NA 399930Recalls & reported problems
Medtronic, Inc. · LEAD 399930 MONARCH 2X4 HINGED 30CM AA · Class 3 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885074087351), received between Mar 30, 2018 and Mar 30, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
9 other recalls cover devices cleared under the same FDA submission (P840001); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
INFORMATION WAS RECEIVED FROM A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). THE PATIENT STATED THAT THEIR ACTIVE LEAD IS ¿RUNNING HIGH¿ SO THEY ARE HAVING TO RE-ADJUST IT TO MAKE IT WORK. THE PATIENT STATED THAT THE ISSUE STARTED 3-4 YEARS AGO AND IS GETTING WORSE. THEIR LEAD IS CONNECTED TO A DIFFERENT MANUFACTURER¿S INS. THE PATIENT WAS REDIRECTED TO FOLLOW UP WITH THAT MANUFACTURER. THE PATIENT STATED THAT THE MANUFACTURE REPRESENTATIVE (REP) AND HEALTHCARE PROFESSIONAL (HCP) WANTED TO KNOW THE LEAD INFORMATION TO TRY AND REPROGRAM THE DEVICE WITH THE LEAD. PATIENT SERVICES REVIEWED THAT THE AVAILABLE REGISTRATION INFORMATION CAN BE FAXED TO A SECURE LINE AND THE REP/HCP CAN CALL DIRECTLY. THERE WERE NO PATIENT SYMPTOMS REPORTED AND NO COMPLICATIONS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7386042 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P840001. One submission often covers many models, so check the product and lot details in each record.
- Z-0372-2025 · Sep 18, 2024 · Open, ClassifiedMedtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
- Z-1691-2024 · Mar 28, 2024 · Open, ClassifiedIntellis AdaptiveStim Product Number 97715, Spinal Cord Stimulator
- Z-0780-2024 · Dec 19, 2023 · Open, ClassifiedModel A710 Intellis Clinician Programmer Application, version 2.0.97
- Z-2159-2021 · Jul 9, 2021 · Open, ClassifiedMedtronic Vectris SureScan MRI, Catalog Numbers: a) 977A260 b) 977A275 c) 977A290
- Z-1926-2016 · May 3, 2016 · TerminatedMedtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or li
- Z-1147-2015 · Jan 9, 2015 · TerminatedThe Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
- Z-1524-2014 · Feb 27, 2014 · TerminatedMedtronic neurostimulators. Restore (37711), RestoreADVANCED (37713), RestoreADVANCED SureScan MRI (97713), RestoreSENSOR (37714), RestoreSENSOR SureScan MRI (97714), PrimeAdvanced
- Z-0278-2014 · Sep 30, 2013 · TerminatedMedtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.
- Z-1536-06 · Aug 2, 2006 · TerminatedMedtronic 3777 Lead Kit, 1x8 Low Impedance Lead Kit for Spinal Cord Stimulation (SCS). Manufactured at: Medtronic, Inc., Villalba, Puerto Rico USA. Medtronic, Inc., Minneapolis
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page