Brilliance CT Big BoreRecalls & reported problems
Philips Medical Systems Nederland B.V. · Class 2
5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 7
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838059450), received between May 9, 2024 and May 9, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Noise, Audible” (1 report).
7 FDA recall records name this device's identifier, 5 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Aug 13, 2025Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Z-2597-2025 · Root cause: Process control · FDA record ↗
- Apr 1, 2025Open, Classified
Unintended motion issues related to Interventional Control that may lead to collision of the Gantry/table with the operator or patient.
Z-1660-2025 · Root cause: Software design · FDA record ↗
- Oct 9, 2023Open, Classified
For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.
Z-0293-2024 · Root cause: Under Investigation by firm · FDA record ↗
- Oct 9, 2023Open, Classified
For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.
Z-0294-2024 · Root cause: Under Investigation by firm · FDA record ↗
- May 1, 2021Open, Classified
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Z-1983-2021 · Root cause: Device Design · FDA record ↗
- Mar 28, 2018Terminated
A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
Z-2273-2018 · Root cause: Other · FDA record ↗
- Feb 23, 2017Terminated
Philips identified a 4D CT pulmonary phase issue on Brilliance BigBore Oncology/Radiology CT with 4.2.0 software version through customer complaints. When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%).
Z-1493-2017 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- No Apparent Adverse Event1
- Noise, Audible1
Patient problems
- Tinnitus1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 9, 2024No Apparent Adverse Event; Noise, Audible; No Apparent Adverse Event; Noise, Audible
THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. THIS ISSUE REPORTED WAS AN INTERVENTIONAL ONCOLOGIST WHO SPENDS EXTENDED PERIODS OF TIME IN THE BB CT ROOM HAS DEVELOPED TINNITUS. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN INITIALLY DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS STARTED AN INVESTIGATION OF THE COMPLAINT.
Patient problems: Tinnitus
MAUDE report 19283137 ↗
Other System, x-ray, tomography, computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls6 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Spectral CT on RailsPhilips Medical Systems Nederland B.V. · System, X-Ray, Tomography, Computed6 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.