Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to… recallZ-0497-2025 · Philips Medical Systems Technologies Ltd. · initiated Oct 10, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Technologies Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in misdiagnosis due to incorrect ischemic map and table values.

FDA root cause category: Software design

What the company is doing

Philips issued URGENT Medical Device Correction letter(# 2024-EI-RI-001) via Fed'x on 10/10/24 to US consignees. Philips Markets organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: To prevent the issue from occurring, Philips recommends that users not use the Perfusion Application Export function (as shown in Figure 4). When used without export functionality, the Perfusion Application performs as expected. " You may continue to use your system(s) in accordance with the intended use and by following the recommendation above. " If a study is reloaded in the Perfusion Application after an export action is performed, and the Perfusion/Diffusion Warning message is observed (shown in Figure 1), the ischemic map and table value calculations should not be used for diagnostic purposes. " Circulate this notice to all users of this device so that they are aware of the potential issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt of this letter via email to: [email protected]. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a time to install a software solution on your system(s) to resolve the issue (reference FCO78000003). Philips plans to begin installing the solution on affected systems in October 2024. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact Informatics Customer Support (1-877-328-2808, Option 4).

Recall details

Recall number
Z-0497-2025
Status
Open, Classified
Initiated by the company
Oct 10, 2024
Posted by the FDA
Nov 20, 2024
Recalling firm
Philips Medical Systems Technologies Ltd.
Quantity
2439 systems
Product code
LLZ
FDA submissions
K110919, K153103
Distribution
US Nationwide and Worldwide Distribution Foreign: Country Argentina Australia Austria Belgium Brazil Chile China Colombia Denmark Finland France Germany Gibraltar Greece Guernsey Holy See (Vatican City State) Hong Kong Hungary Indonesia Ireland Israel Italy Japan Jersey Jordan Lebanon Malaysia Mexico Netherlands New Zealand Norway Oman Panama Philippines Poland Portugal Romania Saudi Arabia Serbia Slovenia Spain Sweden Switzerland Taiwan Thailand United Kingdom

Product description

Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is to provide completely scalable local and wide area PACS (Picture Archiving and Communication System) solutions for hospital and related institutions/sites.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

UDI-DI: (01)00884838100350(11)200221(10)12.2.5; (01)00884838100350(11)200812(10)12.2.6.0; (01)00884838100350(11)201207(10)12.2.6.100; (01)00884838100350(11)210303(10)12.2.6.200; (01)00884838100350(11)210518(10)12.2.6.201; (01)00884838100350(11)210729(10)12.2.6.300; (01)00884838100350(11)210502(10)12.2.8.0; (01)00884838100350(11)220112(10)12.2.8.100; (01)00884838100350(11)221213(10)12.2.8.200; (01)00884838100350(11)230301(10)12.2.8.210; (01)00884838100350(11)230320(10)12.2.8.300; (01)00884838100350(11)230608(10)12.2.8.310; (01)00884838100350(11)230718(10)12.2.8.400; (01)00884838100350(11)230810(10)12.2.8.401; (01)00884838100350(11)230824(10)12.2.8.317; (01)00884838100350(11)231019(10)12.2.8.412; (01)00884838100350(11)231108(10)12.2.8.403; (01)00884838100350(11)231129(10)12.2.8.413; (01)00884838100350(11)231214(10)12.2.8.414; (01)00884838100350(11)231204(10)12.2.8.420; (01)00884838100350(11)240101(10)12.2.8.415; (01)00884838100350(11)240116(10)12.2.8.421; (01)00884838100350(11)240204(10)12.2.8.422; (01)00884838100350(11)240226(10)12.2.8.424; (01)00884838100350(11)240311(10)12.2.8.426; (01)00884838100350(11)240407(10)12.2.8.440; (01)00884838100350(11)240425(10)12.2.8.450; (01)00884838100350(11)240613(10)12.2.8.451; (01)00884838100350(11)240715(10)12.2.8.452; (01)00884838100350(11)240715(10)12.2.8.454; (01)00884838100350(11)230926(10)12.2.8.410

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K110919, K153103 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.