DRX-Compass X-ray System (REF 8624181)Recalls & reported problems
Carestream Health, Inc. · X-RAY SYSTEM CONSOLE PDU 3 P 380-480 C/F · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 60889978624188), received between Jan 31, 2023 and Dec 22, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Loss of Data” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 30, 2023Open, Classified
After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury
Z-1206-2023 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- Loss of Data1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2025Loss of Data; Loss of Data
ON 15-DEC-2025, CARESTREAM HEALTH (CSH) WAS INFORMED OF AN INCIDENT RELATED TO THE DRX-COMPASS SYSTEM THAT OCCURRED ON 28-NOV-2025. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23866154 ↗MalfunctionJan 31, 2023Unintended Movement; Unintended Movement
ON 06-JAN-2023, CARESTREAM HEALTH WAS INFORMED BY A DEALER OF A CUSTOMER ISSUE "WHEN TRYING TO PUSH OTC Z AXIS MOTOR DOWN BUTTON, AFTER RELEASING THE BUTTON, OTC CONTINUED TO MOVE DOWN ON ITS OWN. TECH PRESSED THE "ALL BRAKE RELEASE" BUTTON TO STOP THE MOVEMENT. THE ISSUE WAS INTERMITTENT." THERE WAS NO INJURY TO THE PATIENT OR THE TECH/OPERATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16271767 ↗
Other System, x-ray, stationary devices
- FDR Visionary Suite 566-16130-23Shimadzu Corporation · System, X-Ray, Stationary4 recalls0 reports
- FDR Visionary Suite 566-16130-31Shimadzu Corporation · System, X-Ray, Stationary2 recalls0 reports
- FDR Visionary Suite 566-16130-33Shimadzu Corporation · System, X-Ray, Stationary2 recalls0 reports
- Proteus XRaGE Hualun Medical Systems Co., Ltd · System, x-ray, stationary2 recalls0 reports
- DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Ltd. · System, x-ray, stationary1 recall8 reports
- Radiography 7300 C 712037Philips Medical Systems DMC GmbH · System, X-Ray, Stationary1 recall1 death8 reports
- DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Ltd. · System, x-ray, stationary1 recall5 reports
- Definium TempoGE Hualun Medical Systems Co., Ltd · System, x-ray, stationary1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.