DigitalDiagnost C50 (REF 712204)Recalls & reported problems
Philips Healthcare (Suzhou) Co., Ltd. · Stationary x-ray system · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00884838095199), received between Jul 19, 2022 and Jul 3, 2025 — classified as 2 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K201725); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 9, 2023Open, Classified
If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there is a potential that the Ceiling Suspension of the DigitalDiagnost C50 system may descend inadvertently if power to the system is interrupted, and may result in injuries to patients and users
Z-1383-2023 · Root cause: Equipment maintenance · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem7
- Detachment of Device or Device Component3
- Device Dislodged or Dislocated1
- Flaked1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Bone Fracture(s)1
- Corneal Scar1
- Eye Injury1
- Eye Irritation1
- Foreign Body Sensation in Eye1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 3, 2025Mechanical Problem; Mechanical Problem
THIS COMPLAINT WAS REVIEWED. THE CUSTOMER REPORTED OBSERVING SHAVINGS DESCENDING FROM THE RAILS. THERE'S NO ALLEGATION OF DEATH OR SERIOUS INJURY. HOWEVER, CONSIDERING THAT A SIMILAR INCIDENT HAS PREVIOUSLY OCCURRED ON THE SAME DEVICE LEAD TO AN EYE INJURY, THIS EVENT HAS BEEN DETERMINED TO BE A REPORTABLE ONE. PHILIPS HAS STARTED THE INVESTIGATION OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22392517 ↗MalfunctionApr 27, 2025Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
IT WAS REPORTED THAT THE OVERHEAD TUBE FELL DOWN DUE TO DEFECTIVE CABLE CORRUGATE SUPPORT. THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. HOWEVER, IT IS UNCERTAIN WHETHER THIS EVENT IS LIKELY TO RECUR AND POTENTIALLY CAUSE SERIOUS INJURY. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS INITIATED AN INVESTIGATION INTO THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21915810 ↗MalfunctionApr 27, 2025Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
IT WAS REPORTED THAT THE CABLE CORRUGATED SUPPORT HAS COLLAPSED. ISSUE OCCURRED OUTSIDE OF CLINICAL USE; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. HOWEVER, IT IS UNCERTAIN WHETHER THIS EVENT IS LIKELY TO RECUR AND POTENTIALLY CAUSE SERIOUS INJURY. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS INITIATED AN INVESTIGATION INTO THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21915807 ↗MalfunctionMar 21, 2025Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE POSITIONING OF THE CEILING SUSPENSION, THE HANGER FAILED, CAUSING HEAVY CABLES TO FALL FROM THE CEILING. THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. HOWEVER, IT IS UNCERTAIN WHETHER THIS EVENT IS LIKELY TO RECUR AND POTENTIALLY CAUSE SERIOUS INJURY. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS INITIATED AN INVESTIGATION INTO THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21663441 ↗MalfunctionMar 12, 2025Mechanical Problem; Mechanical Problem
CUSTOMER REPORTED THAT THE CORRUGATED HOSE HAS FELL OVER THE TABLE WHILE A PATIENT WAS ON IT DURING PROCEDURE. HOSE WAS LAYING LOW BUT STILL ABLE TO WORK. THERE'S NO ALLEGATION OF DEATH OR SERIOUS INJURY, HOWEVER, IT IS UNCERTAIN WHETHER THIS EVENT IS LIKELY TO RECUR AND CAUSE A SERIOUS INJURY. THEREFORE, BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS STARTED THE INVESTIGATION OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21580630 ↗MalfunctionFeb 10, 2025Mechanical Problem; Mechanical Problem
