BeneVision Central Monitoring System (UDI 06936415972417)Recalls & reported problems
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 06936415972417), received between Jun 12, 2025 and Feb 25, 2026 — classified as 2 deaths, 3 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 18, 2025Open, Classified
When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing.
Z-0920-2026 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- Device Alarm System5
- No Audible Alarm3
Patient problems
- Insufficient Information3
- Arrhythmia2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
DeathFeb 25, 2026Device Alarm System
NO DEVICE MALFUNCTION WAS REPORTED BY THE CUSTOMER. IT WAS REPORTED THAT BETWEEN (B)(6) 2026, AT APPROXIMATELY 23:30 AND (B)(6) 2026, AT APPROXIMATELY 01:00, A PATIENT CONNECTED TO A TM80 UNIT AND SUBSEQUENTLY CONNECTED TO A BENEVISION DISTRIBUTED MONITORING SYSTEM (DMS) WORKSTATION EXPIRED WHILE NOT CONNECTED TO ECG LEADS. DURING THIS TIME PERIOD, THE PATIENT HAD REMOVED THE ECG LEADS. AS A RESULT, A "PATIENT DISCONNECTED FROM ECG" ALARM WAS PRESENT, AND NO CARDIAC RHYTHM ALARMS WERE GENERATED. CLINICAL STAFF WERE UNABLE TO ACKNOWLEDGE/SILENCE ALARMS FROM WORKSTATION DUE TO PRIOR EVENT.
Patient problems: Arrhythmia
MAUDE report 24446328 ↗DeathFeb 25, 2026Device Alarm System
NO DEVICE MALFUNCTION WAS REPORTED BY THE CUSTOMER. IT WAS REPORTED THAT BETWEEN (B)(6) 2026, AT APPROXIMATELY 23:30 AND (B)(6) 2026, AT APPROXIMATELY 01:00, A PATIENT CONNECTED TO A TM80 UNIT AND SUBSEQUENTLY CONNECTED TO A BENEVISION DISTRIBUTED MONITORING SYSTEM (DMS) WORKSTATION EXPIRED WHILE NOT CONNECTED TO ECG LEADS. DURING THIS TIME PERIOD, THE PATIENT HAD REMOVED THE ECG LEADS. AS A RESULT, A "PATIENT DISCONNECTED FROM ECG" ALARM WAS PRESENT, AND NO CARDIAC RHYTHM ALARMS WERE GENERATED. CLINICAL STAFF WERE UNABLE TO ACKNOWLEDGE/SILENCE ALARMS FROM WORKSTATION DUE TO PRIOR EVENT.
Patient problems: Arrhythmia
MAUDE report 24446326 ↗MalfunctionJun 12, 2025Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm
IT WAS REPORTED THAT ON (B)(6) 2025, A VENTRICULAR TACHYCARDIA (V-TACH) ALARM WAS MISSED; THE SYSTEM DID NOT GENERATE AN ALARM FOR V-TACH. NO PATIENT INJURY WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 22197277 ↗MalfunctionJun 12, 2025Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm
IT WAS REPORTED THAT ON (B)(6) 2025, AT 8:35AM, THE PATIENT ON A TM80 TELEMETRY UNIT (TELE ID 55) WAS IN VENTRICULAR TACHYCARDIA (V-TACH) FOR 5 BEATS AND THE SYSTEM DID NOT GENERATE A V-TACH ALARM. NO PATIENT INJURY WAS REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22197272 ↗MalfunctionJun 12, 2025Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm
IT WAS REPORTED THAT ON (B)(6) 2025, AT 8:35AM, THE PATIENT ON A TM80 TELEMETRY UNIT (TELE ID 55) WAS IN VENTRICULAR TACHYCARDIA (V-TACH) FOR 5 BEATS AND THE SYSTEM DID NOT GENERATE A V-TACH ALARM. NO PATIENT INJURY WAS REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22197267 ↗
Other System, Network And Communication, Physiological Monitors devices
- Xhibit Xhibit CentralSpacelabs Healthcare (Washington), Inc · System, network and communication, physiological monitors3 deaths218 reports
- Infinity M540 MS26372Draeger Medical Systems, Inc. · System,Network And Communication,Physiological Monitors2 recalls9 reports
- CareEventPhilips Medical Systems Hsg · System, network and communication, physiological monitors2 recalls4 deaths7 reports
- XTR 96280Spacelabs Healthcare (Washington), Inc · SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS2 recalls0 reports
- CNS-2101 Central MonitorNihon Kohden Corporation · System, Network And Communication, Physiological Monitors3 deaths76 reports
- eCareManager 4.5Visicu, Inc. · System, network and communication, physiological monitors1 recall1 death1 report
- eCareManager 4.3Visicu, Inc. · System, network and communication, physiological monitors1 recall0 reports
- eCareManager 4.2Visicu, Inc. · System, network and communication, physiological monitors1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.