Skip to content

DigitalDiagnost 712033Recalls & reported problems

Philips Medical Systems DMC GmbH · The DigitalDiagnost is a high-end digital radiography system, that includes - Accessory-lead apron for Vertical stand; - Additional SkyPlate charger; - Adjustable Straps for Patient Support; - Compression Belt (singles. pull); - Detector Holder for Patient Bed; - Hand grips (Pair of); - Handle for Skyplate Large; - Infusion bottle holder; - Mobile Detector Holder; - Paper roll holder; - Patient Support for Stitching; - Portable Panel Protector; - Remote control holder; - Set of 50 Hygienic Bags large; - Set of 50 Hygienic Bags small; - Side Bar; - Stretch Grip for VS/VM Stand; - Table top support · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884838074521), received between Aug 16, 2023 and Apr 1, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Device Fell” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 21, 2023Open, Classified

    Potential for units suspended on the ceiling to fail and fall.

    Z-0272-2024 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Device Fell4
  • Device Handling Problem4
  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 1, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE STITCHING STAND HYDRAULIC RAM HAS COME LOOSE. THE DEVICE WAS NOT IN CLINICAL USE, AND THERE WAS NO HARM TO THE PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24751380 ↗
  • MalfunctionDec 16, 2025Device Handling Problem; Device Fell; Device Handling Problem; Device Fell

    IT WAS REPORTED THAT THE SKYPLATE DETECTOR DROPPED AND HEAVY SHOCK REGISTERED TO THE DETECTOR. THE DEVICE WAS NOT IN CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23818699 ↗
  • MalfunctionSep 9, 2025Device Handling Problem; Device Fell; Device Handling Problem; Device Fell

    IT WAS REPORTED THAT THE DROPPED CASSETTE, IMAGE QUALITY PROBLEM. THE DEVICE WAS IN CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23002642 ↗
  • MalfunctionMar 6, 2025Device Handling Problem; Device Fell; Device Handling Problem; Device Fell

    IT WAS REPORTED THAT THE DETECTOR HAS A DENT IN THE SURFACE. THE DEVICE WAS IN CLINICAL USE AND THERE WAS NO HARM TO PATIENT OR USER. ADDITIONAL INFORMATION RECEIVED FROM REMOTE SERVICE ENGINEER AS DETECTOR DROPPED BY CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21534465 ↗
  • MalfunctionApr 30, 2024Device Handling Problem; Device Fell; Device Handling Problem; Device Fell

    THE CUSTOMER REPORTED THAT THE PORTABLE DETECTOR FELL. A DROPPED DETECTOR CAN LEAD IN WORST CASE TO A FRACTURE IN THE FOOT. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19213370 ↗
  • MalfunctionAug 16, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT THE TUBE IS NEARLY COMING OFF THE CEILING. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17555052 ↗

Other Solid state x-ray imager (flat panel/digital imager) devices

All Solid state x-ray imager (flat panel/digital imager) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.