SureSigns VS4 863283 (UDI 00884838028890)Recalls & reported problems
Philips Medical Systems Hsg · Class 2
8 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 8
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00884838028890), received between Apr 6, 2017 and May 13, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “No Audible Alarm” (4 reports).
8 FDA recall records name this device's identifier, 8 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 19, 2019Open, Classified
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Z-1318-2019 · Root cause: Software design · FDA record ↗
- Apr 19, 2019Open, Classified
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Z-1320-2019 · Root cause: Software design · FDA record ↗
- Apr 19, 2019Open, Classified
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Z-1317-2019 · Root cause: Software design · FDA record ↗
- Apr 19, 2019Open, Classified
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Z-1321-2019 · Root cause: Software design · FDA record ↗
- Apr 19, 2019Open, Classified
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Z-1319-2019 · Root cause: Software design · FDA record ↗
- Apr 19, 2019Open, Classified
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Z-1324-2019 · Root cause: Software design · FDA record ↗
- Apr 19, 2019Open, Classified
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Z-1323-2019 · Root cause: Software design · FDA record ↗
- Apr 19, 2019Open, Classified
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Z-1322-2019 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- No Audible Alarm4
- Detachment of Device or Device Component1
- Device Operates Differently Than Expected1
- Display or Visual Feedback Problem1
- Energy Output Problem1
- Protective Measures Problem1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
PHILIPS RECEIVED A COMPLAINT ON THE SURESIGNS VS4 NBP, SPO2 INDICATING THE SYSTEM CONTINUOUSLY FREEZES. THE DEVICE WAS IN USE MONITORING A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22011065 ↗MalfunctionMar 12, 2024No Audible Alarm; No Audible Alarm
IT WAS REPORTED THE SPEAKER DOES NOT WORK AND A LOUDSPEAKER ALARM IS TRIGGERED, THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT, NO ADVERSE EVENT OR PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18886136 ↗MalfunctionSep 21, 2020Detachment of Device or Device Component
THE CUSTOMER REPORTED THE MONITOR FELL FROM THE HOLDER. THE CUSTOMER CONFIRMED THAT THE DEVICE WAS NOT CONNECTED TO A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10560554 ↗MalfunctionNov 8, 2019No Audible Alarm
THE CUSTOMER REPORTED THE PROBLEM OF AFTER REPLACING THE SPEAKER, STILL NO SOUND AND WOULD LIKE TO ORDER MAIN BOARD. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9294909 ↗MalfunctionMay 29, 2019Protective Measures Problem
THE CUSTOMER REPORTED THE PROBLEM OF ILLEGAL DOWNLOAD MESSAGE SHOWING ON VS4 UNIT IS FROZEN/NOT IN FRONT OF THE UNIT. NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8649310 ↗MalfunctionNov 13, 2018Energy Output Problem
THE CUSTOMER REPORTED A SPEAKER DOES NOT WORK. THE CUSTOMER REPORTED THAT THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE COMPLAINT/EVENT. THERE WAS NO ADVERSE EVENT TO A PATIENT OR USER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8063605 ↗MalfunctionOct 12, 2018No Audible Alarm
THE CUSTOMER REPORTED THAT DEVICE HAS A NO AUDIO/SPEAKER MALFUNCTION. THE CUSTOMER REPORTED THAT IT WAS UNKNOWN AS TO WHETHER THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7960301 ↗MalfunctionJul 11, 2017No Audible Alarm
THE CUSTOMER REPORTED THAT THE DEVICE HAD AN AUDIO FAILURE. THERE WAS NO REPORT OF A DEATH OR SERIOUS INJURY, NOR WAS THERE A REPORT OF ANY ADVERSE IMPACT TO ANY USER OR PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6701026 ↗MalfunctionApr 6, 2017Device Operates Differently Than Expected
THE CUSTOMER STATED THAT THE UNIT IS FREEZING UP. THERE WAS NO PATIENT IMPACT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6464728 ↗
Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,541 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,493 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,172 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths941 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports