CAPSUREFIX® NOVUS 407645 (UDI 00885074200675)Recalls & reported problems

Medtronic, Inc. · LEAD 407645 CAPSUREFIX NOVUS PACE US · Class 3 · Implant

All models of CAPSUREFIX NOVUS (17)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic CapsureFix Novus pacing lead is a wire that carries pacemaker signals to the heart. Of its 1,699 reports, most describe the lead failing to detect the heart's own beats or reading noise as heartbeats, and some describe pulses failing to make the heart beat. The reports include 12 deaths and 558 injuries, and some mention infection, chest pain and shortness of breath. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,706
Deaths reported
12
Injuries reported
561
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,706 adverse event reports that list this exact device (DI 00885074200675), received between Feb 9, 2016 and Sep 30, 2026 — classified as 12 deaths, 561 injuries, 1,133 malfunctions.

The most frequently reported specific device problem is “Under-Sensing” (861 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 11 reportsJan 2024: 22 reportsFeb 2024: 19 reportsMar 2024: 15 reportsApr 2024: 27 reportsMay 2024: 18 reportsJun 2024: 19 reportsJul 2024: 17 reportsAug 2024: 11 reportsSep 2024: 10 reportsOct 2024: 26 reportsNov 2024: 13 reportsDec 2024: 8 reportsJan 2025: 18 reportsFeb 2025: 14 reportsMar 2025: 13 reportsApr 2025: 9 reportsMay 2025: 13 reportsJun 2025: 17 reportsJul 2025: 13 reportsAug 2025: 10 reportsSep 2025: 10 reportsOct 2025: 9 reportsNov 2025: 22 reportsDec 2025: 10 reportsJan 2026: 12 reportsFeb 2026: 10 reportsMar 2026: 18 reportsApr 2026: 17 reportsMay 2026: 16 reportsJun 2026: 11 reportsJul 2026: 8 reportsAug 2026: 6 reportsSep 2026: 10 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202627
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Under-Sensing861
  • Over-Sensing283
  • Adverse Event Without Identified Device or Use Problem214
  • High Capture Threshold174
  • Pacing Problem97
  • Failure to Capture78
  • High impedance73
  • Device Dislodged or Dislocated66

Patient problems

  • No Clinical Signs, Symptoms or Conditions838
  • No Known Impact Or Consequence To Patient449
  • Unspecified Infection161
  • No Consequences Or Impact To Patient76
  • Chest Pain22
  • Dyspnea22
  • Sepsis18
  • Dizziness16

Recent adverse event reports

The 25 most recent of 1,706 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Implantable Cardioverter Defibrillator (ICD) system was explanted due to infection. No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50954560 ↗
  • InjurySep 30, 2026High Capture Threshold

    It was reported that the Right Atrial (RA) lead exhibited high thresholds. The RA lead was capped and replaced. No patient complicat ions have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946359 ↗
  • MalfunctionSep 30, 2026High impedance

    It was reported that the Right Atrial (RA) lead exhibited high impedance. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943032 ↗
  • MalfunctionSep 30, 2026Electrical /Electronic Property Problem; Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing  triggering device classified false termination of Atrial Tachycardia/Atrial Fibrillation (AT/AF) event(s) at times and a Atrial Lead Position Check failure. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942892 ↗
  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Implantable Cardioverter Defibrillator (ICD) patient experienced a pocket infection with swelling and redness. The device system was explanted. No further patient complications have been reported as a result of this event.

    Patient problems: Erythema; Medical device site infection; Swelling/ Edema

    MAUDE report 50903306 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced an infection. The organism was identified as Enterococcus faecalis. It was further noted that the patient experienced bacteremia and discomfort.  The patient was administered antibiotics. The IPG system was explanted. No further patient complications have been reported as a result of this event.

    Patient problems: Bacteremia; Discomfort

    MAUDE report 50882147 ↗
  • MalfunctionSep 11, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times and misclassification of episode type. The Implantable Cardioverter Defibrillator (ICD) de livered Anti-Tachycardia Pacing (ATP) and shocks for AT/AF with Rapid Ventricular Response (RVR) that was detected as Ventricular Tachycardia (VT). The ICD and lead remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50733926 ↗
  • MalfunctionSep 9, 2026High impedance; Pacing Problem; Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) episodes at time and resulted in inappropriate pacing. The RA lead also exhibited high impedance. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50685999 ↗
  • MalfunctionSep 2, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50566862 ↗
  • MalfunctionSep 1, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50546041 ↗
  • MalfunctionAug 20, 2026Under-Sensing

    Patient problems: Chest Pain

    MAUDE report 50356593 ↗
  • InjuryAug 19, 2026Over-Sensing

    It was reported that the Right Atrial (RA) lead exhibited Far Field R-Wave (FFRW) oversensing. The lead remains in use. No patient c omplications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50323722 ↗
  • MalfunctionAug 19, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50320850 ↗
  • MalfunctionAug 18, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50294296 ↗
  • MalfunctionAug 13, 2026Over-Sensing

    It was reported that the Right Atrial (RA) lead exhibited Far Field R-Wave (FFRW) oversensing noted on stored episodes triggering false Atrial Tachycardia/Atrial Fibrillation (AT/AF) detections. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50237230 ↗
  • InjuryAug 13, 2026Adverse Event Without Identified Device or Use Problem

    It was reported the patient's Implantable Cardioverter Defibrillator (ICD) system was explanted due to an infection. No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50234068 ↗
  • MalfunctionJul 30, 2026Over-Sensing

    It was reported that the Right Atrial (RA) lead exhibited oversensing on stored Atrial Tachycardia/Atrial Fibrillation (AT/AF) episodes triggering episode misclassification. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50086536 ↗
  • MalfunctionJul 13, 2026Signal Artifact/Noise

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) EXHIBITED NOISE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25793700 ↗
  • MalfunctionJul 9, 2026Signal Artifact/Noise; Over-Sensing; Under-Sensing; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS. NOISE, OVERSENSING AND UNDERSENSING. THE LEAD WILL BE MONITORED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25763441 ↗
  • MalfunctionJul 9, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25762155 ↗
  • InjuryJul 8, 2026High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS. THE RA LEAD WAS CAPPED AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25744642 ↗
  • MalfunctionJul 7, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) RESULTING IN TERMINATION OF DETECTED AT/AF EVENTS IN PROGRESS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25732818 ↗
  • MalfunctionJul 6, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING ON THE ATRIAL LEAD DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS AND UNNECESSARY PACING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25723199 ↗
  • MalfunctionJul 6, 2026Over-Sensing; Low impedance

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED LOW IMPEDANCE AND OVERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25718934 ↗
  • MalfunctionJun 25, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25626114 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P930039 on Mar 29, 1996. Submitted by Medtronic, Inc. as “Medtronic(R) Capsurefix Lead Model 4068,4067,4568”. The FDA review took 879 days. The FDA has since approved 267 supplements (changes to the device or its labeling), most recently on Jun 24, 2026. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

The FDA lists this type of device as life-sustaining or life-supporting.

FDA definition of this device type: Drug eluting permanent RV or RA pacemaker electrodes are used with compatible pacemakers or implantable cardioverter defibrillators (ICD) or cardiac rhythm therapy (CRT) devices to deliver pacing pulses to the heart.

Common questions about the CAPSUREFIX® NOVUS 407645

Has the CAPSUREFIX® NOVUS 407645 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00885074200675), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the CAPSUREFIX® NOVUS 407645?

The FDA's MAUDE database lists 1,706 adverse event reports for this device, received between Feb 9, 2016 and Sep 30, 2026. In those reports, the most common specific device problems are “Under-Sensing” (861), “Over-Sensing” (283) and “High Capture Threshold” (174); the most common patient problems are “Unspecified Infection” (161), “Chest Pain” (22) and “Dyspnea” (22). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CAPSUREFIX® NOVUS 407645?

12 reports are classified as deaths, 561 as injuries and 1,133 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CAPSUREFIX® NOVUS 407645 FDA approved?

It was approved through premarket approval (PMA) P930039 on Mar 29, 1996.

Is the CAPSUREFIX® NOVUS 407645 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the CAPSUREFIX® NOVUS 407645, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand CAPSUREFIX® NOVUS, model 407645). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page