EPIQ 7 Ultrasound System 795200Recalls & reported problems

Philips Ultrasound LLC · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838107441), received between Jun 3, 2026 and Jun 3, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Battery Problem” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 14, 2022Class IICompleted

    Retroactive: Ultrasound may experience an unexpected automatic reboot, resulting in damage to transducer.

    Z-1007-2026 · Root cause: Software Design Change · Recall details · FDA record ↗

Most reported problems

Device problems

  • Battery Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 3, 2026Battery Problem

    IT WAS REPORTED EPIQ 7G ULTRASOUND SYSTEM'S UNINTERRUPTIBLE POWER SUPPLY (UPS) HAD AN ODOR AND ELECTROLYTE-LIQUID LEAKAGE. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER HARM ASSOCIATED WITH THE EVENT. THE SERVICE ENGINEER DETERMINED THE LIKELY CAUSE OF THE LEAKAGE WAS FROM THE BATTERY. THE EXTERNAL UPS WAS REPLACED TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25373831 ↗

Other System, Image Processing, Radiological devices