Skip to content

ACUSON Redwood Diagnostic Ultrasound System 11503314Recalls & reported problems

Siemens Medical Solutions USA, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 15, 2024Class IIOpen, Classified

    If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

    Z-3216-2024 · Root cause: Software design · Recall details · FDA record ↗

  2. Feb 23, 2024Class IIOpen, Classified

    On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.

    Z-1600-2024 · Root cause: Software design · Recall details · FDA record ↗

  3. Jul 13, 2023Class IIOpen, Classified

    If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6 transducer and viewing in the Dual format visualization function, which may lead to misdiagnosis of a patient's condition or influence patient management decisions in a negative way.

    Z-2582-2023 · Root cause: Software design · Recall details · FDA record ↗

Other System, imaging, pulsed doppler, ultrasonic devices

All System, imaging, pulsed doppler, ultrasonic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page