MOSAIC ULTRA® 305U29Recalls & reported problems

Medtronic, Inc. · VALVE 305U29 MOSAIC AOR ULTRA US · Class 3 · Implant

All models of MOSAIC ULTRA (40)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Mosaic Ultra aortic tissue valve is a replacement heart valve implanted surgically. The 55 reports mostly describe the valve narrowing or leaking after years in place, along with heavy calcium buildup, sometimes leading to a second valve being placed inside it through a catheter. All 55 reports involve injuries, with no deaths reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
55
Deaths reported
0
Injuries reported
55
Recalls naming it
0

Summary

The FDA's MAUDE database contains 55 adverse event reports that list this exact device (DI 00885074253275), received between Jul 7, 2017 and Mar 4, 2026 — classified as 55 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (36 reports).

3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information36
  • Intravalvular regurgitation11
  • Device Stenosis8
  • Gradient Increase8
  • Calcified3
  • Adverse Event Without Identified Device or Use Problem2
  • Incomplete Coaptation2
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Aortic Valve Stenosis26
  • Aortic Valve Insufficiency/ Regurgitation19
  • Insufficient Information17
  • No Information9
  • Aortic Regurgitation7
  • Aortic Insufficiency4
  • Dyspnea3
  • Calcium Deposits/Calcification2

Recent adverse event reports

The 25 most recent of 55 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 4, 2026Gradient Increase; Device Stenosis

    IT WAS REPORTED THAT 15 YEARS AND 10 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS VALVE STENOSIS AND HIGH GRADIENTS. SUBSEQUENTLY ON (B)(6) 2026, THE PATIENT SUCCESSFULLY HAD A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 24506485 ↗
  • InjuryOct 21, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 9 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS LOW EJECTION FRACTION AND DYSSYNCHRONY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Non specific EKG/ECG Changes

    MAUDE report 23349291 ↗
  • InjurySep 16, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 16 YEARS AND 4 MONTHS POST IMPLANT OF THIS MOSAIC AORTIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. EVIDENCE OF CALCIFICATION UNDER LEFT CORONARY CUSP (LCC) AND LEFT VENTRICULAR OUTFLOW TRACT (LVOT). CALCIFICATION ALSO EVIDENT IN LEFT CORONARY ARTERY (LCA). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Calcium Deposits/Calcification

    MAUDE report 23060867 ↗
  • InjuryMay 9, 2025Calcified; Material Split, Cut or Torn; Thickening of Material; Device Stenosis; Intravalvular regurgitation; Calcified; Material Split, Cut or Torn; Thickening of Material; Device Stenosis; Intravalvular regurgitation

    IT WAS REPORTED THAT A 29MM AORTIC TISSUE VALVE WAS EXPLANTED AND A 29MM AORTIC TISSUE VALVE OF A DIFFERENT MODEL WAS IMPLANTED IN THE AORTIC POSITION DURING A PROCEDURE ON (B)(6) 2025. THE REASON FOR THE REPLACEMENT WAS REPORTED AS SEVERE AORTIC REGURGITATION AND AORTIC STENOSIS. THE PATIENT ALSO HAD SYMPTOMS OF DYSPNEA ON EXERTION. A MAZE PROCEDURE WAS ALSO PERFORMED CONCOMITANTLY. DURING THE EXPLANT, IT WAS FOUND THAT THE VALVE HAD HEAVILY CALCIFIED LEAFLETS WITH HOLES AND DENSE PANNUS. NO PATIENT COMPLICATIONS HAVE BEENREPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Cusp Tear; Dyspnea; Foreign Body Reaction

    MAUDE report 21998753 ↗
  • InjuryMar 31, 2025Gradient Increase; Device Stenosis; Intravalvular regurgitation; Gradient Increase; Device Stenosis; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 15 YEARS AND 4 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVAL UATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS MODERATE AORTIC STENOSIS, SEVERE AORTIC REGURGITATION AND INCREASED GRADIENT. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY 23 DAYS LATER, THE PATIENT SUCCESSFULLY HAD AN AORTIC VALVE-IN-VALVE REPLACEMENT WITH A TRANSCATHETER VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis

    MAUDE report 21730113 ↗
  • InjuryMar 11, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION FROM THE PATIENT THAT 13 YEARS AND 9 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE OF A DIFFERENT MANUFACTURER WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21570889 ↗
  • InjuryJan 2, 2025Device Stenosis; Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 8 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 4 DAYS LATER, A DIFFERENT MANUFACTURERS TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 21053041 ↗
  • InjurySep 16, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 6 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM VALVE OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20231601 ↗
  • InjuryJul 15, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 1 MONTH AND 4 DAYS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 29MM VALVE OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19745299 ↗
  • InjuryJun 27, 2024Appropriate Device Problem Term/Code Not Available; Device Stenosis; Intravalvular regurgitation; Appropriate Device Problem Term/Code Not Available; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED SUBSEQUENTLY, 1 MONTH AND 25 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information

    MAUDE report 19625829 ↗
  • InjuryJun 27, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 10 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 27MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19624725 ↗
  • InjuryJun 26, 2024Calcified; Intravalvular regurgitation; Calcified; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 15 YEARS AND 1 MONTH POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC INSUFFICIENCY. IT WAS STATED THAT CALCIUM WAS SEEN UNDER THE VALVE LEAFLETS EXTENDING INTO THE LEFT VENTRICULAR OUTFLOW TRACT (LVOT). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Calcium Deposits/Calcification

    MAUDE report 19619610 ↗
  • InjuryApr 22, 2024Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 3 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC INSUFFICIENCY. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 1 DAY LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 19159438 ↗
  • InjuryApr 9, 2024Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT AS AORTIC STENOSIS AND MODERATE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information

    MAUDE report 19070199 ↗
  • InjuryMar 20, 2024Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFO WAS RECEIVED THAT THE REASON FOR REPLACEMENT WAS REPORTED AS SEVER AORTIC REGURGITATION (AR), MILD AORTIC STENOSIS (AS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information

    MAUDE report 18941207 ↗
  • InjuryFeb 15, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 4 MONTHS POST IMPLANT OF THIS 29 MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. THE PATIENT ULTIMATELY RECEIVED UNKNOWN MEDICATION TREATMENT WITHOUT SURGICAL INTERVENTION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 18711758 ↗
  • InjuryJan 26, 2024Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT AS AORTIC STENOSIS (AS) AND MODERATE AORTIC REGURGITATION (AR). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information

    MAUDE report 18591061 ↗
  • InjuryDec 20, 2023Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 15 YEARS AND 1 MONTH POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC AORTIC REGURGITATION (AR).  NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 18365405 ↗
  • InjuryDec 8, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 15 YEARS AND 10 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 18291465 ↗
  • InjuryNov 30, 2023Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION 13 YEARS AND 3 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE THAT THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 1 DAY LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Insufficient Information

    MAUDE report 18235892 ↗
  • InjuryNov 10, 2023Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation

    MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE REASON FOR INTERVENTION WAS DUE TO SEVERE PULMONARY REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Pulmonary Valve Insufficiency/ Regurgitation

    MAUDE report 18111482 ↗
  • InjuryOct 24, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 6 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 17999033 ↗
  • InjurySep 5, 2023Incomplete Coaptation; Incomplete Coaptation

    MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 11 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS FLAIL LEAFLET AND SEVERE AORTIC INSUFFICIENCY. IT WAS ALSO REPORTED THAT THE PATIENT HAD PULMONARY EDEMA. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Pulmonary Edema

    MAUDE report 17685012 ↗
  • InjuryJul 31, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis

    MAUDE report 17426032 ↗
  • InjuryMay 2, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE REASON FOR INTERVENTION WAS DUE TO SEVERE AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis; Insufficient Information

    MAUDE report 16851960 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.

  • Z-1880-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1881-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1225-2009 · Jan 26, 2009 · Terminated
    Medtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M

Other heart-valve, non-allograft tissue devices

All heart-valve, non-allograft tissue devices →