DRX-Compass X-ray System (REF 8624181)Recalls & reported problems

Carestream Health, Inc. · X-RAY SYSTEM CONSOLE PDU 3 P 380-480 C/F · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 60889978624188), received between Jan 31, 2023 and Dec 22, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Loss of Data” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 30, 2023Open, Classified

    After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury

    Z-1206-2023 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Loss of Data1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 22, 2025Loss of Data; Loss of Data

    ON 15-DEC-2025, CARESTREAM HEALTH (CSH) WAS INFORMED OF AN INCIDENT RELATED TO THE DRX-COMPASS SYSTEM THAT OCCURRED ON 28-NOV-2025. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23866154 ↗
  • MalfunctionJan 31, 2023Unintended Movement; Unintended Movement

    ON 06-JAN-2023, CARESTREAM HEALTH WAS INFORMED BY A DEALER OF A CUSTOMER ISSUE "WHEN TRYING TO PUSH OTC Z AXIS MOTOR DOWN BUTTON, AFTER RELEASING THE BUTTON, OTC CONTINUED TO MOVE DOWN ON ITS OWN. TECH PRESSED THE "ALL BRAKE RELEASE" BUTTON TO STOP THE MOVEMENT. THE ISSUE WAS INTERMITTENT." THERE WAS NO INJURY TO THE PATIENT OR THE TECH/OPERATOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16271767 ↗