DigitalDiagnost C50 (REF 712201)Recalls & reported problems

Philips Healthcare (Suzhou) Co., Ltd. · The DigitalDiagnost C50 is intended for use in generating radiographic images of human anatomy by qualified/trained doctor or technician. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications. · Class 2

All models of DigitalDiagnost C50 (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884838082366), received between Oct 24, 2023 and Jun 3, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Unintended System Motion” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K163410); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 9, 2023Open, Classified

    If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there is a potential that the Ceiling Suspension of the DigitalDiagnost C50 system may descend inadvertently if power to the system is interrupted, and may result in injuries to patients and users

    Z-1382-2023 · Root cause: Equipment maintenance · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unintended System Motion4
  • Device Difficult to Maintain1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 3, 2026Unintended System Motion

    THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. IT WAS REPORTED THAT THE COUNTERBALANCE WAS NOT SUFFICIENT AND THE TUBE DESCENDED DOWNWARD WHEN TURNING THE SYSTEM OFF. WHILE THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY IN THIS SPECIFIC CASE, SIMILAR EVENTS INVOLVING THIS ISSUE HAVE RESULTED IN SERIOUS INJURY IN THE PAST. THEREFORE, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25369799 ↗
  • MalfunctionFeb 9, 2026Mechanical Jam; Device Difficult to Maintain

    THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED. THE REPORTED ISSUE IS THAT THE TUBE IS TOO HEAVY TO MOVE WITH INCREASED RESISTANCE. AS A RESULT, THE USER HAS REPORTED EXPERIENCING PAIN IN THE ARMS, SHOULDERS AND NECK. THE USER HAS BEEN ON ACC NZ LEAVE FOR FOUR DAYS, RECEIVING TREATMENT FROM A PHYSIOTHERAPIST AND GP, AND HAS BEEN ADVISED TO CONSULT A SPECIALIST FOLLOWING AN MRI SCAN. THE FULL EXTENT OF THE INJURY IS CURRENTLY UNDER EVALUATION. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT.

    Patient problems: Pain

    MAUDE report 24290194 ↗
  • MalfunctionMar 12, 2024Unintended System Motion; Unintended System Motion

    THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. CUSTOMER REPORTED THAT AFTER THE SHUTTING THE SYSTEM DOWN, TUBE HEAD AUTOMATICALLY FELL DOWN. THIS EVENT WAS IDENTIFIED OUTSIDE OF CLINICAL USE, DURING REGULAR SHUT DOWN OF THE SYSTEM. HOWEVER, THIS MALFUNCTION WOULD POTENTIALLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY IF THIS WERE TO RECUR. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT.PHILIPS HAS STARTED THE INVESTIGATION OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18891775 ↗
  • MalfunctionNov 5, 2023Unintended System Motion; Unintended System Motion

    THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. THE ISSUE REPORTED WAS BALL TUBE DROPPED FROM TELESCOPIC TO THE LOWEST POSITION AND HIT THE EXAMINATION TABLE AFTER AN EXTERNAL POWER OUTAGE WHILE DEVICE WAS SITTING IDLE. THIS MALFUNCTION WOULD LIKELY TO CAUSE OR CONTRIBUTE TO A SERIOUS INJURY IF THIS WERE TO RECUR. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS COMPLETED THE INVESTIGATION OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18070842 ↗
  • MalfunctionOct 24, 2023Unintended System Motion; Unintended System Motion

    THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THE ISSUE REPORTED WAS THAT TUBE DESCENDED WHEN THE USER WAS MANUALLY ADJUSTING THE TUBE HEIGHT FOR A WALL STAND SCAN. THERE WAS NO INJURY/HARM HAPPENED IN THIS CASE. THIS MALFUNCTION WAS NOT LIKELY TO CAUSE DEATH OR SERIOUS INJURY IF IT WERE TO RECUR BASED ON THE RISK EVALUATION RESULT. THEREFORE, THIS ISSUE HAS BEEN DETERMINED NOT TO BE A REPORTABLE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17993494 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K163410. One submission often covers many models, so check the product and lot details in each record.

  • Z-1383-2023 · Mar 9, 2023 · Open, Classified
    DigitalDiagnost C50 1.1-intended for use in generating radiographic images of human anatomy Model Number : 712204
  • Z-1563-2022 · Feb 25, 2022 · Open, Classified
    DigitalDiagnost C50 1.1
  • Z-2498-2019 · Dec 21, 2018 · Terminated
    DigitalDiagnost C50, Stationary X-ray System

Other System, x-ray, stationary devices

All System, x-ray, stationary devices →