Philips Medical Systems Hsg · CareEvent is a distributed alarming solution. With CareEvent the Philips Alarm distribution solution includes rules and escalation levels to deliver the right alarm to the right caregiver with the right information on a mobile device. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
4
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00884838099128), received between Nov 24, 2021 and May 11, 2026 — classified as 4 deaths, 3 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 28, 2024Class IIOpen, Classified

    It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.

    Z-1387-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Nov 28, 2023Class IIOpen, Classified

    Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition.

    Z-0732-2024 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Alarm System2
  • No Audible Alarm2
  • Wireless Communication Problem2
  • Alarm Not Visible1
  • Delayed Alarm1
  • No Visual Prompts/Feedback1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions2
  • Respiratory Arrest1
  • Tachycardia1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 11, 2026Wireless Communication Problem

    PHILIPS RECEIVED A COMPLAINT ON A CAREEVENT INDICATING A CONNECTION ERROR IN ALL TELEMETRY, CONNECTION ERROR ALSO COMES OUT IN THE CONTROL PANEL. IT SEEMS THAT EVERYTHING IS DOWN. THE DEVICE WAS IN USE ON A PATIENT AT TIME OF EVENT, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25136162 ↗
  • DeathFeb 6, 2025No Audible Alarm; No Audible Alarm

    THE CUSTOMER REPORTED THAT THE ALARMS DID NOT RING ON THE SMARTPHONE AND REQUESTED A LOG REVIEW. THE LOGS INDICATED THAT THE ALARMS OCCURRED ON THE MONITOR AND CENTRAL STATION AND WERE TRANSMITTED TO THE SMARTPHONE. ADDITIONAL INFORMATION IS REQUESTED FOR FURTHER CLARIFICATION. THE DEVICE WAS IN CLINICAL USE AT TIME OF EVENT AND A PATIENT DIED.

    Patient problems: Insufficient Information

    MAUDE report 21322752 ↗
  • DeathMar 22, 2024Delayed Alarm; Delayed Alarm

    A USER REPORT FROM ANSM WAS RECEIVED, INDICATING AROUND 8PM, AN ASYSTOLE ALERT WAS GENERATED ON THE SMARTPHONE. THE USER DISCOVERED THE PATIENT ON THE FLOOR IN CARDIO-RESPIRATORY ARREST. THE INTENSIVIST WAS ALERTED AND CPR WAS INITIATED AT 8:02PM AND CONTINUED FOR 20 MINUTES; HOWEVER, RESUSCITATION WAS UNSUCCESSFUL AND THE PATIENT'S DEATH WAS DECLARED AT 8:22PM. THE DATA AT THE CENTRAL STATION WAS REVIEWED, REVEALING A RED ALARM FOR VENTRICULAR TACHYCARDIA WAS GENERATED AT 7:39PM, WHICH TRANSITIONED TO ASYSTOLE AT 7:48PM. THE TELEMETRY DEVICE WAS SEQUESTERED AS A PRECAUTION AND RELATED SMARTPHONES WERE DETERMINED TO BE COMPLIANT. ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATED THERE WAS A DELAY IN RECEIVING THE ALARMS AT THE PHONE; HOWEVER, A LOG REVIEW REVEALED THE ASYSTOLE WAS GENERATED AT 19:48:50PM, THE ALARM SOUNDED AT THE CENTRAL STATION, AND THE ALARM WAS TRANSMITTED TO THE PHONES AT 19:48:50PM. THE ALARM WAS SILENCED AT 19:50:56 PRIOR TO THE ALARM BEING ACCEPTED AT THE PHONE, WHICH INTERRUPTED THE ALARM TRANSMISSION PROCESS. BASED ON THIS INFORMATION, THERE DOES NOT APPEAR TO BE A MALFUNCTION OF CAREEVENT WHICH CONTRIBUTED TO THE DEATH; HOWEVER, CLINICAL WORKFLOW, ALARM

    Patient problems: Tachycardia

    MAUDE report 18963712 ↗
  • DeathDec 6, 2023No Audible Alarm; No Audible Alarm

    THE CUSTOMER REPORTED RED ALARMS ARE NOT FLOWING TO THE CAREEVENT DEVICE/SYSTEM. THE DEVICE WAS IN USE ON A PATIENT. A PATIENT DEATH WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 18272721 ↗
  • DeathNov 17, 2022Device Alarm System; Wireless Communication Problem; No Visual Prompts/Feedback; No Visual Prompts/Feedback

    IT WAS REPORTED THE PATIENT MONITORING ALERTS RED AND YELLOW ARE NOT RECEIVED ON WAVEWARE SNPP PAGER IT IS UNKNOWN IF THE DEVICE WAS IN CLINICAL USE AT THE TIME THE ISSUE WAS DISCOVERED. A PATIENT DEATH OCCURRED.

    Patient problems: Insufficient Information

    MAUDE report 15816047 ↗
  • MalfunctionJan 1, 2022Alarm Not Visible; Alarm Not Visible

    THE CUSTOMER REPORTED THAT ALARMS FROM THEIR "BBRAUN" PUMPS WERE NOT SENDING ALARMS TO THE CENTRAL STATION FROM ONE ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13137443 ↗
  • MalfunctionNov 24, 2021Device Alarm System; Device Alarm System

    THE CUSTOMER REPORTED THAT A PATIENT WENT INTO SEVERE DESATURATION AND THEY WERE UNSURE IF THE EVENT ALARMED WITHIN CAREEVENT.

    Patient problems: Respiratory Arrest

    MAUDE report 12872721 ↗

FDA clearance and classification

The FDA regulates product code MSX (system, network and communication, physiological monitors) under 21 CFR 870.2300 as a Class II device. 21 CFR 870.2300 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CareEvent

Has the CareEvent been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1387-2024 (Class II), began on Feb 28, 2024: It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the CareEvent?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Nov 24, 2021 and May 11, 2026. In those reports, the most common specific device problems are “Device Alarm System” (2), “No Audible Alarm” (2) and “Wireless Communication Problem” (2); the most common patient problems are “Respiratory Arrest” (1) and “Tachycardia” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CareEvent?

4 reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the CareEvent, and is it still sold?

The FDA's device database lists Philips Medical Systems Hsg as the company responsible for it (brand CareEvent, model CareEvent). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System, network and communication, physiological monitors devices

All System, network and communication, physiological monitors devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page