Philips EPIQ Ultrasound Systems (various models) used in… recallZ-1007-2026 · Philips Ultrasound, LLC · initiated Jul 14, 2022

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Ultrasound, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Retroactive: Ultrasound may experience an unexpected automatic reboot, resulting in damage to transducer.

FDA root cause category: Software Design Change

What the company is doing

On July 18, 2022, Phillips issued a product notification to affected consignees. Philips ask consignees to take the following actions: 1. If the X5-1c transducer is connected to the right-most port (when viewing from the front), this problem will not occur. Please ensure your X5-1c is connected to the right-most connector until the software update (FCO79500561) has been installed. 2. Please retain this letter with your product documentation until the software update (FCO79500561) has been implemented on your system. 3. Please circulate this notice to all users of this device so that they are aware of the issue. 4. A Philips Field Service Engineer will contact you to schedule the installation of the software update to resolve this issue (reference FCO79500561).

Recall details

Recall number
Z-1007-2026
Status
Completed
Initiated by the company
Jul 14, 2022
Posted by the FDA
Dec 31, 2025
Recalling firm
Philips Ultrasound, LLC
Quantity
1,721 systems
Product code
IYN
FDA submissions
K231190
Distribution
Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of Argentina, Australia, Belgium, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of Latvia, Lithuania, Morocco, Netherlands , New Zealand, Oman, Panama, Philippines, Poland, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom.

Product description

Philips EPIQ Ultrasound Systems (various models) used in conjunction with the X5-1c transducer REF: , , 795200, 795201, 795231, 795232, 795234 REF: 795117, 795122

Code information (models, lots, serial numbers)

Software Version Numbers: 9.0, 9.01, 9.02 795234/UDI: (01)00884838107540 795232/UDI: (01)00884838107533 795231/ UDI: (01)00884838097933 795201/UDI: (01)00884838047693 795200/ UDI: (01)00884838107441 795122/ UDI: (01)00884838097933 795117/ UDI: (01)00884838047693

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates