Skip to content

BeneVision Central Monitoring System (UDI 06936415972417)Recalls & reported problems

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Class 2

All models of BeneVision Central Monitoring System (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
2
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 06936415972417), received between Jun 12, 2025 and Feb 25, 2026 — classified as 2 deaths, 3 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 18, 2025Class IIOpen, Classified

    When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing.

    Z-0920-2026 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Alarm System5
  • No Audible Alarm3

Patient problems

  • Insufficient Information3
  • Arrhythmia2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • DeathFeb 25, 2026Device Alarm System

    NO DEVICE MALFUNCTION WAS REPORTED BY THE CUSTOMER. IT WAS REPORTED THAT BETWEEN (B)(6) 2026, AT APPROXIMATELY 23:30 AND (B)(6) 2026, AT APPROXIMATELY 01:00, A PATIENT CONNECTED TO A TM80 UNIT AND SUBSEQUENTLY CONNECTED TO A BENEVISION DISTRIBUTED MONITORING SYSTEM (DMS) WORKSTATION EXPIRED WHILE NOT CONNECTED TO ECG LEADS. DURING THIS TIME PERIOD, THE PATIENT HAD REMOVED THE ECG LEADS. AS A RESULT, A "PATIENT DISCONNECTED FROM ECG" ALARM WAS PRESENT, AND NO CARDIAC RHYTHM ALARMS WERE GENERATED. CLINICAL STAFF WERE UNABLE TO ACKNOWLEDGE/SILENCE ALARMS FROM WORKSTATION DUE TO PRIOR EVENT.

    Patient problems: Arrhythmia

    MAUDE report 24446328 ↗
  • DeathFeb 25, 2026Device Alarm System

    NO DEVICE MALFUNCTION WAS REPORTED BY THE CUSTOMER. IT WAS REPORTED THAT BETWEEN (B)(6) 2026, AT APPROXIMATELY 23:30 AND (B)(6) 2026, AT APPROXIMATELY 01:00, A PATIENT CONNECTED TO A TM80 UNIT AND SUBSEQUENTLY CONNECTED TO A BENEVISION DISTRIBUTED MONITORING SYSTEM (DMS) WORKSTATION EXPIRED WHILE NOT CONNECTED TO ECG LEADS. DURING THIS TIME PERIOD, THE PATIENT HAD REMOVED THE ECG LEADS. AS A RESULT, A "PATIENT DISCONNECTED FROM ECG" ALARM WAS PRESENT, AND NO CARDIAC RHYTHM ALARMS WERE GENERATED. CLINICAL STAFF WERE UNABLE TO ACKNOWLEDGE/SILENCE ALARMS FROM WORKSTATION DUE TO PRIOR EVENT.

    Patient problems: Arrhythmia

    MAUDE report 24446326 ↗
  • MalfunctionJun 12, 2025Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm

    IT WAS REPORTED THAT ON (B)(6) 2025, A VENTRICULAR TACHYCARDIA (V-TACH) ALARM WAS MISSED; THE SYSTEM DID NOT GENERATE AN ALARM FOR V-TACH. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 22197277 ↗
  • MalfunctionJun 12, 2025Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm

    IT WAS REPORTED THAT ON (B)(6) 2025, AT 8:35AM, THE PATIENT ON A TM80 TELEMETRY UNIT (TELE ID 55) WAS IN VENTRICULAR TACHYCARDIA (V-TACH) FOR 5 BEATS AND THE SYSTEM DID NOT GENERATE A V-TACH ALARM. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22197272 ↗
  • MalfunctionJun 12, 2025Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm

    IT WAS REPORTED THAT ON (B)(6) 2025, AT 8:35AM, THE PATIENT ON A TM80 TELEMETRY UNIT (TELE ID 55) WAS IN VENTRICULAR TACHYCARDIA (V-TACH) FOR 5 BEATS AND THE SYSTEM DID NOT GENERATE A V-TACH ALARM. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22197267 ↗

FDA clearance and classification

The FDA regulates product code MSX (system, network and communication, physiological monitors) under 21 CFR 870.2300 as a Class II device. 21 CFR 870.2300 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the BeneVision Central Monitoring System

Has the BeneVision Central Monitoring System been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0920-2026 (Class II), began on Nov 18, 2025: When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the BeneVision Central Monitoring System?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Jun 12, 2025 and Feb 25, 2026. In those reports, the most common specific device problems are “Device Alarm System” (5) and “No Audible Alarm” (3); the most common patient problem is “Arrhythmia” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BeneVision Central Monitoring System?

2 reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the BeneVision Central Monitoring System MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the BeneVision Central Monitoring System, and is it still sold?

The FDA's device database lists Shenzhen Mindray Bio-Medical Electronics Co., Ltd. as the company responsible for it (brand BeneVision Central Monitoring System, model BeneVision). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System, Network And Communication, Physiological Monitors devices

All System, Network And Communication, Physiological Monitors devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page