECISION WAS MADE TO LEAVE THE RETAINED FOREIGN OBJECT WITHIN THE T12 BODY AS RETRIEVAL WOULD BE MORE DAMAGING, DANGEROUS, AND INVASIVE THAN THE RISKS OF LEAVING THE RETAINED OBJECT WITHIN. A DECISION WAS MADE NOT TO CEMENT THE RIGHT SIDE OF THE BODY AS THERE WAS POTENTIAL FOR LATERAL SIDE WALL FRACTURE DURING BALLOON RUPTURE. ON THE BACK TABLE, BONE CEMENT WAS MIXED AND AFTER WITHDRAWAL OF THE BALLOONS FROM THE VERTEBRAE, THE BONE CEMENT WAS INJECTED UNDER FLUOROSCOPIC GUIDANCE INTO THE T12 VERTEBRAE BILATERALLY WITH EXCELLENT BONE FILL, SHOWING FRACTURE FILL WITH THE CEMENT. AFTER WITHDRAWAL OF THE KYPHOPLASTY CANNULA, AP X-RAYS WERE TAKEN DEMONSTRATING EXCELLENT BONE FILL AT T12. NO PATIENT INJURY.

    Patient problems: Foreign Body In Patient

    MAUDE report 22521299 ↗