Vue PACS (Diagnostic) Vue PACS (Diagnostic) - 12Recalls & reported problems
Philips Medical Systems Nederland B.V. · The Vue PACS diagnostic client is web deployable and has a variety of user profiles applicable to different stakeholder in the cycle of care such as radiologists, referring physicians, technologists etc. With advanced streaming technology, the Vue PACS diagnostic client can perform well even over slow communication lines between the client and the server. Enterprise workflow capabilities allow users across the network to read and report from a global worklist, irrespective of the image acquisition location. The Vue Diagnostic workstation includes native clinical applications such as Lesion Management, CT Perfusion, MR Perfusion/Diffusion. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884838100350), received between Oct 28, 2024 and Apr 30, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Application Program Problem” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 5, 2026Class IIOpen, Classified
Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
Z-1692-2026 · Root cause: Software design · Recall details · FDA record ↗
- Oct 10, 2024Class IIOpen, Classified
A software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in misdiagnosis due to incorrect ischemic map and table values.
Z-0497-2025 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Application Program Problem1
- Application Program Version or Upgrade Problem1
- Computer Software Problem1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 30, 2026Computer Software Problem; Insufficient Device Problem Information
SMPA REFERENCE (B)(4). THE CLIENT WAS CRASHING DURING MDT (MULTIDISCIPLINARY TEAM CONFERENCE) AND CT IMAGES FAILED TO DISPLAY. PHILIPS HAS STARTED AN INVESTIGATION OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25040845 ↗MalfunctionDec 23, 2025Application Program Version or Upgrade Problem; Application Program Version or Upgrade Problem
PHILIPS RECEIVED A COMPLAINT ON VUE PACS DIAGNOSTIC INDICATING THAT AFTER UPGRADING THE SYSTEM FROM 12.2.8.570 TO 12.2.8.600 THE CUSTOMER EXPERIENCED BOOKMARKS RETURNING INCORRECT AND INACCURATE RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23881554 ↗MalfunctionOct 28, 2024Application Program Problem; Application Program Problem
IT HAS BEEN REPORTED TO PHILIPS THAT AN IDENTITY ERROR IN PEDIATRIC IMAGING REVEALED THE ABSENCE OF THE POSSIBILITY OF CORRECTION THE AGE ON THE RADIOLOGY PICTURES OF A CHILD WHEN THEY WERE DONE BY MISTAKE IN THE NAME OF ANOTHER CHILD. THE IDENTITY IS CORRECT, PATIENT IDENTIFICATION NUMBER, DATE OF BIRTH, HOWEVER THE AGE OF THE CHILD CORRESPONDS TO THE WRONG IDENTITY ON THE PICTURES DISTRIBUTED ON THE PACS. THE MODIFICATION OR DELETION OF THIS FIELD IS IMPOSSIBLE DURING THE RECONCILIATION OF THE CASE. IN THIS CASE, THIS 7-YEAR-OLD PATIENT HAS ANNOTATED PICTURES OF 11 YEARS. THERE WAS NO REPORTED HARM TO THE PATIENT. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. PHILIPS HAS DECIDED TO REPORT THIS EVENT OUT OF ABUNDANCE OF CAUTION. IT IS UNKNOWN AT THIS TIME IF THIS EVENT IF IT WERE TO RECUR COULD CAUSE SERIOUS INJURY OR DEATH. THIS ISSUE IS CURRENTLY UNDER INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20547647 ↗
Other System, image processing, radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page