Skip to content

Brilliance iCTRecalls & reported problems

Philips Medical Systems Nederland B.V. · Class 2

All models of Brilliance iCT (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
8

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884838059474), received between Dec 8, 2017 and Sep 18, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Device Inoperable” (2 reports).

8 FDA recall records name this device's identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 7, 2026Class IIOpen, Classified

    Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.

    Z-3040-2026 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗

  2. Jun 11, 2026Class IIOpen, Classified

    The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.

    Z-2650-2026 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗

  3. Jan 7, 2026Class IIOpen, Classified

    As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

    Z-1315-2026 · Root cause: Process change control · Recall details · FDA record ↗

  4. Aug 13, 2025Class IIOpen, Classified

    The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

    Z-2598-2025 · Root cause: Process control · Recall details · FDA record ↗

  5. Jun 8, 2021Class IITerminated

    When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.

    Z-2314-2021 · Root cause: Software design · Recall details · FDA record ↗

  6. May 1, 2021Class IIOpen, Classified

    Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

    Z-1990-2021 · Root cause: Device Design · Recall details · FDA record ↗

  7. Nov 10, 2017Class IITerminated

    Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.

    Z-0391-2018 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Nov 10, 2017Class IITerminated

    Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.

    Z-0388-2018 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Inoperable2
  • Arcing1
  • Degraded1
  • No Audible Alarm1
  • Signal Artifact/Noise1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • Insufficient Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 18, 2026Arcing

    The customer reported that the gantry shut down during the arterial phase of a control scan following embolization, resulting in a 30-minute delay. The patient¿s condition had reportedly deteriorated during the waiting period, and the patient was transferred back to the ICU (Intensive Care Unit). The scan was subsequently completed several hours later on a different system.\No intervention was required during the 30-minute waiting period in the Radiology department; however, it remains unknown whether treatment was required after the patient returned to the ICU. No permanent impairment was reported, and the patient has recovered.\\Based on the available information, this issue has been determined to be a reportable event.\\Philips has started an investigation into this complaint. We will file a follow-up eMDR at the completion of the investigation.

    Patient problems: Insufficient Information

    MAUDE report 50819028 ↗
  • MalfunctionMar 29, 2018Device Inoperable

    THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. A PHILIPS FIELD SERVICE ENGINEER (FSE) REPORTED THAT DURING PLANNED MAINTENANCE (PM), THE MICROPHONE WAS FOUND TO HAVE FAILED. THE FSE CONFIRMED THAT THE PATIENT COULD NOT BE HEARD BY THE OPERATOR DUE TO THIS FAILURE. IF THE OPERATOR IS UNABLE TO HEAR THE PATIENT DUE TO A FAILED MICROPHONE, THERE IS POTENTIAL FOR INJURY TO THE PATIENT. THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. THIS EVENT IS CURRENTLY UNDER INVESTIGATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7380950 ↗
  • MalfunctionMar 27, 2018Device Inoperable

    THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. THE CUSTOMER REPORTED THAT THE INTERCOM SYSTEM FAILED. THE PHILIPS FIELD SERVICE ENGINEER (FSE) CONFIRMED THAT THE OPERATOR COULD NOT HEAR THE PATIENT AT ALL TIMES. IF THE OPERATOR IS UNABLE TO HEAR THE PATIENT DUE TO A FAILED MICROPHONE, THERE IS POTENTIAL FOR INJURY TO THE PATIENT. THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7373347 ↗
  • MalfunctionJan 17, 2018No Audible Alarm

    THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. THE CUSTOMER REPORTED THAT THE GANTRY MICROPHONE FAILED AND THEREFORE, THE OPERATOR COULD NOT HEAR THE PATIENT. IF THE OPERATOR IS UNABLE TO HEAR THE PATIENT DUE TO A FAILED MICROPHONE, THERE IS POTENTIAL FOR INJURY TO THE PATIENT. THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. THIS EVENT IS CURRENTLY UNDER INVESTIGATION TO DETERMINE THE CAUSE OF THE EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7201709 ↗
  • MalfunctionDec 8, 2017Signal Artifact/Noise; Degraded

    ENGINEERING ANALYSIS CONCLUDED THAT THIS EVENT HAS THE POTENTIAL TO RESULT IN IMAGE MISINTERPRETATION DUE TO AN ARTIFACT FROM A DEGRADED COMPENSATOR WITHIN THE COLLIMATOR. THEREFORE, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7102098 ↗

FDA clearance and classification

Cleared through 510(k) K250648 on Jun 27, 2025 (decision: Substantially Equivalent). Submitted by Philips Medical Systems Nederland B.V. as “Philips iCT CT system”. The FDA review took 115 days. FDA record ↗

The FDA regulates product code JAK (system, x-ray, tomography, computed) under 21 CFR 892.1750 as a Class II device. 21 CFR 892.1750 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Brilliance iCT

Has the Brilliance iCT been recalled?

Yes. 8 FDA recall records name this device's identifier: 5 still open and 3 closed (terminated) by the FDA. The most recent, Z-3040-2026 (Class II), began on Aug 7, 2026: Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Brilliance iCT?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Dec 8, 2017 and Sep 18, 2026. In those reports, the most common specific device problems are “Device Inoperable” (2), “Arcing” (1) and “Degraded” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Brilliance iCT?

None of the reports are classified as deaths, 1 as an injury and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Brilliance iCT FDA approved?

It was cleared through 510(k) K250648 on Jun 27, 2025. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Brilliance iCT, and is it still sold?

The FDA's device database lists Philips Medical Systems Nederland B.V. as the company responsible for it (brand Brilliance iCT, model Brilliance iCT). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page