Skip to content

SureSigns VS4 recallZ-1322-2019 · Philips North America, LLC · initiated Apr 19, 2019

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.

FDA root cause category: Software design

What the company is doing

On April 19, 2019 Philips Healthcare issued URGENT MEDICAL DEVICE CORRECTION notices to customers via courier service. Actions to be taken by customer: Upon receipt of this notification, promptly perform the software upgrade to each of your SureSigns VS3/VS4 monitors, as called out in the attached Pulse Rate Software-Installation Instructions. It details how to obtain and install the system software update. It also provides instructions on how to export the file that is created during software installation and instructions where to email the file back to Philips. If you need any further information or support concerning this issue, please contact your local Philips representative or call us at 1-800-722-9377.

Recall details

Recall number
Z-1322-2019
Status
Open, Classified
Initiated by the company
Apr 19, 2019
Recalling firm
Philips North America, LLC
Quantity
12719
Product code
MHX
FDA submissions
K163649
Distribution
Worldwide distribution. US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea Republic, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Martinique, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Philippines, Portugal, Qatar, Reunion, Romania, Rwanda, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Kingdom, United Arab Emirates,Venezuela, Vietnam, and Zimbabwe.

Product description

SureSigns VS3 NBP, SpO2, Temp, Rec, Product Number 863074. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).

Code information (models, lots, serial numbers)

Serial Numbers: All Including UDIs: (01)00884838028890(21) and (01)00884838087075(21) followed by serial number.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 82684).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page