EPIQ 7 Ultrasound System 795200Recalls & reported problems
Philips Ultrasound LLC · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838107441), received between Jun 3, 2026 and Jun 3, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Battery Problem” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 14, 2022Class IICompleted
Retroactive: Ultrasound may experience an unexpected automatic reboot, resulting in damage to transducer.
Z-1007-2026 · Root cause: Software Design Change · Recall details · FDA record ↗
Most reported problems
Device problems
- Battery Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 3, 2026Battery Problem
IT WAS REPORTED EPIQ 7G ULTRASOUND SYSTEM'S UNINTERRUPTIBLE POWER SUPPLY (UPS) HAD AN ODOR AND ELECTROLYTE-LIQUID LEAKAGE. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER HARM ASSOCIATED WITH THE EVENT. THE SERVICE ENGINEER DETERMINED THE LIKELY CAUSE OF THE LEAKAGE WAS FROM THE BATTERY. THE EXTERNAL UPS WAS REPLACED TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25373831 ↗
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page