THE ISSUE REPORTED WAS THE CABLE HOSE HOLDER CAME LOOSE AND HOSE DROPPED. THERE'S NO ALLEGATION OF DEATH OR SERIOUS INJURY. HOWEVER, IT IS UNCERTAIN WHETHER THIS EVENT IS LIKELY TO RECUR AND CAUSE A SERIOUS INJURY. THEREFORE, BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21343683 ↗InjuryFeb 10, 2025Flaked; Mechanical Problem; Flaked; Mechanical Problem
THE CUSTOMER REPORTED THAT DURING THE POSITIONING OF THE C50 CEILING SUSPENSION FOR A PATIENT'S X-RAY, THE TECHNICIAN, WHILE LOOKING UP TO MOVE THE TUBE, SUSTAINED AN INJURY WHEN A FLAKE OR SLIVER OF METAL ENTERED HER EYE. DESPITE WEARING GLASSES, THE METAL FLAKE CAUSED IRRITATION AND BECAME "EMBEDDED IN THE EYE." THE DOCTOR HAD TO REMOVE PART OF THE METAL FLAKE, WITH FOLLOW-UP TREATMENT REQUIRED TO REMOVE THE REMAINING FRAGMENT. UPON EXAMINATION OF THE EQUIPMENT, IT WAS NOTED THAT "METAL FLAKES WERE FOUND SURROUNDING THE TUBE HEAD BASE AND HAD ORIGINATED FROM THE TOP RAILS OF THE CEILING SUSPENSION." BASED ON AVAILABLE INFORMATION, THE REPORTED EVENT (METAL FLAKE IN THE EYE REQUIRING REMOVAL X2 TO PREVENT PERMANENT DAMAGE) MEETS CRITERIA FOR SERIOUS INJURY. THEREFORE, THIS EVENT HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS STARTED THE INVESTIGATION OF THIS EVENT.
Patient problems: Corneal Scar; Eye Injury; Eye Irritation; Foreign Body Sensation in Eye
MAUDE report 21343352 ↗InjuryJul 19, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
DURING AN ANKLE EXAM ON THE EXAM BED. THE EQUIPMENT WAS POWERED OFF, AND THE TUBE WAS SUSPENDED AND PRESSED DOWN ON THE PATIENT. THE PATIENT BELIEVED THAT THE ANKLE FRACTURE WAS CAUSED BY THE PRESSURE OF OUR PHILIPS EQUIPMENT. THE PATIENT THINKS THAT OUR PHILIPS EQUIPMENT WAS PRESSED DOWN AND CAUSED THE ANKLE FRACTURE. NOW THE HOSPITAL ASKS: IF THE PATIENT CANNOT ACCEPT THE HOSPITAL'S EXPLANATION, THE HOSPITAL NEEDS PHILIPS TO COOPERATE WITH THE EXPLANATION. PATIENT BELIEVES HE HAS BEEN HARMED. SUDDEN POWER FAILURE DURING NORMAL OPERATION.
Patient problems: Bone Fracture(s)
MAUDE report 15051938 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K201725. One submission often covers many models, so check the product and lot details in each record.
- Z-1382-2023 · Mar 9, 2023 · Open, ClassifiedPhilips DigitalDiagnost C50 -intended for use in generating radiographic images of human anatomy Model Number : 712201
- Z-1563-2022 · Feb 25, 2022 · Open, ClassifiedDigitalDiagnost C50 1.1
Other System, x-ray, stationary devices
- FDR Visionary Suite 566-16130-23Shimadzu Corporation · System, X-Ray, Stationary4 recalls0 reports
- FDR Visionary Suite 566-16130-31Shimadzu Corporation · System, X-Ray, Stationary2 recalls0 reports
- FDR Visionary Suite 566-16130-33Shimadzu Corporation · System, X-Ray, Stationary2 recalls0 reports
- Proteus XRaGE Hualun Medical Systems Co., Ltd · System, x-ray, stationary2 recalls0 reports
- Radiography 7300 C 712037Philips Medical Systems DMC GmbH · System, X-Ray, Stationary1 recall1 death8 reports
- DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Ltd. · System, x-ray, stationary1 recall5 reports
- DRX-Compass X-ray SystemCarestream Health, Inc. · System, x-ray, stationary1 recall2 reports
- Definium TempoGE Hualun Medical Systems Co., Ltd · System, x-ray, stationary1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.