Skip to content

Brilliance CT Big BoreRecalls & reported problems

Philips Medical Systems Nederland B.V. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
7

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838059450), received between May 9, 2024 and May 9, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Noise, Audible” (1 report).

7 FDA recall records name this device's identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 13, 2025Class IIOpen, Classified

    The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

    Z-2597-2025 · Root cause: Process control · Recall details · FDA record ↗

  2. Apr 1, 2025Class IIOpen, Classified

    Unintended motion issues related to Interventional Control that may lead to collision of the Gantry/table with the operator or patient.

    Z-1660-2025 · Root cause: Software design · Recall details · FDA record ↗

  3. Oct 9, 2023Class IIOpen, Classified

    For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.

    Z-0293-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Oct 9, 2023Class IIOpen, Classified

    For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.

    Z-0294-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  5. May 1, 2021Class IIOpen, Classified

    Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

    Z-1983-2021 · Root cause: Device Design · Recall details · FDA record ↗

  6. Mar 28, 2018Class IITerminated

    A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.

    Z-2273-2018 · Root cause: Other · Recall details · FDA record ↗

  7. Feb 23, 2017Class IITerminated

    Philips identified a 4D CT pulmonary phase issue on Brilliance BigBore Oncology/Radiology CT with 4.2.0 software version through customer complaints. When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%).

    Z-1493-2017 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Apparent Adverse Event1
  • Noise, Audible1

Patient problems

  • Tinnitus1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 9, 2024No Apparent Adverse Event; Noise, Audible; No Apparent Adverse Event; Noise, Audible

    THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. THIS ISSUE REPORTED WAS AN INTERVENTIONAL ONCOLOGIST WHO SPENDS EXTENDED PERIODS OF TIME IN THE BB CT ROOM HAS DEVELOPED TINNITUS. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN INITIALLY DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS STARTED AN INVESTIGATION OF THE COMPLAINT.

    Patient problems: Tinnitus

    MAUDE report 19283137 ↗

FDA clearance and classification

The FDA regulates product code JAK (system, x-ray, tomography, computed) under 21 CFR 892.1750 as a Class II device. 21 CFR 892.1750 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Brilliance CT Big Bore

Has the Brilliance CT Big Bore been recalled?

Yes. 7 FDA recall records name this device's identifier: 5 still open and 2 closed (terminated) by the FDA. The most recent, Z-2597-2025 (Class II), began on Aug 13, 2025: The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Brilliance CT Big Bore?

The FDA's MAUDE database lists 1 adverse event report for this device, received on May 9, 2024. That report lists device problem “Noise, Audible” (1); patient problem “Tinnitus” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Brilliance CT Big Bore?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Brilliance CT Big Bore, and is it still sold?

The FDA's device database lists Philips Medical Systems Nederland B.V. as the company responsible for it (brand Brilliance CT Big Bore, model Brilliance CT Big Bore). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page