SureSigns VS4 recallZ-1323-2019 · Philips North America, LLC · initiated Apr 19, 2019
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
FDA root cause category: Software design
What the company is doing
On April 19, 2019 Philips Healthcare issued URGENT MEDICAL DEVICE CORRECTION notices to customers via courier service. Actions to be taken by customer: Upon receipt of this notification, promptly perform the software upgrade to each of your SureSigns VS3/VS4 monitors, as called out in the attached Pulse Rate Software-Installation Instructions. It details how to obtain and install the system software update. It also provides instructions on how to export the file that is created during software installation and instructions where to email the file back to Philips. If you need any further information or support concerning this issue, please contact your local Philips representative or call us at 1-800-722-9377.
Recall details
- Recall number
- Z-1323-2019
- Status
- Open, Classified
- Initiated by the company
- Apr 19, 2019
- Recalling firm
- Philips North America, LLC
- Quantity
- 61522
- Product code
- MHX
- FDA submissions
- K163649
- Distribution
- Worldwide distribution. US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea Republic, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Martinique, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Philippines, Portugal, Qatar, Reunion, Romania, Rwanda, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Kingdom, United Arab Emirates,Venezuela, Vietnam, and Zimbabwe.
Product description
SureSigns VS4 NBP, SpO2, Product Number 863283. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
Code information (models, lots, serial numbers)
Serial Numbers: All Including UDIs: (01)00884838028890(21) and (01)00884838087075(21) followed by serial number.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SureSigns VS4 863283Philips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)8 recalls9 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 82684).
- Z-1317-2019 · Open, ClassifiedSureSigns VS3 NBP, Product Number 863069. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
- Z-1318-2019 · Open, ClassifiedSureSigns VS3 NBP Temp, Product Number 863070. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
- Z-1319-2019 · Open, ClassifiedSureSigns VS3 NBP, SpO2, Product Number 863071. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
- Z-1320-2019 · Open, ClassifiedSureSigns VS3 NBP, SpO2, Rec, Product Number 863072. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
- Z-1321-2019 · Open, ClassifiedSureSigns VS3 NBP, SpO2, Temp, Product Number 863073. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
- Z-1322-2019 · Open, ClassifiedSureSigns VS3 NBP, SpO2, Temp, Rec, Product Number 863074. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
- Z-1324-2019 · Open, ClassifiedSureSigns VS4 Government Bundle, Product Number 863286. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page