Transducer S7-3t 989605439232Recalls & reported problems
Philips Ultrasound LLC · Class 2
All models of Transducer S7-3t (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884838067530), received between Nov 17, 2017 and Jan 19, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Defective Device” (1 report).
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jul 3, 2025Class IIIOpen, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Z-2395-2025 · Root cause: Device Design · Recall details · FDA record ↗
- May 28, 2025Class IIOpen, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
Z-2060-2025 · Root cause: Process control · Recall details · FDA record ↗
- May 28, 2025Class IIOpen, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
Z-2061-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Defective Device1
- Device Operates Differently Than Expected1
- Mechanical Jam1
- Positioning Problem1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 19, 2021Mechanical Jam; Mechanical Jam
A CUSTOMER REPORTED AN S7-3T MODEL TRANSDUCER HAD AN ARTICULATION ISSUE DURING USE. THERE WAS NO INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11197935 ↗MalfunctionJul 1, 2020Defective Device
A CUSTOMER REPORTED AN S7-3T MODEL TRANSDUCER HAD NO ARTICULATION IN ANY DIRECTION. THERE WAS NO INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10221360 ↗MalfunctionMay 15, 2020Positioning Problem
A CUSTOMER REPORTED AN S7-3T MODEL TRANSDUCER HAD A LOSS OF ANGLE MOVEMENT DURING A TEE EXAMINATION. THERE WAS NO INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10063221 ↗MalfunctionNov 17, 2017Device Operates Differently Than Expected
THE CUSTOMER REPORTED THE S7-3T TRANSDUCER WILL NOT ARTICULATE. THERE WAS NO INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7042650 ↗
Other System, imaging, pulsed doppler, ultrasonic devices
- Transducer C5-1 989605412042Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic11 recalls0 reports
- Transducer X5-1 989605400802Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic10 recalls0 reports
- EPIQ 7 Diagnostic Ultrasound System 989605386721Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls2 deaths146 reports
- Transducer S7-3t 989605406772Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic5 recalls1 death18 reports
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed doppler, ultrasonic199 reports
- Transducer 3D9-3v 989605395622Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls1 report
- ACUSON Redwood Diagnostic Ultrasound System 11503314Siemens Medical Solutions USA, Inc. · System, imaging, pulsed doppler, ultrasonic3 recalls0 reports
- Transducer S8-3t 989605420183Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic2 recalls10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page